Detection and classification of gaseous microemboli during Pulsatile and nonpulsatile perfusion in a simulated neonatal CPB model

被引:29
|
作者
Uendar, Akif
Ji, Bingyang
Kunselman, Allen R.
Myers, John L.
机构
[1] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Pediat, Hershey, PA USA
[2] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Surg, Hershey, PA USA
[3] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Bioengn, Hershey, PA USA
[4] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Hlth Evaluat Sci, Hershey, PA USA
关键词
D O I
10.1097/MAT.0b013e3181588dc3
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
We compared the effects of perfusion modes (pulsatile vs. nonpulsatile) on gaseous microemboli delivery using the Emboli Detection and Classification (EDACTM) Quantifier at postpump, postoxygenator, and postarterial filter sites in a simulated pediatric cardiopulmonary bypass (CPB) model. The mock loop was subjected to five different pump flow rates of equal 100 ml/min intervals, ranging from 400 to 800 ml/min. When the target pump flow rate was achieved, 5 cc air was introduced into the venous line. The EDACTM system recorded gaseous microemboli counts simultaneously at three locations in 5-minute intervals. Regardless of the type of perfusion mode, when the pump flow rate was increased, more gaseous microemboli were generated at postpump site. Compared with nonpulsatile flow, pulsatile flow did deliver significantly more gaseous microemboli at postpump site, but there was no difference between two groups at postoxygenator and postarterial filter sites. Capiox Baby-RX hollow-fiber membrane oxygenator significantly reduced the gaseous microemboli counts in both groups at all five pump flow rates with either pulsatile flow or nonpulsatile flow in this model. Our results suggest that using this novel EDAC (TM) system, we could detect the size of gaseous microemboli, as small as 10 mu m, and the percentage of detected gaseous microemboli, <40 mu m, was about 90% in total gaseous microemboli counts at any flow rate with pulsatile or nonpulsatile flow.
引用
收藏
页码:725 / 729
页数:5
相关论文
共 50 条
  • [41] Error associated with the choice of an aortic cannula in measuring regional cerebral blood flow with microspheres during pulsatile CPB in a neonatal piglet model
    Undar, A
    Lodge, AJ
    Daggett, CW
    Runge, TM
    Ungerleider, RM
    Calhoon, JH
    ASAIO JOURNAL, 1997, 43 (05) : M482 - M486
  • [42] Quantification of perfusion modes in terms of surplus hemodynamic energy levels in a simulated pediatric CPB model
    Undar, Akif
    Ji, Bingyang
    Lukic, Branka
    Zapanta, Conrad M.
    Kunselman, Allen R.
    Reibson, John D.
    Weiss, William J.
    Rosenberg, Gerson
    Myers, John L.
    ASAIO JOURNAL, 2006, 52 (06) : 712 - 717
  • [43] Evaluation of Conventional Nonpulsatile and Novel Pulsatile Extracorporeal Life Support Systems in a Simulated Pediatric Extracorporeal Life Support Model
    Wang, Shigang
    Evenson, Alissa
    Chin, Brian J.
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2015, 39 (01) : E1 - E9
  • [44] Impact of Oxygenator Selection on Hemodynamic Energy Indicators Under Pulsatile and Nonpulsatile Flow in a Neonatal Extracorporeal Life Support Model
    Vasavada, Rahul
    Khan, Sameer
    Qiu, Feng
    Kunselman, Allen
    Uendar, Akif
    ARTIFICIAL ORGANS, 2011, 35 (06) : E101 - E107
  • [45] Evaluation of Quadrox-i and Capiox FX neonatal oxygenators with integrated arterial filters in eliminating gaseous microemboli and retaining hemodynamic properties during simulated cardiopulmonary bypass
    Lin, J.
    Dogal, N. M.
    Mathis, R. K.
    Qiu, F.
    Kunselman, A.
    Uendar, A.
    PERFUSION-UK, 2012, 27 (03): : 235 - 243
  • [46] Pulsatile perfusion improves regional myocardial blood flow during and after hypothermic cardiopulmonary bypass in a neonatal piglet model
    Ündar, A
    Masai, T
    Yang, SQ
    Eichstaedt, HC
    McGarry, MC
    Vaughn, WK
    Fraser, CD
    ASAIO JOURNAL, 2002, 48 (01) : 90 - 95
  • [47] Stolen blood flow: Effect of an open arterial filter purge line in a simulated neonatal CPB model
    Wang, Shigang
    Miller, Akemi
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (04) : 432 - 435
  • [48] Microvascular fluid exchange during pulsatile cardiopulmonary bypass perfusion with the combined use of a nonpulsatile pump and intra-aortic balloon pump
    Lundemoen, Steinar
    Kvalheim, Venny Lise
    Mongstad, Arve
    Andersen, Knut Sverre
    Grong, Ketil
    Husby, Paul
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2013, 146 (05): : 1275 - 1282
  • [49] In Vitro Comparison of the Delivery of Gaseous Microemboli and Hemodynamic Energy for a Diagonal and a Roller Pump During Simulated Infantile Cardiopulmonary Bypass Procedures
    Dhami, Ranjodh
    Wang, Shigang
    Kunselman, Allen R.
    Uendar, Akif
    ARTIFICIAL ORGANS, 2014, 38 (01) : 56 - 63
  • [50] A randomized trial of pulsatile perfusion using an intra-aortic balloon pump versus nonpulsatile perfusion on short-term changes in kidney function during cardiopulmonary bypass during myocardial Reperfusion
    Onorati, Francesco
    Presta, Pierangela
    Fuiano, Giorgio
    Mastroroberto, Pasquale
    Comi, Nicolino
    Pezzo, Francesco
    Tozzo, Carmela
    Renzulli, Attilio
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2007, 50 (02) : 229 - 238