Handling ability of gaseous microemboli of two pediatric arterial filters in a simulated CPB model

被引:6
|
作者
Strother, A. [1 ,2 ]
Wang, S. [2 ]
Kunselman, A. R.
Uendar, A. [2 ,3 ]
机构
[1] Ann Arbor Publ Sch, Huron High Sch, Ann Arbor, MI USA
[2] Penn State Hershey Pediat Cardiovasc Res Ctr, Dept Pediat, Hershey, PA USA
[3] Penn State Hershey Childrens Hosp, Penn State Hershey Coll Med, Penn State Milton S Hershey Med Ctr, Hershey, PA USA
来源
PERFUSION-UK | 2013年 / 28卷 / 03期
关键词
gaseous microemboli; pediatric arterial filter; cardiopulmonary bypass; NEONATAL CARDIOPULMONARY BYPASS; QUADROX-I; NONPULSATILE PERFUSION; GENERATION; PULSATILE; FLOW; CLASSIFICATION; OXYGENATORS; FILTRATION; CIRCUITS;
D O I
10.1177/0267659112475106
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objective: The purpose of this experiment was to compare the Sorin KIDS D131 and the Terumo Capiox AF02 pediatric arterial filters in a simulated CPB procedure to determine which filter is the better for clinical use. Methods: The experimental circuit was primed with an 800 ml combination of lactated Ringer's solution and human blood (hematocrit (Hct) 30%). The two filters were tested under flow rates of 500, 1000, and 1500 ml/min at room temperature and their purge lines opened and closed as 5cc of air was injected into the circuit. Results: As the flow rates increased, the number of gaseous microemboli (GME) being returned to the pseudo patient increased for both of the pediatric arterial filters. Having an open purge line increased the number of GME removed from the CPB circuit, caused less of a pressure drop than when closed and increased the total hemodynamic energy loss than when closed. Both of the filters performed and reacted similarly in decreasing GME, hemodynamic energy loss and pressure drop. The only minor difference was that the Capiox AF02 had slightly less stolen blood flow (109.5 +/- 1.7 ml/min at 500 ml/min, 114.7 +/- 1.1 ml/min at 1000 ml/min and 105.8 +/- 4.2 ml/min at 1500 ml/min) from the open purge line than the KIDS D131 (119.5 +/- 2.5 ml/min at 500 ml/min, 128.3 +/- 1.0 ml/min at 1000 ml/min and 126.3 +/- 3.1 ml/min at 1500 ml/min). Conclusion: Our study confirmed that both the Sorin KIDS D131 and the Terumo Capiox AF02 were equivalent in their ability to remove significant numbers of GME, the amount of pressure drop and the total hemodynamic energy loss across the arterial filters at the various flow rates. An arterial filter is not an option, but a necessity for removing microemboli delivered to the patient.
引用
收藏
页码:244 / 252
页数:9
相关论文
共 22 条
  • [1] The capability of trapping gaseous microemboli of two pediatric arterial filters with pulsatile and nonpulsatile flow in a simulated infant CPB model
    Wang, Shigang
    Win, Khin N.
    Kunselman, Allen R.
    Woitas, Karl
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 519 - 522
  • [2] Gaseous microemboli detection in a simulated pediatric CPB circuit using a novel ultrasound system
    Miller, Akemi
    Wang, Shigang
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 504 - 508
  • [3] Detection and classification of gaseous microemboli during Pulsatile and nonpulsatile perfusion in a simulated neonatal CPB model
    Uendar, Akif
    Ji, Bingyang
    Kunselman, Allen R.
    Myers, John L.
    ASAIO JOURNAL, 2007, 53 (06) : 725 - 729
  • [4] Comparison of two different blood pumps on delivery of gaseous microemboli during pulsatile and nonpulsatile perfusion in a simulated infant CPB model
    Wang, Shigang
    Kunselman, Allen R.
    Myers, John L.
    Undar, Akif
    ASAIO JOURNAL, 2008, 54 (05) : 538 - 541
  • [5] In Vitro Comparison of Pediatric Oxygenators With and Without Integrated Arterial Filters in Maintaining Optimal Hemodynamic Stability and Managing Gaseous Microemboli
    Moroi, Morgan
    Force, Madison
    Wang, Shigang
    Kunselman, Allen R.
    Undar, Akif
    ARTIFICIAL ORGANS, 2018, 42 (04) : 420 - 431
  • [6] Is it safe to conduct CPB procedures without arterial filters in neonatal and pediatric patients?
    Uendar, A.
    Wang, S.
    Palanzo, D.
    Baer, L. D.
    PERFUSION-UK, 2015, 30 (02): : 172 - 173
  • [7] Evaluation of Combined Extracorporeal Life Support and Continuous Renal Replacement Therapy on Hemodynamic Performance and Gaseous Microemboli Handling Ability in a Simulated Neonatal ECLS System
    Shank, Kaitlyn R.
    Profeta, Elizabeth
    Wang, Shigang
    O'Connor, Christian
    Kunselman, Allen R.
    Woitas, Karl
    Myers, John L.
    Undar, Akif
    ARTIFICIAL ORGANS, 2018, 42 (04) : 365 - 376
  • [8] An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Melchior, Richard W.
    Rosenthal, Tami
    Glatz, Andrew C.
    PERFUSION-UK, 2010, 25 (04): : 255 - 263
  • [9] An in vitro comparison of the ability of three commonly used pediatric cardiopulmonary bypass circuits to filter gaseous microemboli
    Qiu, Feng
    Talor, Jonathan
    Uendar, Akif
    PERFUSION-UK, 2011, 26 (02): : 167 - 168
  • [10] Evaluation of Quadrox-i and Capiox FX neonatal oxygenators with integrated arterial filters in eliminating gaseous microemboli and retaining hemodynamic properties during simulated cardiopulmonary bypass
    Lin, J.
    Dogal, N. M.
    Mathis, R. K.
    Qiu, F.
    Kunselman, A.
    Uendar, A.
    PERFUSION-UK, 2012, 27 (03): : 235 - 243