Microemboli detection and classification by innovative ultrasound technology during simulated neonatal cardiopulmonary bypass at different flow rates, perfusion modes, and perfusate temperatures

被引:26
|
作者
Schreiner, Robert S. [1 ]
Rider, Alan R. [1 ]
Myers, John W. [1 ]
Ji, Bingyang [1 ]
Kunselman, Allen R. [2 ]
Myers, John L. [1 ,3 ]
Uendar, Akif [1 ,3 ,4 ]
机构
[1] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Pediat, Hershey, PA 17033 USA
[2] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Hlth Evaluat Sci, Hershey, PA 17033 USA
[3] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Surg, Hershey, PA 17033 USA
[4] Penn State Childrens Hosp, Penn State Milton S Hershey Med Ctr, Penn State Coll Med, Dept Bioengn, Hershey, PA 17033 USA
关键词
D O I
10.1097/MAT.0b013e31816ecfff
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The objective of this study was to detect and classify the number and size of gaseous microemboli in a simulated pediatric model of cardiopulmonary bypass. Tests were conducted at five different flow rates (400-1,200 ml/min in 200 ml/min increments), pulsatile versus nonpulsatile perfusion modes, and under normothermic, hypothermic, and deep hypothermic (35 degrees C, 25 degrees C, and 15 degrees C) conditions, yielding 180 total experiments. The circuit was primed with lactated Ringer's solution and filled with heparinized bovine blood. At the beginning of each experiment, 5 ml of air were injected into the venous line via the luer port of the oxygenator. Microemboli were quantified and classified by size for 5 minute segments at three transducer sites: postpump, postoxygenator, and postarterial filter. The purge line of the arterial filter was closed during all experiments. In all but one experiment, 90% of emboli at the postpump site were found to be smaller than 40 mu m. At the postarterial filter site, nearly 99% of the emboli were smaller than 40 mu m. Additionally, increasing microemboli counts were observed when the flow rate was increased and when the temperature was decreased. Lower temperatures, higher flow rates, and pulsatile perfusion were all associated with higher emboli counts. The majority of gaseous microemboli found in the simulated circuit was significantly below 40 mu m; the smallest level detectable by traditional Doppler devices.
引用
收藏
页码:316 / 324
页数:9
相关论文
共 3 条
  • [1] Detection and classification of gaseous microemboli during Pulsatile and nonpulsatile perfusion in a simulated neonatal CPB model
    Uendar, Akif
    Ji, Bingyang
    Kunselman, Allen R.
    Myers, John L.
    ASAIO JOURNAL, 2007, 53 (06) : 725 - 729
  • [2] Delivery of gaseous microemboli with vacuum-assisted venous drainage during pulsatile and nonpulsatile perfusion in a simulated neonatal cardiopulmonary bypass model
    Wang, Shigang
    Baer, Larry
    Kunselman, Allen R.
    Myers, John L.
    Uendar, Akif
    ASAIO JOURNAL, 2008, 54 (04) : 416 - 422
  • [3] Comparison of four different pediatric 10F aortic cannulae during Pulsatile versus nonpulsatile perfusion in a simulated neonatal model of cardiopulmonary bypass
    Uendar, Akif
    Ji, Bingyang
    Rider, Alan
    Lukic, Branka
    Kunselman, Allen R.
    Weiss, William J.
    Myers, John L.
    ASAIO JOURNAL, 2007, 53 (06) : 778 - 784