A theoretical study of CsI:Tl columnar scintillator image quality parameters by analytical modeling

被引:12
|
作者
Kalyvas, N. [1 ]
Valais, I. [1 ]
Michail, C. [1 ]
Fountos, G. [1 ]
Kandarakis, I. [1 ]
Cavouras, D. [1 ]
机构
[1] Technol Educ Inst TEI Athens, Dept Biomed Engn, Athens 12210, Greece
关键词
CsI:Tl; Columnar phosphors; Modeling; MTF; DQE; FLAT-PANEL IMAGER; SPATIAL-RESOLUTION; RAY; EFFICIENCY; PERFORMANCE; DETECTOR; NOISE; PHOSPHORS; QUANTUM; OPTIMIZATION;
D O I
10.1016/j.nima.2015.01.027
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
Medical X-ray digital imaging systems such as mammography, radiography and computed tomography (CT), are composed from efficient radiation detectors, which can transform the X-rays to electron signal. Scintillators are materials that emit light when excited by X-rays and incorporated in X-ray medical imaging detectors. Columnar scintillator, like CsI:Tl is very often used for X-ray detection due to its higher performance. The columnar form limits the lateral spread of the optical photons to the scintillator output, thus it demonstrates superior spatial resolution compared to granular scintillators. The aim of this work is to provide an analytical model for calculating the MTF, the DQE and the emission efficiency of a columnar scintillator. The model parameters were validated against published Monte Carlo data The model was able to predict the overall performance of CsliTl scintillators and suggested an optimum thickness of 300 pm for radiography applications. (C) 2015 Elsevier B.V. All rights reserved
引用
收藏
页码:18 / 24
页数:7
相关论文
共 50 条
  • [21] Dual-energy x-ray computed tomography study based on CsI:Tl and LYSO:Ce scintillator combination
    Xia, Xiaoqin
    Hu, Xiaodong
    Zou, Jing
    JOURNAL OF APPLIED PHYSICS, 2021, 130 (23)
  • [22] Image quality of a photon-counting mammography system compared to digital mammography based on amorphous silicon with CsI-scintillator
    Fischmann, Arne
    Steidle, Guenther
    DIGITAL MAMMOGRAPHY, PROCEEDINGS, 2006, 4046 : 441 - 446
  • [23] Image analytical monitoring of paper quality - a feasibility study
    Dahl, Casper K.
    Minkkinen, Pentti
    Esbensen, Kim H.
    TAPPI JOURNAL, 2006, 5 (11): : 18 - 24
  • [24] An analytical method for determination of quality parameters in cotton plumes by digital image and chemometrics
    Goncalves, Maria Ivanda S.
    Vilar, Welma T. S.
    Medeiros, Everaldo Paulo
    Coelho Pontes, Marcio Jose
    COMPUTERS AND ELECTRONICS IN AGRICULTURE, 2016, 123 : 89 - 94
  • [25] Theoretical study on analytical potential function and spectroscopic parameters for CaF molecule
    Yang, CL
    Zhang, X
    Han, KL
    JOURNAL OF MOLECULAR STRUCTURE-THEOCHEM, 2004, 678 (1-3): : 183 - 188
  • [26] Analytical modeling of quality parameters in casting process - learning-based approach
    Suthar, Janak
    Persis, Jinil
    Gupta, Ruchita
    INTERNATIONAL JOURNAL OF QUALITY & RELIABILITY MANAGEMENT, 2023, 40 (08) : 1821 - 1858
  • [27] Image Quality Analysis of an aSi:H/CsI(Tl) Flat-Panel Based Digital Radiography System Using a Chest Phantom
    Kim, Sung-Hyun
    Kim, Do-Il
    Yoon, Jai-Woong
    Lee, Hyoung-Koo
    Suh, Tae-Seok
    Choe, Bo-Yong
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING 2006, VOL 14, PTS 1-6, 2007, 14 : 1559 - 1562
  • [28] α-Si:H/CsI(Tl) flat-panel versus computed radiography for chest imaging applications:: image quality metrics measurement
    Liu, XM
    Shaw, CC
    MEDICAL PHYSICS, 2004, 31 (01) : 98 - 110
  • [29] A theoretical study of intraocular lens tilt and decentration on perceptual image quality
    Turuwhenua, J
    OPHTHALMIC AND PHYSIOLOGICAL OPTICS, 2005, 25 (06) : 556 - 567
  • [30] PERCEPTUAL QUALITY OF VIDEO WITH QUANTIZATION VARIATION : A SUBJECTIVE STUDY AND ANALYTICAL MODELING
    Ou, Yen-Fu
    Zeng, Huiqi
    Wang, Yao
    2012 IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING (ICIP 2012), 2012, : 1505 - 1508