Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition

被引:29
|
作者
Santoso, Andrew P. [1 ]
Song, Kwang H. [2 ,3 ]
Qin, Yujiao [3 ]
Gardner, Stephen J. [3 ]
Liu, Chang [3 ]
Chetty, Indrin J. [3 ]
Movsas, Benjamin [3 ]
Ajlouni, Munther [3 ]
Wen, Ning [3 ]
机构
[1] Wayne State Univ, Dept Radiat Oncol, Sch Med, Detroit, MI 48201 USA
[2] Texas Oncol, Ft Worth, TX 76104 USA
[3] Henry Ford Hlth Syst, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
Cone-beam CT; Image guidance; Dosimetry; Motion management; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; RADIOTHERAPY; PROTOCOL; MOTION; CBCT;
D O I
10.1186/s13014-016-0677-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager (R). Methods: A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan (R) phantom with real-time position management (TM) system for gantry speeds varying between 1.0 to 6.0 degrees/s. Superior-inferior motion of the phantom was executed using a sinusoidal waveform with five second period. Scans were retrospectively sorted into 4 phases (CBCT-4 ph) and 10 phases (CBCT-10 ph); average 4DCBCT (CBCT-ave), using all image data from the 4DCBCT acquisitions was also evaluated. The 4DCBCT images were evaluated using the following image quality metrics: spatial resolution, contrast-to-noise ratio (CNR), and uniformity index (UI). Additionally, Hounsfield unit (HU) sensitivity compared to a baseline CBCT and percent differences and RMS errors (RMSE) of excursion were also determined. Imaging dose was evaluated using an IBA CC13 ion chamber placed within CIRS Thorax phantom using the same sinusoidal motion and image acquisition settings as mentioned above. Results: Spatial resolution decreased linearly from 5.93 to 3.82 lp/cm as gantry speed increased from 1.0 to 6.0 degrees/s. CNR decreased linearly from 4.80 to 1.82 with gantry speed increasing from 1.0 to 6.0 degrees/s, respectively. No noteworthy variations in UI, HU sensitivity, or excursion metrics were observed with changes in gantry speed. Ion chamber dose rates measured ranged from 2.30 (lung) to 5.18 (bone) E-3 cGy/mAs. Conclusions: A quantitative analysis of the Varian OBI's 4DCBCT capabilities was explored. Changing gantry speed changes the number of projections used for reconstruction, affecting both image quality and imaging dose if x-ray tube current is held constant. From the results of this study, a gantry speed between 2 and 3 degrees/s was optimal when considering image quality, dose, and reconstruction time. The future of 4DCBCT clinical utility relies on further investigation of image acquisition and reconstruction optimization.
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition
    Andrew P. Santoso
    Kwang H. Song
    Yujiao Qin
    Stephen J. Gardner
    Chang Liu
    Indrin J. Chetty
    Benjamin Movsas
    Munther Ajlouni
    Ning Wen
    Radiation Oncology, 11
  • [2] Gantry Speed Significantly Affects Image Quality and Imaging Dose for 4D Cone-Beam Computed Tomography On the Varian Edge Platform
    Santoso, A.
    Song, K.
    Gardner, S.
    Chetty, I.
    Wen, N.
    MEDICAL PHYSICS, 2015, 42 (06) : 3269 - 3270
  • [3] Actively triggered 4d cone-beam CT acquisition
    Fast, Martin F.
    Wisotzky, Eric
    Oelfke, Uwe
    Nill, Simeon
    MEDICAL PHYSICS, 2013, 40 (09)
  • [4] Optimizing 4D cone-beam CT acquisition protocol for external beam radiotherapy
    Li, Tianfang
    Xing, Lei
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2007, 67 (04): : 1211 - 1219
  • [5] Evaluation of 4D CT to 4D Cone-Beam CT Deformable Image Registration for Lung Cancer Adaptive Radiation Therapy
    Balik, S.
    Hugo, G.
    Weiss, E.
    Jan, N.
    Roman, N.
    Sleeman, W.
    Fatyga, M.
    Christensen, G.
    Murphy, M.
    Lu, J.
    Keall, P.
    Williamson, J.
    MEDICAL PHYSICS, 2012, 39 (06) : 3875 - 3875
  • [6] Individualizing 4D cone-beam CT (CBCT) acquisition protocol for external beam radiotherapy
    Mao, W.
    Li, T.
    Munro, P.
    Chao, M.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2006, 66 (03): : S146 - S147
  • [7] 4D Cone-Beam CT Acquisition Using Respiratory Phase Predication Technique
    Yan, P.
    Zhou, L.
    Li, Z.
    Chi, Y.
    Chang, J.
    Chao, C.
    MEDICAL PHYSICS, 2012, 39 (06) : 3602 - 3602
  • [8] Image Quality and Dose Measurement for Partial Cone-Beam CT
    Abouei, E.
    Ford, N.
    MEDICAL PHYSICS, 2014, 41 (06) : 143 - 143
  • [9] Cone-Beam CT for Breast Imaging: Radiation Dose, Breast Coverage, and Image Quality
    O'Connell, Avice
    Conover, David L.
    Zhang, Yan
    Seifert, Posy
    Logan-Young, Wende
    Lin, Chuen-Fu Linda
    Sahler, Lawrence
    Ning, Ruola
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2010, 195 (02) : 496 - 509
  • [10] Cone-beam CT versus Multidetector CT in Postoperative Cochlear Implant Imaging: Evaluation of Image Quality and Radiation Dose
    Helal, R. A.
    Jacob, R.
    Elshinnawy, M. A.
    Othman, A., I
    Al-Dhamari, I. M.
    Paulus, D. W.
    Abdelaziz, T. T.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2021, 42 (02) : 362 - 367