Evaluation of gantry speed on image quality and imaging dose for 4D cone-beam CT acquisition

被引:29
|
作者
Santoso, Andrew P. [1 ]
Song, Kwang H. [2 ,3 ]
Qin, Yujiao [3 ]
Gardner, Stephen J. [3 ]
Liu, Chang [3 ]
Chetty, Indrin J. [3 ]
Movsas, Benjamin [3 ]
Ajlouni, Munther [3 ]
Wen, Ning [3 ]
机构
[1] Wayne State Univ, Dept Radiat Oncol, Sch Med, Detroit, MI 48201 USA
[2] Texas Oncol, Ft Worth, TX 76104 USA
[3] Henry Ford Hlth Syst, Dept Radiat Oncol, Detroit, MI 48202 USA
关键词
Cone-beam CT; Image guidance; Dosimetry; Motion management; 4-DIMENSIONAL COMPUTED-TOMOGRAPHY; RADIATION-THERAPY; RADIOTHERAPY; PROTOCOL; MOTION; CBCT;
D O I
10.1186/s13014-016-0677-8
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: This study investigates the effect of gantry speed on 4DCBCT image quality and dose for the Varian On-Board Imager (R). Methods: A thoracic 4DCBCT protocol was designed using a 125 kVp spectrum. Image quality parameters were evaluated for 4DCBCT acquisition using Catphan (R) phantom with real-time position management (TM) system for gantry speeds varying between 1.0 to 6.0 degrees/s. Superior-inferior motion of the phantom was executed using a sinusoidal waveform with five second period. Scans were retrospectively sorted into 4 phases (CBCT-4 ph) and 10 phases (CBCT-10 ph); average 4DCBCT (CBCT-ave), using all image data from the 4DCBCT acquisitions was also evaluated. The 4DCBCT images were evaluated using the following image quality metrics: spatial resolution, contrast-to-noise ratio (CNR), and uniformity index (UI). Additionally, Hounsfield unit (HU) sensitivity compared to a baseline CBCT and percent differences and RMS errors (RMSE) of excursion were also determined. Imaging dose was evaluated using an IBA CC13 ion chamber placed within CIRS Thorax phantom using the same sinusoidal motion and image acquisition settings as mentioned above. Results: Spatial resolution decreased linearly from 5.93 to 3.82 lp/cm as gantry speed increased from 1.0 to 6.0 degrees/s. CNR decreased linearly from 4.80 to 1.82 with gantry speed increasing from 1.0 to 6.0 degrees/s, respectively. No noteworthy variations in UI, HU sensitivity, or excursion metrics were observed with changes in gantry speed. Ion chamber dose rates measured ranged from 2.30 (lung) to 5.18 (bone) E-3 cGy/mAs. Conclusions: A quantitative analysis of the Varian OBI's 4DCBCT capabilities was explored. Changing gantry speed changes the number of projections used for reconstruction, affecting both image quality and imaging dose if x-ray tube current is held constant. From the results of this study, a gantry speed between 2 and 3 degrees/s was optimal when considering image quality, dose, and reconstruction time. The future of 4DCBCT clinical utility relies on further investigation of image acquisition and reconstruction optimization.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Respiratory signal extraction for 4D CT imaging of the thorax from cone-beam CT projections
    Rit, S
    Sarrut, D
    Ginestet, C
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S533 - S534
  • [22] Respiratory signal extraction for 4D CT imaging of the thorax from cone-beam CT projections
    Rit, S
    Sarrut, D
    Ginestet, C
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2005, PT 1, 2005, 3749 : 556 - 563
  • [23] Optimization of a Linac-based 4D Cone-beam CT Acquisition Protocol for Image-guided Radiation Therapy
    Lu, J.
    Zheng, D.
    Weiss, E.
    Williamson, J.
    Keall, P.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2009, 75 (03): : S628 - S628
  • [24] Development of the 4D Cone-Beam CT Modelling Software to Investigate Effects of Imaging Parameters
    Kim, D.
    Kim, T.
    Kang, S.
    Suh, T.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [25] Iterative volume of interest based 4D cone-beam CT
    Martin, Rachael
    Ahmad, Moiz
    Hugo, Geoffrey
    Pan, Tinsu
    MEDICAL PHYSICS, 2017, 44 (12) : 6515 - 6528
  • [26] Low-dose cardiac 4D cone-beam CT image reconstruction using two-cycle data
    Wi, Sunhee
    Lee, Yunjeong
    Lee, Jiseoc
    Abbas, Sajid
    Cho, Seungryong
    Son, Kihong
    Kim, Jin Sung
    2014 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2014,
  • [27] Effect of dose reduction on image registration and image quality for cone-beam CT in radiotherapy
    Loutfi-Krauss, B.
    Koehn, J.
    Bluemer, N.
    Freundl, K.
    Koch, T.
    Kara, E.
    Scherf, C.
    Roedel, C.
    Ramm, U.
    Licher, J.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2015, 191 (02) : 192 - 200
  • [28] Image quality of an investigational imaging panel for use with the imaging beam line cone-beam CT
    Beltran, Chris
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2012, 13 (01): : 76 - 83
  • [29] Weight-bearing gantry-free cone-beam CT of the lumbar spine: Image quality analysis and dose efficiency
    Feldle, Philipp
    Grunz, Jan-Peter
    Kunz, Andreas Steven
    Patzer, Theresa Sophie
    Huflage, Henner
    Hendel, Robin
    Luetkens, Karsten Sebastian
    Erguen, Suleyman
    Bley, Thorsten Alexander
    Conrads, Nora
    EUROPEAN JOURNAL OF RADIOLOGY, 2023, 165
  • [30] Faster and lower dose imaging: evaluating adaptive, constant gantry velocity and angular separation in fast low-dose 4D cone beam CT imaging
    Lau, Benjamin K. F.
    Dillon, Owen
    Vinod, Shalini K. K.
    O'Brien, Ricky T. T.
    Reynolds, Tess
    MEDICAL PHYSICS, 2024, 51 (02) : 1364 - 1382