Evaluation and empirical analysis of an exact FBP algorithm for spiral cone-beam CT

被引:9
|
作者
Katsevich, A [1 ]
Lauritsch, G [1 ]
Bruder, H [1 ]
Flohr, T [1 ]
Stierstorfer, K [1 ]
机构
[1] Univ Cent Florida, Dept Math, Orlando, FL 32816 USA
关键词
exact inversion; cone-beam data; spiral tomography; filtered back-projection algorithm;
D O I
10.1117/12.481348
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Recently one of the authors proposed a reconstruction algorithm, which is theoretically exact and has the truly shift-invariant filtering and backprojection structure. Each voxel is reconstructed using the theoretically minimum section of the spiral, which is located between the endpoints of the PI segment of the voxel. Filtering is one-dimensional, performed along lines with variable tilt on the detector, and consists of five terms. We will present evaluation of the performance of the algorithm. We will also discuss and illustrate empirically the contributions of the five filtering terms to the overall image. A thorough evaluation proved the validity of the algorithm. Excellent image results were achieved even for high pitch values. Overall image quality can be regarded as at least equivalent to the less efficient, exact, Radon-based methods. However, the new algorithm significantly increases efficiency. Thus, the method has the potential to be applied in clinical scanners of the future. The empirical analysis leads to a simple, intuitive understanding of the otherwise obscure terms of the algorithm. Identification and skipping of the practically irrelevant fifth term allows significant speed-up of the algorithm due to uniform distance weighting.
引用
收藏
页码:663 / 674
页数:12
相关论文
共 50 条
  • [41] Evaluation of helical cone-beam CT reconstruction algorithms
    Köhler, T
    Bontus, C
    Brown, K
    Heuscher, D
    Grass, M
    Shechter, G
    Proksa, R
    2002 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-3, 2003, : 1217 - 1220
  • [42] An in vitro evaluation of cone-beam breast CT methods
    Zeng, Kai
    Fajardo, Laurie L.
    Kao, Simon
    Franken, Edmund A.
    Park, Jeong Mi
    Jing, Zhenxue
    Bai, Er-wei
    Wang, Ge
    JOURNAL OF X-RAY SCIENCE AND TECHNOLOGY, 2008, 16 (03) : 171 - 187
  • [43] Evaluation of the OSC-TV iterative reconstruction algorithm for cone-beam optical CT
    Matenine, Dmitri
    Mascolo-Fortin, Julia
    Goussard, Yves
    Despres, Philippe
    MEDICAL PHYSICS, 2015, 42 (11) : 6376 - 6386
  • [44] Performance evaluation of the backprojection filtered (BPF) algorithm in circular fan-beam and cone-beam CT
    LI Liang CHEN Zhi-Qiang ZHANG Li KANG Ke-Jun Department of Engineering Physics
    Nuclear Science and Techniques, 2006, (02) : 113 - 117
  • [45] Optimization of derivative kernels for exact cone-beam ROI reconstruction in spiral computed tomography
    Lauritsch, G
    Sourbelle, K
    Tam, KC
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2002, 49 (03) : 728 - 732
  • [46] Optimization of derivative kernels for exact cone-beam ROI reconstruction in Spiral Computed Tomography
    Lauritsch, G
    Sourbelle, K
    Tam, KC
    2001 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORDS, VOLS 1-4, 2002, : 1652 - 1656
  • [47] Approximate and exact cone-beam reconstruction with standard and non-standard spiral scanning
    Wang, Ge
    Ye, Yangbo
    Yu, Hengyong
    PHYSICS IN MEDICINE AND BIOLOGY, 2007, 52 (06): : R1 - R13
  • [48] Approximate and exact cone-beam reconstruction with standard and non-standard spiral scanning
    Wang, Ge
    Ye, Yangbo
    Yu, Hengyong
    DEVELOPMENTS IN X-RAY TOMOGRAPHY V, 2006, 6318
  • [49] Analytical geometric parameter calibration algorithm for cone-beam CT
    Chen, Lian
    Wu, Zhifang
    Liu, Ximing
    Yao, Ming
    Qinghua Daxue Xuebao/Journal of Tsinghua University, 2010, 50 (03): : 418 - 421
  • [50] The Improvement of Katsevich Reconstruction Algorithm Based on Cone-beam CT
    Wang, Li Fang
    MEASUREMENT TECHNOLOGY AND ITS APPLICATION, PTS 1 AND 2, 2013, 239-240 : 1148 - 1151