摘要
Recently one of the authors proposed a reconstruction algorithm, which is theoretically exact and has the truly shift-invariant filtering and backprojection structure. Each voxel is reconstructed using the theoretically minimum section of the spiral, which is located between the endpoints of the PI segment of the voxel. Filtering is one-dimensional, performed along lines with variable tilt on the detector, and consists of five terms. We will present evaluation of the performance of the algorithm. We will also discuss and illustrate empirically the contributions of the five filtering terms to the overall image. A thorough evaluation proved the validity of the algorithm. Excellent image results were achieved even for high pitch values. Overall image quality can be regarded as at least equivalent to the less efficient, exact, Radon-based methods. However, the new algorithm significantly increases efficiency. Thus, the method has the potential to be applied in clinical scanners of the future. The empirical analysis leads to a simple, intuitive understanding of the otherwise obscure terms of the algorithm. Identification and skipping of the practically irrelevant fifth term allows significant speed-up of the algorithm due to uniform distance weighting.
引用
收藏
页码:663 / 674
页数:12
相关论文
共 50 条
  • [31] Performance evaluation of local ROI algorithms for exact ROI reconstruction in spiral cone-beam computed tomography
    Sourbelle, K
    Lauritsch, G
    Tam, KC
    Noo, F
    Kalender, WA
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2001, 48 (03) : 697 - 702
  • [32] Exact local regions-of-interest reconstruction in spiral cone-beam filtered-backprojection CT: theory
    Tam, KC
    MEDICAL IMAGING 2000: IMAGE PROCESSING, PTS 1 AND 2, 2000, 3979 : 506 - 519
  • [33] Analysis of image noise with exact cone-beam reconstruction
    Lee, SW
    Wang, G
    RADIOLOGY, 2002, 225 : 253 - 253
  • [34] Advanced single-slice rebinning in cone-beam spiral CT
    Kachelriess, M
    Schaller, S
    Kalender, WA
    MEDICAL PHYSICS, 2000, 27 (04) : 754 - 772
  • [35] A backprojection-filtration algorithm for nonstandard spiral cone-beam CT with an n-PI-window
    Yu, HY
    Ye, YB
    Zhao, SY
    Wang, G
    PHYSICS IN MEDICINE AND BIOLOGY, 2005, 50 (09): : 2099 - 2111
  • [36] Exact reconstruction of volumetric images in reverse helical cone-beam CT
    Cho, Seungryong
    Xia, Dan
    Pelizzari, Charles A.
    Pan, Xiaochuan
    MEDICAL PHYSICS, 2008, 35 (07) : 3030 - 3040
  • [37] Motion artefact reduction for exact cone-beam CT reconstruction algorithms
    Koken, P
    Köhler, T
    Bontus, C
    Proksa, R
    2003 IEEE NUCLEAR SCIENCE SYMPOSIUM, CONFERENCE RECORD, VOLS 1-5, 2004, : 2985 - 2988
  • [38] Exact image reconstruction for selected ROI in helical cone-beam CT
    Pan, XC
    Zou, Y
    Xia, D
    Sidky, E
    Jiang, TZ
    2004 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-7, 2004, : 3773 - 3776
  • [39] Physical phantom studies of helical cone-beam CT with exact reconstruction
    Tan, Jun
    Li, H. Harold
    Klein, Eric
    Li, Hua
    Parikh, Parag
    Yang, Deshan
    MEDICAL PHYSICS, 2012, 39 (08) : 4695 - 4704
  • [40] Evaluation of Dose Calculation Based On Cone-Beam CT
    Abe, T.
    Tateoka, K.
    Fujimoto, K.
    Nakazawa, T.
    Saitou, Y.
    Nakata, A.
    Yano, M.
    Sakata, K.
    MEDICAL PHYSICS, 2014, 41 (06) : 259 - 259