Validation of SIMIND Monte Carlo Simulation Software for Modelling a Siemens Symbia T SPECT Scintillation Camera

被引:5
|
作者
Ejeh, John E. [1 ,2 ]
van Staden, Johan A. [1 ]
du Raan, Hanlie [1 ]
机构
[1] Univ Free State, Dept Med Phys, Bloemfontein, South Africa
[2] Univ Coll Hosp Ibadan, Dept Nucl Med, Ibadan, Nigeria
关键词
Computer simulation; Code validation; SIMIND; Septal penetration fraction; Monte Carlo; SEGMENTATION; SCATTER;
D O I
10.1007/978-981-10-9035-6_106
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
The use of Monte Carlo (MC) simulations in nuclear medicine is on the rise due to its usefulness in modelling many physical phenomena which are difficult or impossible to measure. To use MC for gamma camera image simulations, it is important to validate the simulated data with the measured data to ensure that what is simulated is a true representation of the physical system being modelled. Therefore, in this study, the SIMIND MC code for modelling a Siemens Symbia T dual head SPECT/CT gamma camera fitted with high energy all-purpose collimators (HEAP) was validated. The validation was done by acquiring and simulating certain NEMA stipulated performance measurements for I-131. These measured and simulated performance data were processed and the results compared. The FWHM and FWTM for the gamma camera for measured and simulated system spatial resolution were 14.4, 29.7 mm versus 13.9, 25.9 mm. System sensitivity performance measurements and simulation produced results of 49 cps/MBq and 48 cps/MBq respectively. The corresponding measured and simulated values for the septal penetration were 3.8 and 4.2%. From this study it is evident that the SIMIND MC Code can be used with confidence to mimic I-131 studies acquired with a Siemens Symbia T SPECT/CT gamma camera fitted with HEAP collimators.
引用
收藏
页码:573 / 576
页数:4
相关论文
共 50 条
  • [31] Monte Carlo simulation of scintillation photons for the design of a high-resolution SPECT detector dedicated to human brain
    Yoshiyuki Hirano
    Tsutomu Zeniya
    Hidehiro Iida
    Annals of Nuclear Medicine, 2012, 26 : 214 - 221
  • [32] Evaluation of six scatter correction methods based on spectral analysis in Tc-99m SPECT imaging using SIMIND Monte Carlo simulation
    Asl, Mahsa Noori
    Sadremomtaz, Alireza
    Bitarafan-Rajabi, Ahmad
    JOURNAL OF MEDICAL PHYSICS, 2013, 38 (04) : 189 - 197
  • [33] Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT
    Payvand Taherparvar
    Alireza Sadremomtaz
    Australasian Physical & Engineering Sciences in Medicine, 2018, 41 : 31 - 39
  • [34] Development of GATE Monte Carlo simulation for a CsI pixelated gamma camera dedicated to high resolution animal SPECT
    Taherparvar, Payvand
    Sadremomtaz, Alireza
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2018, 41 (01) : 31 - 39
  • [35] Towards a comprehensive validation of a Monte Carlo simulation for clinical SPECT acquisition protocol optimisation and image correction
    Pells, S.
    Cullen, D. M.
    Robinson, A. P.
    Pietras, B.
    Calvert, N.
    Deidda, D.
    Fenwick, A.
    Ferreira, K.
    Hamilton, D.
    Heetun, W.
    Julyan, P.
    Needham, G.
    Page, E.
    Price, E.
    Tipping, J.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2020, 47 (SUPPL 1) : S418 - S419
  • [36] Monte Carlo simulation for In-111 imaging using a scintillation camera with a low-energy general purpose collimator
    Kojima, A.
    Uehara, S.
    Kawamura, S.
    Inoue, Y.
    Okada, K.
    Matsumoto, M.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2007, 34 : S345 - S345
  • [37] Comparison of photon counting and conventional scintillation detectors in a pinhole SPECT system for small animal imaging: Monte carlo simulation studies
    Young-Jin Lee
    Su-Jin Park
    Seung-Wan Lee
    Dae-Hong Kim
    Ye-Seul Kim
    Hee-Joung Kim
    Journal of the Korean Physical Society, 2013, 62 : 1317 - 1322
  • [38] Comparison of photon counting and conventional scintillation detectors in a pinhole SPECT system for small animal imaging: Monte carlo simulation studies
    Lee, Young-Jin
    Park, Su-Jin
    Lee, Seung-Wan
    Kim, Dae-Hong
    Kim, Ye-Seul
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 62 (09) : 1317 - 1322
  • [39] Initial assessment of the Monte-Carlo simulation of SPECT recording with the new region-centric CZT "DSPECT" camera
    Imbert, L.
    Marie, P. Y.
    Galbrun, E.
    Poussier, S.
    Wolf, D.
    Karcher, G.
    Noel, A.
    2012 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE RECORD (NSS/MIC), 2012, : 2543 - 2545
  • [40] Monte-Carlo Simulation and Experimental Validation of a Gamma-Camera Prototype Featuring a Novel Collimation Technique
    Meinderink, T.
    Osterloh, P.
    Osterloh, K.
    Goetz, T. I.
    Kuwert, T.
    Ritt, P.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2018, 45 : S18 - S18