Iterative denoising accelerated 3D SPACE FLAIR sequence for brain MR imaging at 3T

被引:11
|
作者
Eliezer, Michael [1 ,2 ]
Vaussy, Alexis [3 ]
Toupin, Solenn [3 ]
Barbe, Remy [1 ]
Kannengiesser, Stephan [4 ]
Stemmer, Alto [4 ]
Houdart, Emmanuel [1 ,2 ]
机构
[1] Lariboisiere Univ Hosp, Dept Neuroradiol, F-75010 Paris, France
[2] Univ Paris, Fac Med, F-75010 Paris, France
[3] Siemens Healthineers France, F-93210 St Denis, France
[4] Siemens Healthineers, D-91052 Erlangen, Germany
关键词
Magnetic resonance imaging (MRI); 3D FLAIR; Iterative denoising; Brain MRI; MULTIPLE-SCLEROSIS; 2D FLAIR; SUPPRESSION;
D O I
10.1016/j.diii.2021.09.004
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: The purpose of this study was to prospectively evaluate image quality of three-dimensional fluid attenuated inversion recovery (3D-FLAIR) sequence acquired with a high acceleration factor and reconstructed with iterative denoising (ID) for brain magnetic resonance imaging (MRI) at 3-T. Material and methods: Patients with brain tumor who underwent brain MRI were consecutively included. Two 3D-FLAIR sequences were successively performed for each patient. A first conventional FLAIR acquisition (conv-FLAIR) was performed with an acceleration factor of 6. The second acquisition was performed with an increased acceleration factor of 9. Two series one without ID (acc-FLAIR) and one with ID (acc-FLAIR-ID) were reconstructed. Two neuroradiologists independently assessed image quality, deep brain nuclei visualization and white matter/gray matter (WM/GM) differentiation on a 4-point scale. Results: Thirty patients with brain tumor were consecutively included in this study. There were 16 women and 14 men with a mean age of 54 +/- 17 (SD) years (range: 22-78 years). Scanning time of Acc-FLAIR-ID and Acc-FLAIR (4 min 40 sec) was 37% shorter than that of conv-FLAIR (2 min 50 sec) (P < 0.01). Improved image quality score was significantly different for both conv-FLAIR and acc-FLAIR-ID compared to acc-FLAIR (P < 0.01 for both). WM/GM differentiation score of conv-FLAIR was not significantly different compared to accFLAIR-ID (P = 0.10). Improved WM/GM differentiation score was different for both sequences compared to acc-FLAIR (P = 0.017 and P < 0.001). Deep brain nuclei visualization score was not different between convFLAIR and acc-FLAIR-ID (P = 0.71). However, the improved deep brain nuclei visualization score was signifi-cantly different for both sequences compared to acc-FLAIR (P < 0.001 for both). Conclusion: Scanning time of 3D-FLAIR sequence using a high acceleration factor reconstructed with ID algorithm can be reduced by 37% while preserving image quality for brain MRI. (c) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:13 / 20
页数:8
相关论文
共 50 条
  • [31] Accelerated time-resolved 3D contrast-enhanced MR angiography at 3T: clinical experience in 31 patients
    Alex Frydrychowicz
    Thorsten A. Bley
    Jan T. Winterer
    Andreas Harloff
    Mathias Langer
    Jürgen Hennig
    Michael Markl
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2006, 19 : 187 - 195
  • [32] In vivo sodium MR imaging of the abdomen at 3T
    James, Judy R.
    Panda, Anshuman
    Lin, Chen
    Dydak, Ulrike
    Dale, Brian M.
    Bansal, Navin
    ABDOMINAL IMAGING, 2015, 40 (07): : 2272 - 2280
  • [33] "Value of 3D STIR Gated MRI Sequence in Cardiac and Chest Imaging with 3T MRI": Initial Experiences
    Kandasamy, Geetha
    Peter, S. Babu
    Priya, R. Bharathi
    JOURNAL OF CLINICAL AND DIAGNOSTIC RESEARCH, 2019, 13 (01) : TC6 - TC9
  • [34] 3T MR imaging: The pros and cons - Reply
    Bradley, WG
    Safar, FG
    Hurtado, C
    Ord, J
    Alksne, JF
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2004, 25 (09) : 1455 - 1455
  • [35] Comparison of 1.5-T 3D SPACE MR Imaging of the Knee vs Conventional MR Imaging of the Knee
    Magee, T.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2012, 198 (05)
  • [36] In vivo sodium MR imaging of the abdomen at 3T
    Judy R. James
    Anshuman Panda
    Chen Lin
    Ulrike Dydak
    Brian M. Dale
    Navin Bansal
    Abdominal Imaging, 2015, 40 : 2272 - 2280
  • [37] MR safety and imaging of Neuroform stents at 3T
    Nehra, A
    Moran, CJ
    Cross, DT
    Derdeyn, CP
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2004, 25 (09) : 1476 - 1478
  • [38] 3D MR thermometry of frozen tissue: Feasibility and accuracy during cryoablation at 3T
    Overduin, Christiaan G.
    Futterer, Jurgen J.
    Scheenen, Tom W. J.
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2016, 44 (06) : 1572 - 1579
  • [39] Optimisation of the 3D MDEFT sequence for anatomical brain imaging: Technical implications at 1.5 and 3 T
    Deichmann, R
    Schwarzbauer, C
    Turner, R
    NEUROIMAGE, 2004, 21 (02) : 757 - 767
  • [40] Improving Detection of Multiple Sclerosis Lesions in the Posterior Fossa Using an Optimized 3D-FLAIR Sequence at 3T
    Lecler, A.
    El Sanharawi, I.
    El Methni, J.
    Gout, O.
    Koskas, P.
    Savatovsky, J.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2019, 40 (07) : 1170 - 1176