3D MR thermometry of frozen tissue: Feasibility and accuracy during cryoablation at 3T

被引:10
|
作者
Overduin, Christiaan G. [1 ]
Futterer, Jurgen J. [1 ,2 ]
Scheenen, Tom W. J. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Radiol & Nucl Med, POB 9101 767, NL-6500 HB Nijmegen, Netherlands
[2] Univ Twente, MIRA Inst Biomed Engn & Tech Med, Enschede, Netherlands
关键词
UTE; MR thermometry; frozen tissue; cryoablation; GUIDED PERCUTANEOUS CRYOTHERAPY; PULSED NMR; INITIAL-EXPERIENCE; SIGNAL INTENSITY; RELAXATION-TIMES; STRIATED-MUSCLE; WATER; CRYOSURGERY; TUMOR; SEQUENCES;
D O I
10.1002/jmri.25301
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo assess the feasibility and accuracy of 3D ultrashort echo time (UTE) magnetic resonance (MR) thermometry of frozen tissue during cryoablation on a clinical 3T MR system. Materials and MethodsEx vivo porcine muscle specimens (n=4) were imaged during two cycles of 10:3 minutes freeze-thaw on a 3T clinical MR scanner. Continuous MR monitoring was performed using a 3D radial ramp-sampled UTE sequence with a shortest TE of 70 s. Fiber optic sensors were used for temperature reference. Data of three experiments were used as reference sets. Signal intensity values were normalized to baseline before cooling and related to temperature. Data for subzero temperatures were fit to a monoexponential function. In the separate validation set, the obtained fit was used to generate 3D MR temperature maps of frozen tissue at each imaging timepoint. Statistical analysis was performed to assess accuracy of the MR temperature maps. ResultsWith 3D UTE imaging, MR signal was measured from frozen tissue down to temperatures as low as -40 degrees C. Temperatures predicted from the MR temperature maps strongly correlated with sensor recorded values (r=0.977, P < 0.001). Bland-Altman analysis demonstrated a mean difference between MR-estimated temperatures and sensor readings of -1.22.7 degrees C with upper and lower limits of agreement of +4.1 and -6.5 degrees C, respectively. Conclusion3D MR thermometry of frozen tissue using UTE signal intensity was feasible during cryoablation on a clinical 3T MR system. Down to temperatures as low as -40 degrees C, accuracy of the MR temperature maps was within clinically acceptable limits. J. Magn. Reson. Imaging 2016;44:1572-1579.
引用
收藏
页码:1572 / 1579
页数:8
相关论文
共 50 条
  • [1] Towards optimized MR thermometry of the human heart at 3T
    Hey, S.
    Cernicanu, A.
    de Senneville, B. D.
    Roujol, S.
    Ries, M.
    Jais, P.
    Moonen, C. T. W.
    Quesson, B.
    NMR IN BIOMEDICINE, 2012, 25 (01) : 35 - 43
  • [2] Feasibility of 3D T1W sequences in contrast-material enhanced MR cisternography at 3T
    Algin, Oktay
    Koc, Ural
    Ayberk, Giyas
    TURKISH JOURNAL OF MEDICAL SCIENCES, 2022, 52 (06) : 1943 - 1949
  • [3] 3D sensitivity encoded ellipsoidal MR spectroscopic imaging of gliomas at 3T
    Ozturk-Isik, Esin
    Chen, Albert. P.
    Crane, Jason C.
    Bian, Wei
    Xu, Duan
    Han, Eric T.
    Chang, Susan M.
    Vigneron, Daniel B.
    Nelson, Sarah J.
    MAGNETIC RESONANCE IMAGING, 2009, 27 (09) : 1249 - 1257
  • [4] Development and Evaluation of a Novel 3D Spiral Staircase Ultrashort-Echo-Time (SSCUTE) Sequence for Thermometry of Frozen Tissue during MRI-Guided Cryoablation
    Browne, J. E.
    Krishnamoorthy, G.
    Lu, A.
    Woodrum, D. A.
    Pipe, J.
    MEDICAL PHYSICS, 2024, 51 (10) : 7761 - 7761
  • [5] Intra-operative visualisation of 3D temperature maps and 3D navigation during tissue cryoablation
    Samset, E
    Mala, T
    Aurdal, L
    Balasingham, I
    COMPUTERIZED MEDICAL IMAGING AND GRAPHICS, 2005, 29 (06) : 499 - 505
  • [6] Iterative Denoising Accelerated 3D FLAIR Sequence for Hydrops MR Imaging at 3T
    Quint, R.
    Vaussy, A.
    Stemmer, A.
    Hautefort, C.
    Houdart, E.
    Eliezer, M.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2023, 44 (09) : 1064 - 1069
  • [7] Iterative denoising accelerated 3D SPACE FLAIR sequence for brain MR imaging at 3T
    Eliezer, Michael
    Vaussy, Alexis
    Toupin, Solenn
    Barbe, Remy
    Kannengiesser, Stephan
    Stemmer, Alto
    Houdart, Emmanuel
    DIAGNOSTIC AND INTERVENTIONAL IMAGING, 2022, 103 (01) : 13 - 20
  • [8] The Holy Grail in diagnostic neuroradiology: 3T or 3D?
    Frederik Barkhof
    Petra J. W. Pouwels
    Mike P. Wattjes
    European Radiology, 2011, 21 : 449 - 456
  • [9] The Holy Grail in diagnostic neuroradiology: 3T or 3D?
    Barkhof, Frederik
    Pouwels, Petra J. W.
    Wattjes, Mike P.
    EUROPEAN RADIOLOGY, 2011, 21 (03) : 449 - 456
  • [10] Optimized Breath-Hold Compressed-Sensing 3D MR Cholangiopancreatography at 3T: Image Quality Analysis and Clinical Feasibility Assessment
    Song, Ji Soo
    Kim, Seung Hun
    Kuehn, Bernd
    Paek, Mun Young
    DIAGNOSTICS, 2020, 10 (06)