3D MR thermometry of frozen tissue: Feasibility and accuracy during cryoablation at 3T

被引:10
|
作者
Overduin, Christiaan G. [1 ]
Futterer, Jurgen J. [1 ,2 ]
Scheenen, Tom W. J. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Dept Radiol & Nucl Med, POB 9101 767, NL-6500 HB Nijmegen, Netherlands
[2] Univ Twente, MIRA Inst Biomed Engn & Tech Med, Enschede, Netherlands
关键词
UTE; MR thermometry; frozen tissue; cryoablation; GUIDED PERCUTANEOUS CRYOTHERAPY; PULSED NMR; INITIAL-EXPERIENCE; SIGNAL INTENSITY; RELAXATION-TIMES; STRIATED-MUSCLE; WATER; CRYOSURGERY; TUMOR; SEQUENCES;
D O I
10.1002/jmri.25301
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
PurposeTo assess the feasibility and accuracy of 3D ultrashort echo time (UTE) magnetic resonance (MR) thermometry of frozen tissue during cryoablation on a clinical 3T MR system. Materials and MethodsEx vivo porcine muscle specimens (n=4) were imaged during two cycles of 10:3 minutes freeze-thaw on a 3T clinical MR scanner. Continuous MR monitoring was performed using a 3D radial ramp-sampled UTE sequence with a shortest TE of 70 s. Fiber optic sensors were used for temperature reference. Data of three experiments were used as reference sets. Signal intensity values were normalized to baseline before cooling and related to temperature. Data for subzero temperatures were fit to a monoexponential function. In the separate validation set, the obtained fit was used to generate 3D MR temperature maps of frozen tissue at each imaging timepoint. Statistical analysis was performed to assess accuracy of the MR temperature maps. ResultsWith 3D UTE imaging, MR signal was measured from frozen tissue down to temperatures as low as -40 degrees C. Temperatures predicted from the MR temperature maps strongly correlated with sensor recorded values (r=0.977, P < 0.001). Bland-Altman analysis demonstrated a mean difference between MR-estimated temperatures and sensor readings of -1.22.7 degrees C with upper and lower limits of agreement of +4.1 and -6.5 degrees C, respectively. Conclusion3D MR thermometry of frozen tissue using UTE signal intensity was feasible during cryoablation on a clinical 3T MR system. Down to temperatures as low as -40 degrees C, accuracy of the MR temperature maps was within clinically acceptable limits. J. Magn. Reson. Imaging 2016;44:1572-1579.
引用
收藏
页码:1572 / 1579
页数:8
相关论文
共 50 条
  • [21] Image Analysis in Time-Resolved Large Field of View 3D MR-Angiography at 3T
    Frydrychowicz, Alex
    Bley, Thorsten A.
    Zadeh, Zahra Abdullah
    Harloff, Andreas
    Winterer, Jan T.
    Hennig, Juergen
    Langer, Mathias
    Markl, Michael
    JOURNAL OF MAGNETIC RESONANCE IMAGING, 2008, 28 (05) : 1116 - 1124
  • [22] Carotid Artery Stenosis: Comparison of 3D Time-of-Flight MR Angiography and Contrast-Enhanced MR Angiography at 3T
    Platzek, Ivan
    Sieron, Dominik
    Wiggermann, Philipp
    Laniado, Michael
    RADIOLOGY RESEARCH AND PRACTICE, 2014, 2014
  • [23] 3D-Video-fMRI: 3D Motion Tracking in a 3T MRI Environment
    Fernandes, Jose Maria
    Tafula, Sergio
    Silva Cunha, Joao Paulo
    IMAGE ANALYSIS AND RECOGNITION: 8TH INTERNATIONAL CONFERENCE, ICIAR 2011, PT II: 8TH INTERNATIONAL CONFERENCE, ICIAR 2011, 2011, 6754 : 59 - 67
  • [24] 3T MR findings in vascular parkinsonism
    Lim, S. H.
    Ng, K. M.
    Tan, H. E.
    Rumpel, H.
    Yuen, Y.
    Tan, E. K.
    Chan, L. L.
    JOURNAL OF THE NEUROLOGICAL SCIENCES, 2009, 285 : S295 - S295
  • [25] Frequency drift in MR spectroscopy at 3T
    Hui, Steve C. N.
    Mikkelsen, Mark
    Zollner, Helge J.
    Ahluwalia, Vishwadeep
    Alcauter, Sarael
    Baltusis, Laima
    Barany, Deborah A.
    Barlow, Laura R.
    Becker, Robert
    Berman, Jeffrey, I
    Berrington, Adam
    Bhattacharyya, Pallab K.
    Blicher, Jakob Udby
    Bogner, Wolfgang
    Brown, Mark S.
    Calhoun, Vince D.
    Castillo, Ryan
    Cecil, Kim M.
    Choi, Yeo Bi
    Chu, Winnie C. W.
    Clarke, William T.
    Craven, Alexander R.
    Cuypers, Koen
    Dacko, Michael
    de la Fuente-Sandoval, Camilo
    Desmond, Patricia
    Domagalik, Aleksandra
    Dumont, Julien
    Duncan, Niall W.
    Dydak, Ulrike
    Dyke, Katherine
    Edmondson, David A.
    Ende, Gabriele
    Ersland, Lars
    Evans, C. John
    Fermin, Alan S. R.
    Ferretti, Antonio
    Fillmer, Ariane
    Gong, Tao
    Greenhouse, Ian
    Grist, James T.
    Gu, Meng
    Harris, Ashley D.
    Hatz, Katarzyna
    Heba, Stefanie
    Heckova, Eva
    Hegarty, John P.
    Heise, Kirstin-Friederike
    Honda, Shiori
    Jacobson, Aaron
    NEUROIMAGE, 2021, 241
  • [26] A Phantom Validation Study of a 3D Background Phase Model for MR Thermometry
    Yung, J.
    Fuentes, D.
    Hazle, J.
    Stafford, R.
    MEDICAL PHYSICS, 2012, 39 (06) : 3664 - 3664
  • [27] High resolution 3D diffusion MRI of carotid plaque on 3T
    Yibin Xie
    Wei Yu
    Zhaoyang Fan
    Christopher T Nguyen
    Jing An
    Zhaoqi Zhang
    Debiao Li
    Journal of Cardiovascular Magnetic Resonance, 16 (Suppl 1)
  • [28] Accelerated Pseudo 3D Dynamic Speech MR Imaging at 3T Using Unsupervised Deep Variational Manifold Learning
    Rusho, Rushdi Zahid
    Zou, Qing
    Alam, Wahidul
    Erattakulangara, Subin
    Jacob, Mathews
    Lingala, Sajan Goud
    MEDICAL IMAGE COMPUTING AND COMPUTER ASSISTED INTERVENTION, MICCAI 2022, PT VI, 2022, 13436 : 697 - 706
  • [29] Evaluation of optimised 3D turbo spin echo and gradient echo MR pulse sequences of the knee at 3T and 1.5T
    Abdulaal, O. M.
    Rainford, L.
    MacMahon, P. J.
    Kenny, P.
    Carty, F.
    Galligan, M.
    Cradock, A.
    Alhazmi, F. H.
    McGee, A.
    RADIOGRAPHY, 2021, 27 (02) : 389 - 397
  • [30] 3D High-Spatial-Resolution Cerebral MR Venography at 3T: A Contrast-Dose-Reduction Study
    Tomasian, A.
    Salamon, N.
    Krishnam, M. S.
    Finn, J. P.
    Villablanca, J. P.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2009, 30 (02) : 349 - 355