Analytical properties of the GEM Premier 3000 analyzer evaluated

被引:0
|
作者
Vukelic, Nada [1 ]
Futac, Daljia Papic [1 ]
Topic, Elizabeta [1 ]
机构
[1] Univ Zagreb, Sestre Milosrdnice Univ Hosp, Univ Dept Chem, HR-10000 Zagreb, Croatia
关键词
point of care testing; acid-base balance parameters; electrolytes; metabolites;
D O I
暂无
中图分类号
R446 [实验室诊断]; R-33 [实验医学、医学实验];
学科分类号
1001 ;
摘要
Introduction: The development of sensor technology has led to the advent of analyzers for fast and simple determination of analyte concentration in whole blood samples that can be performed in a laboratory and as point of care testing (POCT). POCT analyzers are generally put in at intensive care units to enable rapid test results and earliest possible therapy introduction. Aim: To evaluate analytical properties of the GEM Premier 3000 analyzer and compatibility of the results obtained by use of this analyzer with those reported from analyzers used in central laboratory before the GEM Premier 3000 POCT analyzer fitting and placing in operation at Intensive Care Unit, University Department of Surgery. Material and methods: Analytical properties of the analyzer were evaluated by determination of within-run imprecision, between-run imprecision, inaccuracy, and comparative determination of analyte concentrations in patient samples on the study analyzer and the analyzers used for the respective analytes at central laboratory as reference analyzers. The study was performed in accordance with ECCLS recommendations. Results: Satisfactory results were recorded on within-run imprecision for all study analytes (CV:<= 3.46%), and on between-run imprecision for all analytes (CV <= 4.65%) except for lactate (Level 2, CV = 9.35%; Level 3, CV =5.55%). Inaccuracy evaluation yielded satisfactory results for all study analytes (R <= 4.53%) except for lactate (Level 1, R = 13.59%; Level 2, R = -11.00%; Level 3, R = 10.00%). Correlation coefficients were within range 0.9727-0.9920, except for sodium (r = 0.9207) and showed high correlation of all analyte concentrations determined in patient samples on the study analyzerand reference analyzers. Passing Bablok regression analysis also indicated a very high compatibility of analyte concentrations measured in patient samples by use of the study analyzer and reference analyzers. Conclusion: Results of analyte concentration determination in patient samples obtained by use of the study analyzer were compatible with those produced by reference analyzers at central laboratory. The GEM Premier 3000 analyzer has proved to be simple to operate, while automated and independent quality control performed after each individual sample and automated error elimination ensures safe work and reliable test results.d
引用
收藏
页码:231 / 241
页数:11
相关论文
共 50 条
  • [21] GEM Premier 3000血气分析仪常见操作问题分析及维护
    兰莫莉
    钟晓燕
    雷雪娜
    医疗装备, 2014, 27 (04) : 139 - 140
  • [22] GEM Premier 3000血气分析仪试剂包条形码解析
    俞晓青
    医疗卫生装备, 2013, 34 (01) : 105 - 106
  • [23] Analytical evaluation of direct bicarbonate measurement with the new gem premier chemstat in hemodialysis patients
    Balboni, Fiamma
    Terreni, Alessandro
    Gallo, Marco
    Guzzi, Francesco
    Caparrini, Chiara
    Burbui, Stefania
    Vezzosi, Marco
    Galora, Silvia
    Lori, Giulia
    Lippi, Giuseppe
    SCANDINAVIAN JOURNAL OF CLINICAL & LABORATORY INVESTIGATION, 2021, 81 (05): : 418 - 421
  • [24] Does mixing improve the comparability between hemoglobin measured on the GEM Premier 4000 blood gas analyzer and the Sysmex XN hematology analyzer?
    Ismail, Ola
    Balion, Cynthia
    Hill, Stephen
    Martin, Janet
    CLINICAL BIOCHEMISTRY, 2019, 74 : 97 - 97
  • [25] Gem premier 3000血气分析仪常见故障及解决方法
    赖东珍
    中国医学装备, 2006, (10) : 37 - 39
  • [26] GEM premier3000血气分析仪使用中的注意事项
    丁华
    刘和录
    何玥
    徐志康
    陈伟光
    陈望
    梁鸿
    医疗卫生装备, 2011, 32 (11) : 110 - 111
  • [27] Novel generations of laboratory instruments should not worsen analytical quality: The case of GEM Premier 5000
    Aloisio, Elena
    Carnevale, Assunta
    Dolci, Alberto
    Panteghini, Mauro
    CLINICAL BIOCHEMISTRY, 2018, 58 : 128 - 130
  • [28] Gem premier 3000血气分析试剂包故障分析及自行解决方法
    赖东珍
    中国医疗设备, 2013, 28 (04) : 144 - 147
  • [29] GEM Premier 3000血气分析仪非传统检测项目实用价值的探讨
    余建洪
    检验医学, 2016, 31 (01) : 26 - 30
  • [30] New Chemistry Cartridge for the GEM Premier 4000 Critical Care Analyzer, Including Electrochemical Sensors for Creatinine, Urea, and Measured Bicarbonate
    Cervera, Jose
    Conant, Joann
    POINT OF CARE, 2010, 9 (01): : 1 - 7