Coefficient-Free Blood Pressure Estimation Based on Pulse Transit Time-Cuff Pressure Dependence

被引:60
|
作者
Forouzanfar, Mohamad [1 ]
Ahmad, Saif [1 ]
Batkin, Izmail [1 ,2 ]
Dajani, Hilmi R. [1 ]
Groza, Voicu Z. [1 ]
Bolic, Miodrag [1 ]
机构
[1] Univ Ottawa, Sch Elect Engn & Comp Sci, Ottawa, ON K1N 6N5, Canada
[2] Ottawa Gen Hosp, Ottawa, ON K1H 8LC, Canada
基金
加拿大自然科学与工程研究理事会;
关键词
Blood pressure (BP); estimation; mathematical model; oscillometry; pulse transit time (PTT); OSCILLOMETRIC MAXIMUM; HYPERTENSION; VALIDATION; DEVICE;
D O I
10.1109/TBME.2013.2243148
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Oscillometry is a popular technique for automatic estimation of blood pressure (BP). However, most of the oscillometric algorithms rely on empirical coefficients for systolic and diastolic pressure evaluation that may differ in various patient populations, rendering the technique unreliable. A promising complementary technique for automatic estimation of BP, based on the dependence of pulse transit time (PTT) on cuff pressure (CP) (PTT-CP mapping), has been proposed in the literature. However, a theoretical grounding for this technique and a nonparametric BP estimation approach are still missing. In this paper, we propose a novel coefficient-free BP estimation method based on PTT-CP dependence. PTT is mathematically modeled as a function of arterial lumen area under the cuff. It is then analytically shown that PTT-CP mappings computed from various points on the arterial pulses can be used to directly estimate systolic, diastolic, and mean arterial pressure without empirical coefficients. Analytical results are cross-validated with a pilot investigation on ten healthy subjects where 150 simultaneous electrocardiogram and oscillometric BP recordings are analyzed. The results are encouraging whereby the mean absolute errors of the proposed method in estimating systolic and diastolic pressures are 5.31 and 4.51 mmHg, respectively, relative to the Food and Drug Administration approved Omron monitor. Our work thus shows promise toward providing robust and objective BP estimation in a variety of patients and monitoring situations.
引用
收藏
页码:1814 / 1824
页数:11
相关论文
共 50 条
  • [21] Near threshold pulse transit time processor for central blood pressure estimation
    Veirano, Francisco
    Perez-Nicoli, Pablo
    Gammarano, Nicolas
    Fierro, German
    Silveira, Fernando
    2022 IEEE 13TH LATIN AMERICAN SYMPOSIUM ON CIRCUITS AND SYSTEMS (LASCAS), 2022, : 172 - 175
  • [22] Characters available in photoplethysmogram for blood pressure estimation: beyond the pulse transit time
    Yanjun Li
    Zengli Wang
    Lin Zhang
    Xianglin Yang
    Jinzhong Song
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 367 - 376
  • [23] Characters available in photoplethysmogram for blood pressure estimation: beyond the pulse transit time
    Li, Yanjun
    Wang, Zengli
    Zhang, Lin
    Yang, Xianglin
    Song, Jinzhong
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (02) : 367 - 376
  • [24] Cuff-less blood pressure measurement with pulse transit time: The importance of rigorous assessment
    van Helmond, Noud
    Plante, Timothy B.
    JOURNAL OF CLINICAL HYPERTENSION, 2021, 23 (01): : 71 - 72
  • [25] Continuous blood pressure measurement based on pulse transit time: a comparison to ambulatory blood pressure measurement
    Kuechler, G.
    Gehring, J.
    Wesely, M.
    Thimel, K.
    Haberl, R.
    EUROPEAN HEART JOURNAL, 2016, 37 : 1267 - 1267
  • [26] An implantable optical blood pressure sensor based on pulse transit time
    Fiala, Jens
    Bingger, Philipp
    Ruh, Dominic
    Foerster, Katharina
    Heilmann, Claudia
    Beyersdorf, Friedhelm
    Zappe, Hans
    Seifert, Andreas
    BIOMEDICAL MICRODEVICES, 2013, 15 (01) : 73 - 81
  • [27] An implantable optical blood pressure sensor based on pulse transit time
    Jens Fiala
    Philipp Bingger
    Dominic Ruh
    Katharina Foerster
    Claudia Heilmann
    Friedhelm Beyersdorf
    Hans Zappe
    Andreas Seifert
    Biomedical Microdevices, 2013, 15 : 73 - 81
  • [28] Perturbationless Calibration of Pulse Transit Time to Blood Pressure
    Gao, Mingwu
    Mukkamala, Ramakrishna
    2012 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2012, : 232 - 235
  • [29] Limitation of Pulse Transit Time to Estimate Blood Pressure
    Sklarsky, Amanda
    Garvin, Nathan M.
    Pawelczyk, James A.
    FASEB JOURNAL, 2019, 33
  • [30] Cuff-less and continuous blood pressure measurement based on pulse transit time from carotid and toe photoplethysmograms
    Zuhair Sameen A.
    Jaafar R.
    Zahedi E.
    Kok Beng G.
    Journal of Medical Engineering and Technology, 2022, 46 (07): : 567 - 589