Empirical dead-time corrections for synchrotron sources

被引:19
|
作者
Walko, D. A. [1 ]
Arms, D. A. [1 ]
Landahl, E. C. [1 ]
机构
[1] Argonne Natl Lab, Adv Photon Source, Argonne, IL 60439 USA
关键词
dead-time; photon detector; synchrotron fill pattern; data correction;
D O I
10.1107/S0909049508022358
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An experimental comparison of models for performing dead-time corrections of photon-counting detectors at synchrotron sources is presented. The performance of several detectors in the three operating modes of the Advanced Photon Source is systematically compared, with particular emphasis on asymmetric fill patterns. Several simple and well known correction formulas are evaluated. The results demonstrate the critical importance of detector speed and synchrotron fill pattern in selecting the proper dead-time correction.
引用
收藏
页码:612 / 617