Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate 192Ir brachytherapy sources

被引:13
|
作者
Kaveckyte, Vaiva [1 ]
Malusek, Alexandr [1 ]
Benmakhlouf, Hamza [2 ]
Carlsson, Gudrun Alm [1 ]
Tedgren, Asa Carlsson [1 ,2 ]
机构
[1] Linkoping Univ, Dept Med & Hlth Sci, Radiat Phys, SE-58185 Linkoping, Sweden
[2] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
brachytherapy; dosimetry; Iridium-192; microDiamond detector; radiation detectors; CRYSTAL DIAMOND DETECTOR; PHOTON ENERGIES RELEVANT; HDR BRACHYTHERAPY; RATE DEPENDENCE; THERMOLUMINESCENT DOSIMETERS; ELECTRON-BEAMS; RADIATION; MOSFET; PROTOCOL; SPECTRA;
D O I
10.1002/mp.12694
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Experimental dosimetry of high-dose-rate (HDR) Ir-192 brachytherapy (BT) sources is complicated due to high dose and dose-rate gradients, and softening of photon energy spectrum with depth. A single crystal synthetic diamond detector microDiamond (PTW 60019, Freiburg, Germany) has a small active volume, high sensitivity, direct readout, and nearly water-equivalent active volume. The purpose of this study was to evaluate the suitability of microDiamond detectors for the determination of absorbed dose to water around HDR Ir-192 BT sources. Three microDiamond detectors were used, allowing for the comparison of their properties. Methods: In-phantom measurements were performed using microSelectron and VariSource iX HDR Ir-192 BT treatment units. Their treatment planning systems (TPSs), Oncentra (v. 4.3) and BrachyVision (v. 13.6), respectively, were used to create irradiation plans for a cubic PMMA phantom with the microDiamond positioned at one of three source-to-detector distances (SDDs) (1.5, 2.5, and 5.5 cm) at a time. The source was stepped in increments of 0.5 cm over a total length of 6 cm to yield absorbed dose of 2 Gy at the nominal reference-point of the detector. Detectors were calibrated in Co-60 beam in terms of absorbed dose to water, and Monte Carlo (MC) calculated beam quality correction factors were applied to account for absorbed-dose energy dependence. Phantom correction factors were applied to account for differences in dimensions between the measurement phantom and a water phantom used for absorbed dose calculations made with a TPS. The same measurements were made with all three of the detectors. Additionally, dose-rate dependence and stability of the detectors were evaluated in Co-60 beam. Results: The percentage differences between experimentally determined and TPS-calculated absorbed doses to water were from -1.3% to +2.9%. The values agreed to within experimental uncertainties, which were from 1.9% to 4.3% (k = 2) depending on the detector, SDD and treatment delivery unit. No dose-rate or intrinsic energy dependence corrections were applied. All microDiamonds were comparable in terms of preirradiation dose, stability of the readings and energy response, and showed a good agreement. Conclusions: The results indicate that the microDiamond is potentially suitable for the determination of absorbed dose to water around HDR Ir-192 BT sources and may be used for independent verification of TPS's calculations, as well as for QA measurements of HDR Ir-192 BT treatment delivery units at clinical sites. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:429 / 437
页数:9
相关论文
共 50 条
  • [1] A microDiamondfor determination of absorbed dose around high-dose-rate 192Ir brachytherapy sources
    Kaveckyte, V.
    Malusek, A.
    Benmaklouf, H.
    Carlsson, G. Alm
    Tedgren, A. Carlsson
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S407 - S407
  • [2] Monte Carlo and experimental high dose rate 192Ir brachytherapy dosimetry with microDiamond detectors
    Rossi, Giulio
    Gainey, Mark
    Thomann, Benedikt
    Kollefrath, Michael
    Wuerfel, Jan
    Allgaier, Bernd
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2019, 29 (03): : 272 - 281
  • [3] Determination of the dose rate around a HDR 192Ir brachytherapy source with the microDiamond and the microSilicon detector
    Rossi, Giulio
    Failing, Thomas
    Gainey, Mark
    Kollefrath, Michael
    Hensley, Frank
    Zink, Klemens
    Baltas, Dimos
    ZEITSCHRIFT FUR MEDIZINISCHE PHYSIK, 2023, 33 (04): : 463 - 478
  • [4] Comparison of calibration procedures for 192IR high-dose-rate brachytherapy sources
    Baltas, D
    Geramani, K
    Ioannidis, GT
    Hierholz, K
    Rogge, B
    Kolotas, C
    Müller-Sievers, K
    Milickovic, N
    Kober, B
    Zamboglou, N
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1999, 43 (03): : 653 - 661
  • [5] Characterization of microMOSFET detectors for in vivo dosimetry in high-dose-rate brachytherapy with 192Ir
    Ruiz-Arrebola, Samuel
    Fabregat-Borras, Rosa
    Rodriguez, Eduardo
    Fernandez-Montes, Manuel
    Perez-Macho, Mercedes
    Ferri, Maria
    Garcia, Ana
    Cardenal, Juan
    Pacheco, Maria T.
    Anchuelo, Javier
    Tornero-Lopez, Ana M.
    Prada, Pedro J.
    Guirado, Damian
    MEDICAL PHYSICS, 2020, 47 (05) : 2242 - 2253
  • [6] A novel absorbed dose standard for the calibration of 192Ir sources used in high dose rate brachytherapy
    D'Arienzo, M.
    Guerra, A. S.
    Loreti, S.
    Caporali, C.
    Pimpinella, M.
    Bolzan, C.
    RADIOTHERAPY AND ONCOLOGY, 2014, 111 : S146 - S147
  • [7] High-Dose-Rate 192Ir Brachytherapy Dose Verification: A Phantom Study
    Nikoofar, Alireza
    Hoseinpour, Zohreh
    Mahdavi, Seied Rabi
    Hasanzadeh, Hadi
    Tavirani, Mostafa Rezaei
    IRANIAN JOURNAL OF CANCER PREVENTION, 2015, 8 (03)
  • [8] Suitability of the microDiamond detector for experimental determination of the anisotropy function of High Dose Rate192Ir brachytherapy sources
    Rossi, Giulio
    Gainey, Mark
    Kollefrath, Michael
    Hofmann, Elena
    Baltas, Dimos
    MEDICAL PHYSICS, 2020, 47 (11) : 5838 - 5851
  • [9] Quality control of 192Ir High Dose Rate Brachytherapy sources
    Pereira, A
    Pirraco, R
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S136 - S137
  • [10] Dosimetric characteristics of the 192Ir high-dose-rate afterloading brachytherapy source
    Sadeghi, Mahdi
    Taghdiri, Fatemah
    Saidi, Pooneh
    JAPANESE JOURNAL OF RADIOLOGY, 2011, 29 (05) : 324 - 329