Suitability of microDiamond detectors for the determination of absorbed dose to water around high-dose-rate 192Ir brachytherapy sources

被引:13
|
作者
Kaveckyte, Vaiva [1 ]
Malusek, Alexandr [1 ]
Benmakhlouf, Hamza [2 ]
Carlsson, Gudrun Alm [1 ]
Tedgren, Asa Carlsson [1 ,2 ]
机构
[1] Linkoping Univ, Dept Med & Hlth Sci, Radiat Phys, SE-58185 Linkoping, Sweden
[2] Karolinska Univ Hosp, Dept Med Radiat Phys & Nucl Med, SE-17176 Stockholm, Sweden
关键词
brachytherapy; dosimetry; Iridium-192; microDiamond detector; radiation detectors; CRYSTAL DIAMOND DETECTOR; PHOTON ENERGIES RELEVANT; HDR BRACHYTHERAPY; RATE DEPENDENCE; THERMOLUMINESCENT DOSIMETERS; ELECTRON-BEAMS; RADIATION; MOSFET; PROTOCOL; SPECTRA;
D O I
10.1002/mp.12694
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Experimental dosimetry of high-dose-rate (HDR) Ir-192 brachytherapy (BT) sources is complicated due to high dose and dose-rate gradients, and softening of photon energy spectrum with depth. A single crystal synthetic diamond detector microDiamond (PTW 60019, Freiburg, Germany) has a small active volume, high sensitivity, direct readout, and nearly water-equivalent active volume. The purpose of this study was to evaluate the suitability of microDiamond detectors for the determination of absorbed dose to water around HDR Ir-192 BT sources. Three microDiamond detectors were used, allowing for the comparison of their properties. Methods: In-phantom measurements were performed using microSelectron and VariSource iX HDR Ir-192 BT treatment units. Their treatment planning systems (TPSs), Oncentra (v. 4.3) and BrachyVision (v. 13.6), respectively, were used to create irradiation plans for a cubic PMMA phantom with the microDiamond positioned at one of three source-to-detector distances (SDDs) (1.5, 2.5, and 5.5 cm) at a time. The source was stepped in increments of 0.5 cm over a total length of 6 cm to yield absorbed dose of 2 Gy at the nominal reference-point of the detector. Detectors were calibrated in Co-60 beam in terms of absorbed dose to water, and Monte Carlo (MC) calculated beam quality correction factors were applied to account for absorbed-dose energy dependence. Phantom correction factors were applied to account for differences in dimensions between the measurement phantom and a water phantom used for absorbed dose calculations made with a TPS. The same measurements were made with all three of the detectors. Additionally, dose-rate dependence and stability of the detectors were evaluated in Co-60 beam. Results: The percentage differences between experimentally determined and TPS-calculated absorbed doses to water were from -1.3% to +2.9%. The values agreed to within experimental uncertainties, which were from 1.9% to 4.3% (k = 2) depending on the detector, SDD and treatment delivery unit. No dose-rate or intrinsic energy dependence corrections were applied. All microDiamonds were comparable in terms of preirradiation dose, stability of the readings and energy response, and showed a good agreement. Conclusions: The results indicate that the microDiamond is potentially suitable for the determination of absorbed dose to water around HDR Ir-192 BT sources and may be used for independent verification of TPS's calculations, as well as for QA measurements of HDR Ir-192 BT treatment delivery units at clinical sites. (C) 2017 American Association of Physicists in Medicine
引用
收藏
页码:429 / 437
页数:9
相关论文
共 50 条
  • [21] Clinical application of 192Ir High-Dose-Rate brachytherapy in metastatic lymph nodes of the neck
    Hongling Lu
    Yunchuan Sun
    Yan Gao
    Li Xiao
    Jianxi Zhou
    Xiaoming Yin
    Wei Guo
    Kui Fan
    Discover Oncology, 14
  • [22] Calibration of new high dose rate 192Ir sources
    Stump, KE
    DeWerd, LA
    Micka, JA
    Anderson, DR
    MEDICAL PHYSICS, 2002, 29 (07) : 1483 - 1488
  • [23] Calibration of 192Ir High Dose Rate Brachytherapy Source Using Water Phantom
    Yu, L.
    Yang, B.
    Liu, X.
    Pang, T.
    Liu, N.
    Li, B.
    Dong, T.
    Wang, B.
    Wang, Z.
    Qiu, J.
    MEDICAL PHYSICS, 2018, 45 (06) : E212 - E212
  • [24] High dose rate 192Ir afterloading brachytherapy for cancer of the vagina
    Kushner, DM
    Fleming, PA
    Kennedy, AW
    Wilkinson, DA
    Lee, E
    Saffle, PA
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (910): : 719 - 725
  • [25] A technique for calibrating a high dose rate 192Ir brachytherapy source
    Fourie, O. L.
    Crabtree, T. G.
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2012, 35 (01) : 85 - 92
  • [26] A technique for calibrating a high dose rate 192Ir brachytherapy source
    O. L. Fourie
    T. G. Crabtree
    Australasian Physical & Engineering Sciences in Medicine, 2012, 35 : 85 - 92
  • [27] NPL's new absorbed dose standard for the calibration of HDR 192Ir brachytherapy sources
    Sander, T.
    Duane, S.
    Lee, N. D.
    Thomas, C. G.
    Owen, P. J.
    Bailey, M.
    Palmans, H.
    METROLOGIA, 2012, 49 (05) : S184 - S188
  • [28] Beam quality correction factors for dose measurements around 192Ir brachytherapy sources
    Thrapsanioti, Zoi
    Peppa, Vasiliki
    Hourdakis, Costas J.
    Karaiskos, Pantelis
    Lekatou, Aristea
    Pantelis, Evaggelos
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (01):
  • [29] Clinical application of the Fricke-glucomannan gel dosimeter for high-dose-rate 192Ir brachytherapy
    Noda, S.
    Suzuki, Y.
    Hoshino, Y.
    Furukawa, S.
    Katoh, H.
    Kurotaki, K.
    Nakano, T.
    PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (14): : 3985 - 3993
  • [30] First experience of 192Ir source stuck event during high-dose-rate brachytherapy in Japan
    Kumagai, Shinobu
    Arai, Norikazu
    Takata, Takeshi
    Kon, Daisuke
    Saitoh, Toshiya
    Oba, Hiroshi
    Furui, Shigeru
    Kotoku, Jun'ichi
    Shiraishi, Kenshiro
    JOURNAL OF CONTEMPORARY BRACHYTHERAPY, 2020, 12 (01) : 53 - 60