Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples

被引:5
|
作者
Yong, Sharon [1 ]
Liu, Hong [1 ]
Wong, Lingkai [1 ]
Teo, Hui Ling [1 ]
Chen, Yizhao [1 ]
Liu, Qinde [1 ]
Teo, Tang Lin [1 ]
Lee, Tong Kooi [1 ]
机构
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词
Urea; Human serum; Isotope dilution mass spectrometry; Liquid chromatography-isotope dilution; tandem mass spectrometry; Reference measurement procedure; External quality assessment;
D O I
10.1016/j.ijms.2017.01.013
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 50 条
  • [41] Determination of serum cortisol using isotope dilution–liquid chromatography–mass spectrometry as a candidate reference method
    Pyong Gil Jung
    Byungjoo Kim
    Sang-Ryoul Park
    Hun-Young So
    Lian Hua Shi
    Yongseong Kim
    Analytical and Bioanalytical Chemistry, 2004, 380 : 782 - 788
  • [42] Rapid determination of serum testosterone by liquid chromatography-isotope dilution tandem mass spectrometry and a split sample comparison with three automated immunoassays
    Chen, Yu
    Yazdanpanah, Mehrdad
    Hoffman, Barry R.
    Diamandis, Eleftherios P.
    Wong, Pui-Yuen
    CLINICAL BIOCHEMISTRY, 2009, 42 (06) : 484 - 490
  • [43] Determination of serum creatinine by liquid chromatography isotope dilution tandem mass spectrometry (LC-IDMS/MS): A candidate reference method
    Rasoul, Alikhani-Koupaei
    Mehrdad, Yazdanpanah
    Eleftherios, Diamandis
    Barry, Hoffman
    Khosrow, Adeli
    Jeffrey, Tong
    CLINICAL BIOCHEMISTRY, 2008, 41 (14-15) : 1273 - 1273
  • [44] Isotope dilution-liquid chromatography electrospray ionization-tandem mass spectrometry for the determination of serum thyroxine as a potential reference method
    De Brabandere, VI
    Hou, P
    Stockl, D
    Thienpont, LM
    De Leenheer, AP
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 1998, 12 (16) : 1099 - 1103
  • [45] Gas chromatography-isotope dilution mass spectrometry method for multi-level serum cholesterol analysis
    Edwards, S. H.
    Pyatt, S. D.
    Stribling, S. L.
    Washburn, R.
    Kimberly, M. M.
    Myers, G. L.
    CLINICAL CHEMISTRY, 2007, 53 (06) : A43 - A44
  • [46] An isotope dilution-liquid chromatography-tandem mass spectrometry-based candidate reference measurement procedure for the quantification of cortisol in human serum and plasma
    Ott, Myriam
    Singh, Neeraj
    Bauland, Friederike
    Koeppl, Daniel
    Kandler, Kerstin
    Gaudl, Alexander
    Geistanger, Andrea
    Ceglarek, Uta
    Rauh, Manfred
    Geletneky, Christian
    Taibon, Judith
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2025,
  • [47] Quantification of cholesterol in human serum by isotope dilution liquid chromatography/mass spectrometry
    Shin, Hyesun
    Lee, Hwashim
    Lee, Gaeho
    ANALYTICAL SCIENCE AND TECHNOLOGY, 2008, 21 (06): : 502 - 509
  • [48] Liquid Chromatography-Isotope Dilution-Mass Spectrometry as a New Basis for the Reference Measurement Procedure for Hemoglobin A1c Determination
    Kaiser, Patricia
    Akerboom, Theodorus
    Ohlendorf, Ruediger
    Reinauer, Hans
    CLINICAL CHEMISTRY, 2010, 56 (05) : 750 - 754
  • [49] An automated turbulent flow liquid chromatography-isotope dilution mass spectrometry (LC-IDMS) method for quantitation of serum creatinine
    Harlan, Robert
    Clarke, William
    Di Bussolo, Joseph M.
    Kozak, Marta
    Straseski, Joely
    Meany, Danni Li
    CLINICA CHIMICA ACTA, 2010, 411 (21-22) : 1728 - 1734
  • [50] Fast and Sensitive Analysis of Cefiderocol in Human Plasma Microsamples by Liquid Chromatography-Isotope Dilution Tandem Mass Spectrometry for Therapeutic Drug Monitoring
    Barone, Rossella
    Conti, Matteo
    Cojutti, Pier Giorgio
    Gatti, Milo
    Viale, Pierluigi
    Pea, Federico
    ANTIBIOTICS-BASEL, 2023, 12 (02):