Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples

被引:5
|
作者
Yong, Sharon [1 ]
Liu, Hong [1 ]
Wong, Lingkai [1 ]
Teo, Hui Ling [1 ]
Chen, Yizhao [1 ]
Liu, Qinde [1 ]
Teo, Tang Lin [1 ]
Lee, Tong Kooi [1 ]
机构
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词
Urea; Human serum; Isotope dilution mass spectrometry; Liquid chromatography-isotope dilution; tandem mass spectrometry; Reference measurement procedure; External quality assessment;
D O I
10.1016/j.ijms.2017.01.013
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 50 条
  • [21] Simultaneous measurement of total estradiol and testosterone in human serum by isotope dilution liquid chromatography tandem mass spectrometry
    Hui Zhou
    Yuesong Wang
    Matthew Gatcombe
    Jacob Farris
    Julianne C. Botelho
    Samuel P. Caudill
    Hubert W. Vesper
    Analytical and Bioanalytical Chemistry, 2017, 409 : 5943 - 5954
  • [22] Development of a candidate reference method for determination of tyrosine in serum by isotope dilution liquid chromatography-tandem mass spectrometry
    Lee, Hwashim
    Park, Sangryoul
    Kim, Byungjoo
    Lee, Gaeho
    ANALYTICAL SCIENCES, 2007, 23 (03) : 299 - 303
  • [23] An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma
    Zhang, Qiaoxuan
    Han, Liqiao
    Zheng, Songbai
    Ouyang, Fen
    Wu, Xiaobin
    Yan, Jun
    Zhan, Min
    Ke, Peifeng
    Zhuang, Junhua
    Huang, Xianzhang
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2021, 413 (17) : 4471 - 4481
  • [24] An isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma
    Qiaoxuan Zhang
    Liqiao Han
    Songbai Zheng
    Fen Ouyang
    Xiaobin Wu
    Jun Yan
    Min Zhan
    Peifeng Ke
    Junhua Zhuang
    Xianzhang Huang
    Analytical and Bioanalytical Chemistry, 2021, 413 : 4471 - 4481
  • [25] Development of a Candidate Reference Method for Determination of Tyrosine in Serum by Isotope Dilution Liquid Chromatography-Tandem Mass Spectrometry
    Hwashim Lee
    Sangryoul Park
    Byungjoo Kim
    Gaeho Lee
    Analytical Sciences, 2007, 23 : 299 - 303
  • [26] Measurement of serum uric acid by isotope dilution liquid chromatography tandem mass spectrometry: Modification of a candidate reference measurement method and its clinical application
    Yan, Jun
    Lin, Haibiao
    Xu, Yuzhu
    Yu, Lintao
    Chen, Song
    Zhang, Qiaoxuan
    Han, Liqiao
    Ou, Yangfen
    Zhan, Min
    Wang, Jianbing
    Ke, Peifeng
    Wang, Yunxiu
    Huang, Xianzhang
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2022, 482
  • [27] The analysis of thyroxine in human serum by an 'exact matching' isotope dilution method with liquid chromatography/tandem mass spectrometry
    Hopley, CJ
    Stokes, P
    Webb, KS
    Baynham, M
    RAPID COMMUNICATIONS IN MASS SPECTROMETRY, 2004, 18 (10) : 1033 - 1038
  • [28] Developing a reference measurement procedure for free glycerol in human serum by two-step gas chromatography-isotope dilution mass spectrometry
    Chen, Yizhao
    Teo, Hui Ling
    Liu, Qinde
    Lee, Tong Kooi
    CLINICAL BIOCHEMISTRY, 2015, 48 (13-14) : 897 - 903
  • [29] Quantitation of sibutramine in human hair using gas chromatography-isotope dilution tandem mass spectrometry
    Yang, Hongkun
    Wurita, Amin
    Liu, Jinlei
    Wang, Yue
    Hasegawa, Koutaro
    FORENSIC TOXICOLOGY, 2022, 40 (02) : 366 - 373
  • [30] Determination of Cortisol in Human Serum Using Ultrasonic Extraction Coupled with Liquid Chromatography-Isotope Dilution Mass Spectrometry
    Zhou Jian
    Wang Min
    Tang Xiao-Yan
    Mao Xue-Fei
    Dai Xin-Hua
    CHINESE JOURNAL OF ANALYTICAL CHEMISTRY, 2013, 41 (09) : 1439 - 1443