Liquid chromatography-isotope dilution tandem mass spectrometry method for the measurement of urea in human serum and assignment of reference values to external quality assessment samples

被引:5
|
作者
Yong, Sharon [1 ]
Liu, Hong [1 ]
Wong, Lingkai [1 ]
Teo, Hui Ling [1 ]
Chen, Yizhao [1 ]
Liu, Qinde [1 ]
Teo, Tang Lin [1 ]
Lee, Tong Kooi [1 ]
机构
[1] Chem Metrol Div, Appl Sci Grp, Hlth Sci Author, Singapore, Singapore
关键词
Urea; Human serum; Isotope dilution mass spectrometry; Liquid chromatography-isotope dilution; tandem mass spectrometry; Reference measurement procedure; External quality assessment;
D O I
10.1016/j.ijms.2017.01.013
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
Reference measurement procedures are essential for the accurate measurement of urea in human serum. The widely used gas chromatography-isotope dilution mass spectrometry (GC-IDMS) involves tedious derivatization processes. We developed and validated a simple liquid chromatography-isotope dilution tandem mass spectrometry (LC-IDMS/MS) method for the measurement of urea in human serum. In this method, C-13,N-15(2)-urea was spiked into serum samples as the internal standard, and pure urea certified reference material (CRM) was used as the calibration standard to ensure the traceability of the measurements to the International System of Units (SI). The sample cleanup procedure involved a simple protein precipitation step, followed by centrifugation and filtration. The developed LC-IDMS/MS method was found to be accurate and highly reproducible. The coefficient of variation (CV) was generally below 1%. Good accuracy was demonstrated by measuring serum CRM5 from two National Metrology Institutes (NMIs) with relative deviations from the certified values being close to or below 1%, and measurement uncertainties of 2.0%-2.8%. The method was used to assign the reference values of serum urea in ten cycles of an external quality assessment (EQA) program for clinical laboratories over a period of four years. The robust mean of the reported results from clinical laboratories using automated analyzers were close to the assigned values from LC-IDMS/MS method. With good accuracy and precision, as well as simple sample cleanup procedure, the developed LCIDMS/MS method can be used as a reference measurement procedure for the measurement of urea in human serum. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:87 / 93
页数:7
相关论文
共 50 条
  • [31] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Susan S.-C. Tai
    Bei Xu
    Michael J. Welch
    Karen W. Phinney
    Analytical and Bioanalytical Chemistry, 2007, 388 : 1087 - 1094
  • [32] Development and evaluation of a candidate reference measurement procedure for the determination of testosterone in human serum using isotope dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    Phinney, Karen W.
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2007, 388 (5-6) : 1087 - 1094
  • [33] Serum cholesterol measured by isotope dilution liquid chromatography tandem mass spectrometry
    Zhou, Weiyan
    Li, Hongxia
    Dong, Jun
    Wang, Shu
    Yang, Ruiyue
    Zhang, Chuanbao
    Chen, Wenxiang
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2011, 49 (04) : 669 - 676
  • [34] A candidate reference measurement procedure for quantification of glycocholic acid in human serum based on isotope dilution liquid chromatography-tandem mass spectrometry
    Zhang, Pingping
    Wang, Huimin
    Liang, Man
    Wang, Zhifang
    Liu, Chunlong
    Han, Yanlin
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2024, 416 (23) : 5121 - 5131
  • [35] Measurement of progesterone in human serum by isotope dilution liquid chromatography–tandem mass spectrometry and comparison with the commercial chemiluminescence immunoassay
    Hwashim Lee
    Chang Joon Park
    Gaeho Lee
    Analytical and Bioanalytical Chemistry, 2010, 396 : 1713 - 1719
  • [36] Development and evaluation of a candidate reference measurement procedure for the determination of progesterone in human serum using isotope-dilution liquid chromatography/tandem mass spectrometry
    Tai, Susan S. -C.
    Xu, Bei
    Welch, Michael J.
    ANALYTICAL CHEMISTRY, 2006, 78 (18) : 6628 - 6633
  • [37] Candidate reference measurement procedure for determination of urea in serum by liquid chromatography-tandem mass spectrometry
    Han, Liqiao
    Huang, Xiaoting
    Zhang, Lu
    Zhang, Qiaoxuan
    Wang, Jianbing
    Lin, Haibiao
    Yan, Jun
    Zhuang, Junhua
    Huang, Xianzhang
    JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 2019, 162 : 124 - 129
  • [38] Measurement of progesterone in human serum by isotope dilution liquid chromatography-tandem mass spectrometry and comparison with the commercial chemiluminescence immunoassay
    Lee, Hwashim
    Park, Chang Joon
    Lee, Gaeho
    ANALYTICAL AND BIOANALYTICAL CHEMISTRY, 2010, 396 (05) : 1713 - 1719
  • [39] Measurement of estradiol, estrone, and testosterone in postmenopausal human serum by isotope dilution liquid chromatography tandem mass spectrometry without derivatization
    Wooding, Kerry M.
    Hankin, Joseph A.
    Johnson, Chris A.
    Chosich, Justin D.
    Baek, Sung W.
    Bradford, Andrew P.
    Murphy, Robert C.
    Santoro, Nanette
    STEROIDS, 2015, 96 : 89 - 94
  • [40] DETERMINATION OF URINE CREATINE, CREATININE. GUANIDINOACETATE, AND ARGININE BY LIQUID CHROMATOGRAPHY-ISOTOPE DILUTION TANDEM MASS SPECTROMETRY
    Yazdanpanah, M.
    Adeli, K.
    vonBoth, I.
    Schulze, A.
    Kyriakopoulou, G. L.
    MOLECULAR GENETICS AND METABOLISM, 2009, 98 (1-2) : 134 - 134