CRITICAL APPRAISAL OF VOLUMETRIC MODULATED ARC THERAPY IN STEREOTACTIC BODY RADIATION THERAPY FOR METASTASES TO ABDOMINAL LYMPH NODES

被引:61
|
作者
Bignardi, Mario [2 ]
Cozzi, Luca [1 ]
Fogliata, Antonella
Lattuada, Paola [2 ]
Mancosu, Pietro [2 ]
Navarria, Piera [2 ]
Urso, Gaetano [2 ]
Vigorito, Sabrina [2 ]
Scorsetti, Marta [2 ]
机构
[1] Oncol Inst So Switzerland, Dept Radiat Oncol, Med Phys Unit, CH-6504 Bellinzona, Switzerland
[2] Ist Clin Humanitas, Rozzano, Italy
关键词
RapidArc; Stereotactic body radiation therapy; IMRT; Volumetric modulated are therapy; MONTE-CARLO CALCULATIONS; CLINICAL IMPLEMENTATION; CONFORMAL RADIOTHERAPY; INTENSITY MODULATION; DOSE CALCULATION; LIVER; IMRT; CANCER; TUMORS; LUNG;
D O I
10.1016/j.ijrobp.2009.05.035
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A planning study was performed comparing volumetric modulated arcs, RapidArc (RA), fixed beam IMRT (IM), and conformal radiotherapy (CRT) with multiple static fields or short conformal arcs in a series of patients treated with hypofractionated stereotactic body radiation therapy (SBRT) for solitary or oligo-metastases from different tumors to abdominal lymph nodes. Methods and Materials: Fourteen patients were included in the study. Dose prescription was set to 45 Gy (mean dose to clinical target volume [CTV]) in six fractions of 7.5 Gy. Objectives for CTV and planning target volume (PTV) were as follows: Dose(min) >95%, Dose(max) <107%. For organs at risk the following objectives were used: Maximum dose to spine <18 Gy; V-15Gy <35% for both kidneys, V-36Gy <1% for duodenum, V36Gy <3% for stomach and small bowel, V-15Gy <(total liver volume - 700 cm(3)) for liver. Dose-volume histograms were evaluated to assess plan quality. Results: Planning objectives on CTV and PTV were achieved by all techniques. Use of RA improved PTV coverage (V-95% = 90.2% +/- 5.2% for RA compared with 82.5% +/- 9.6% and 84.5% +/- 8.2% for CRT and IM, respectively). Most planning objectives for organs at risk were met by all techniques except for the duodenum, small bowel, and stomach, in which the CRT plans exceeded the dose/volume constraints in some patients. The MU/fraction values were as follows: 2186 +/- 211 for RA, 2583 +/- 699 for IM, and 1554 +/- 153 for CRT. Effective treatment time resulted as follows: 3.7 +/- 0.4 min for RA, 10.6 +/- 1.2 min for IM, and 6.3 +/- 0.5 min for CRT. Conclusions: Delivery of SBRT by RA showed improvements in conformal avoidance with respect to standard conformal irradiation. Delivery parameters confirmed logistical advantages of RA, particularly compared with IM. (C) 2009 Elsevier Inc.
引用
收藏
页码:1570 / 1577
页数:8
相关论文
共 50 条
  • [41] Stereotactic Body Radiation Therapy for Mediastinal and Hilar Lymph Node Metastases
    Shahi, Jeevin
    Poon, Ian
    Ung, Yee C.
    Tsao, May
    Bjarnason, Georg A.
    Malik, Nauman H.
    Zhang, Liying
    Louie, Alexander, V
    Cheung, Patrick
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2021, 109 (03): : 764 - 774
  • [42] Prostate Stereotactic Ablative Radiation Therapy Using Volumetric Modulated Arc Therapy to Dominant Intraprostatic Lesions
    Murray, Louise J.
    Lilley, John
    Thompson, Christopher M.
    Cosgrove, Vivian
    Mason, Josh
    Sykes, Jonathan
    Franks, Kevin
    Sebag-Montefiore, David
    Henry, Ann M.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2014, 89 (02): : 406 - 415
  • [43] A comparative dosimetric evaluation of dynamic conformal arc therapy and volumetric modulated arc therapy for lung stereotactic body radiotherapy
    Thanh, Nguyen Van
    Tuong, Pham Nguyen
    Tai, Duong Thanh
    Loan, Truong Thi Hong
    Sandwall, Peter
    Bradley, David
    Chow, James C. L.
    NUCLEAR ENGINEERING AND TECHNOLOGY, 2025, 57 (04)
  • [44] On the use of volumetric-modulated arc therapy for single-fraction thoracic vertebral metastases stereotactic body radiosurgery
    Pokhrel, Damodar
    Sood, Sumit
    McClinton, Christopher
    Shen, Xinglei
    Badkul, Rajeev
    Jiang, Hongyu
    Mallory, Matthew
    Mitchell, Mellissa
    Wang, Fen
    Lominska, Christopher
    MEDICAL DOSIMETRY, 2017, 42 (01) : 69 - 75
  • [45] Optimal combination of collimator angles for dual-arc volumetric modulated arc therapy planning in stereotactic body radiotherapy for spinal metastases
    Tsurumaki, Fumiya
    Nakajima, Yujiro
    Ito, Kei
    Kito, Satoshi
    Kikumura, Riki
    Murofushi, Keiko Nemoto
    Yorozu, Atsunori
    Fujita, Yukio
    MEDICAL DOSIMETRY, 2025, 50 (01) : 74 - 79
  • [46] Single fraction spine stereotactic ablative body radiotherapy with volumetric modulated arc therapy
    Matthew M. Gestaut
    Nitika Thawani
    Sangroh Kim
    Veera R. Gutti
    Sameer Jhavar
    Niloyjyoti Deb
    Andrew Morrow
    Russell A. Ward
    Jason H. Huang
    Mehul Patel
    Journal of Neuro-Oncology, 2017, 133 : 165 - 172
  • [47] Single fraction spine stereotactic ablative body radiotherapy with volumetric modulated arc therapy
    Gestaut, Matthew M.
    Thawani, Nitika
    Kim, Sangroh
    Gutti, Veera R.
    Jhavar, Sameer
    Deb, Niloyjyoti
    Morrow, Andrew
    Ward, Russell A.
    Huang, Jason H.
    Patel, Mehul
    JOURNAL OF NEURO-ONCOLOGY, 2017, 133 (01) : 165 - 172
  • [48] Dosimetric comparison of flattened and unflattened beams for stereotactic body radiation therapy: Impact of the size of the PTV on dynamic conformal arc and volumetric modulated arc therapy
    Vieillevigne, Laure
    Bessieres, Sandra
    Ouali, Monia
    Lanaspeze, Christel
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (11): : 1405 - 1414
  • [49] Novel Hybrid Volumetric Modulated Arc Therapy (VMAT) for Stereotactic Body Radiation Therapy (SBRT) of Lung Tumors in Proximity to the Chest Wall
    Ku, L.
    Reyes-Molyneux, N.
    Wang, S. J.
    Kang, M. Y.
    McDonald, T. L.
    JOURNAL OF THORACIC ONCOLOGY, 2012, 7 (09) : S258 - S258
  • [50] Dosimetric study for spine stereotactic body radiation therapy: magnetic resonance guided linear accelerator versus volumetric modulated arc therapy
    Yadav, Poonam
    Musunuru, Hima B.
    Witt, Jacob S.
    Bassetti, Michael
    Bayouth, John
    Baschnagel, Andrew M.
    RADIOLOGY AND ONCOLOGY, 2019, 53 (03) : 362 - 368