CRITICAL APPRAISAL OF VOLUMETRIC MODULATED ARC THERAPY IN STEREOTACTIC BODY RADIATION THERAPY FOR METASTASES TO ABDOMINAL LYMPH NODES

被引:61
|
作者
Bignardi, Mario [2 ]
Cozzi, Luca [1 ]
Fogliata, Antonella
Lattuada, Paola [2 ]
Mancosu, Pietro [2 ]
Navarria, Piera [2 ]
Urso, Gaetano [2 ]
Vigorito, Sabrina [2 ]
Scorsetti, Marta [2 ]
机构
[1] Oncol Inst So Switzerland, Dept Radiat Oncol, Med Phys Unit, CH-6504 Bellinzona, Switzerland
[2] Ist Clin Humanitas, Rozzano, Italy
关键词
RapidArc; Stereotactic body radiation therapy; IMRT; Volumetric modulated are therapy; MONTE-CARLO CALCULATIONS; CLINICAL IMPLEMENTATION; CONFORMAL RADIOTHERAPY; INTENSITY MODULATION; DOSE CALCULATION; LIVER; IMRT; CANCER; TUMORS; LUNG;
D O I
10.1016/j.ijrobp.2009.05.035
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Purpose: A planning study was performed comparing volumetric modulated arcs, RapidArc (RA), fixed beam IMRT (IM), and conformal radiotherapy (CRT) with multiple static fields or short conformal arcs in a series of patients treated with hypofractionated stereotactic body radiation therapy (SBRT) for solitary or oligo-metastases from different tumors to abdominal lymph nodes. Methods and Materials: Fourteen patients were included in the study. Dose prescription was set to 45 Gy (mean dose to clinical target volume [CTV]) in six fractions of 7.5 Gy. Objectives for CTV and planning target volume (PTV) were as follows: Dose(min) >95%, Dose(max) <107%. For organs at risk the following objectives were used: Maximum dose to spine <18 Gy; V-15Gy <35% for both kidneys, V-36Gy <1% for duodenum, V36Gy <3% for stomach and small bowel, V-15Gy <(total liver volume - 700 cm(3)) for liver. Dose-volume histograms were evaluated to assess plan quality. Results: Planning objectives on CTV and PTV were achieved by all techniques. Use of RA improved PTV coverage (V-95% = 90.2% +/- 5.2% for RA compared with 82.5% +/- 9.6% and 84.5% +/- 8.2% for CRT and IM, respectively). Most planning objectives for organs at risk were met by all techniques except for the duodenum, small bowel, and stomach, in which the CRT plans exceeded the dose/volume constraints in some patients. The MU/fraction values were as follows: 2186 +/- 211 for RA, 2583 +/- 699 for IM, and 1554 +/- 153 for CRT. Effective treatment time resulted as follows: 3.7 +/- 0.4 min for RA, 10.6 +/- 1.2 min for IM, and 6.3 +/- 0.5 min for CRT. Conclusions: Delivery of SBRT by RA showed improvements in conformal avoidance with respect to standard conformal irradiation. Delivery parameters confirmed logistical advantages of RA, particularly compared with IM. (C) 2009 Elsevier Inc.
引用
收藏
页码:1570 / 1577
页数:8
相关论文
共 50 条
  • [21] Implementation of a volumetric modulated arc therapy treatment planning solution for kidney and adrenal stereotactic body radiation therapy
    Sonier, Marcus
    Chu, William
    Lalani, Nafisha
    Erler, Darby
    Cheung, Patrick
    Korol, Renee
    Medical Dosimetry, 2016, 41 (04) : 323 - 328
  • [22] Abdominal Compression During Liver Stereotactic Body Radiation Therapy (SBRT) by Volumetric Modulated Arc Therapy (VMAT) Affords Significant Liver Sparing
    Snider, J. W., III
    Molitoris, J.
    Zhang, B.
    Chuong, M. D.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : S116 - S117
  • [23] Dosimetric comparison of volumetric modulated arc therapy and intensity-modulated radiation therapy for liver stereotactic ablative body radiotherapy
    Thibault, Isabelle
    Wronski, Matt
    Korol, Renee
    Chu, William
    Helou, Joelle
    Erler, Darby
    Chung, Hans T.
    JOURNAL OF RADIATION ONCOLOGY, 2016, 5 (01) : 63 - 69
  • [24] A Collimator Setting Optimization Algorithm for Dual-Arc Volumetric Modulated Arc Therapy in Pancreas Stereotactic Body Radiation Therapy
    Li, Xinyi
    Wu, Jackie
    Palta, Manisha
    Zhang, You
    Sheng, Yang
    Zhang, Jiahan
    Wang, Chunhao
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2019, 18
  • [25] A Collimator Setting Optimization Algorithm for Dual-Arc Volumetric Modulated Arc Therapy in Pancreas Stereotactic Body Radiation Therapy
    Li, X.
    Wu, Q.
    Palta, M.
    Zhang, Y.
    Sheng, Y.
    Zhang, J.
    Wang, C.
    MEDICAL PHYSICS, 2019, 46 (06) : E472 - E472
  • [26] A review of stereotactic body radiotherapy - is volumetric modulated arc therapy the answer?
    Sapkaroski, Daniel
    Osborne, Catherine
    Knight, Kellie A.
    JOURNAL OF MEDICAL RADIATION SCIENCES, 2015, 62 (02): : 142 - 151
  • [27] Volumetric modulated arc therapy treatment planning of thoracic vertebral metastases using stereotactic body radiotherapy
    Mallory, Matthew
    Pokhrel, Damodar
    Badkul, Rajeev
    Jiang, Hongyu
    Lominska, Christopher
    Wang, Fen
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2018, 19 (02): : 54 - 61
  • [28] Clinical Outcomes Using Stereotactic Body Radiation Therapy (SBRT) for Metastases to Retroperitoneal Lymph Nodes
    Khwaja, Shariq
    Barney, Brandon
    Olivier, Kenneth
    Miller, Robert
    Haddock, Michael
    AMERICAN JOURNAL OF CLINICAL ONCOLOGY-CANCER CLINICAL TRIALS, 2011, 34 (02): : 206 - 206
  • [29] Efficient Treatment Delivery of Lung Stereotactic Body Radiation Therapy (SBRT) Using Volumetric Modulated Arc Therapy (VMAT)
    Comsa, D. C.
    Purdie, T. G.
    Chao, J.
    Moseley, D. J.
    MEDICAL PHYSICS, 2009, 36 (09) : 4320 - 4320
  • [30] A COMPARISON OF STEREOTACTIC BODY RADIOTHERAPY MODALITIES: VOLUMETRIC MODULATED ARC THERAPY VERSUS INTENSITY MODULATED RADIATION THERAPY IN TREATMENT OF KIDNEY CANCER
    Chaudhary, Simone
    Swaminath, Anand
    Lukka, Himansu
    Chow, Tom
    Vansantvoort, Jasmin
    RADIOTHERAPY AND ONCOLOGY, 2019, 139 : S63 - S63