Finite element comparative analysis of three different internal fixation methods in the treatment of Pauwels type III femoral neck fractures

被引:18
|
作者
Ma, Ji [1 ]
Zhao, Ziying [2 ]
Zhi, Xiaodong [1 ,3 ]
Wang, Hao [1 ,3 ]
Wang, Wei [1 ,3 ,4 ]
机构
[1] Jinzhou Med Univ, Dept Orthped, Affiliated Hosp 1, Jinzhou 121000, Liaoning Provin, Peoples R China
[2] Jinzhou Med Univ, Dept Endocrinol, Affiliated Hosp 1, Jinzhou 121000, Liaoning Provin, Peoples R China
[3] Liaoning Key Lab Med Tissue Engn, Jinzhou 121000, Liaoning Provin, Peoples R China
[4] Jinzhou Med Univ, Inst Orthped, Jinzhou 121000, Liaoning Provin, Peoples R China
关键词
Femoral neck fracture; Pauwels III type; Finite element analysis; Biomechanics; HIP SCREW; YOUNG-PATIENTS; ADULTS;
D O I
10.1186/s12891-022-06003-3
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Comparison of 4 cannulated lag screws (3 inverted triangular cannulated screws + anti-rotating screws;4 CLS), dynamic hip screws + derotational screws (DHS + DS), and femoral neck fixation system (FNS) in the treatment of Biomechanical properties of middle-aged Pauwels type III femoral neck fractures. Methods: The femur CT data of a healthy young volunteer was selected and imported into Mimics software to construct a three-dimensional model of a normal femur. Pauwels type III femoral neck fractures were simulated according to the 70 degrees fracture line. Use Geomagic and SolidWorks software to optimize and build CLS, DHS + DS, and FNS fracture internal fixation models. Finally, Ansys software was used to analyze the stress distribution, peak value, and maximum displacement of the proximal fracture fragment and internal fixation; the displacement distribution, and peak value of the fracture surface at the fracture end. Results: (1) The stress peaks of the proximal fracture fragments in the three groups were concentrated near the femoral calcar. The peak stress of the FNS group was the largest, and the DHS + DS group was the smallest.(2)The displacement of the fracture fragments was all located at the top of the femur. The peak displacement of the FNS group was the largest, and the DHS + DS group was the smallest.(3) The internal fixation stress of the three groups is concentrated in the middle part of the device. The stress distribution of the first two groups of models is more uniform than that of FNS. The peak stress of FNS is the largest and the CLS is the smallest.(4) The internal fixed displacements are all located at the top of the model. The peak displacement of the CLS is the largest, and the DHS + DS is the smallest. (5) The displacement of the fracture surface is in the upper part of the fractured end. The peak displacement of the FNS group was the largest, and the DHS + DS group was the smallest. Conclusion: Compared with the other two internal fixation methods, dynamic hip screw + derotational screw (DHS + DS) showed good biomechanical stability. When Pauwels type III femoral neck fracture occurs in young adults, DHS + DS can be given priority as the preferred treatment for this type of fracture.
引用
收藏
页数:9
相关论文
共 50 条
  • [21] Biomechanical comparison of four different fixation methods in the management of Pauwels type III femoral neck fractures: Is there a clear winner?
    Yildirim, Cem
    Demirel, Mehmet
    Karahan, Goekhan
    Cetinkaya, Engin
    Misir, Abduelhamit
    Yamak, Fatih
    Bozdag, Erguen
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2022, 53 (10): : 3124 - 3129
  • [22] Modified F configuration in the treatment of Pauwels type III femoral neck fracture: a finite element analysis
    Lin, Shiyuan
    Shang, Jie
    Xing, Baizhou
    Wu, Biao
    Peng, Rong
    Wang, Gang
    Lu, Hua-Ding
    BMC MUSCULOSKELETAL DISORDERS, 2021, 22 (01)
  • [23] Modified F configuration in the treatment of Pauwels type III femoral neck fracture: a finite element analysis
    Shiyuan Lin
    Jie Shang
    Baizhou Xing
    Biao Wu
    Rong Peng
    Gang Wang
    Hua-Ding Lu
    BMC Musculoskeletal Disorders, 22
  • [24] Pre-sliding of femoral neck system improves fixation stability in pauwels type III femoral neck fracture: a finite element analysis
    Yonghan Cha
    Jun Young Chung
    Chang-Ho Jung
    Jin-Woo Kim
    Jeyoon Lee
    Jun-Il Yoo
    Jung-Taek Kim
    Yongho Jeon
    BMC Musculoskeletal Disorders, 24
  • [25] Pre-sliding of femoral neck system improves fixation stability in pauwels type III femoral neck fracture: a finite element analysis
    Cha, Yonghan
    Chung, Jun Young
    Jung, Chang-Ho
    Kim, Jin-Woo
    Lee, Jeyoon
    Yoo, Jun-Il
    Kim, Jung-Taek
    Jeon, Yongho
    BMC MUSCULOSKELETAL DISORDERS, 2023, 24 (01)
  • [26] Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures
    Jichao Liu
    Zhengwei Li
    Jie Ding
    Bingzhe Huang
    Chengdong Piao
    BMC Musculoskeletal Disorders, 23
  • [27] Biomechanical analysis of two medial buttress plate fixation methods to treat Pauwels type III femoral neck fractures
    Liu, Jichao
    Li, Zhengwei
    Ding, Jie
    Huang, Bingzhe
    Piao, Chengdong
    BMC MUSCULOSKELETAL DISORDERS, 2022, 23 (01)
  • [28] Correction to: Comparison of three different internal fixation implants in treatment of femoral neck fracture—a finite element analysis
    Jia Li
    Zhe Zhao
    Pengbin Yin
    Licheng Zhang
    Peifu Tang
    Journal of Orthopaedic Surgery and Research, 14
  • [29] Biomechanical study of internal fixation methods for femoral neck fractures based on Pauwels angle
    Wang, Fuyang
    Liu, Yuchen
    Huo, Yi
    Wang, Ziming
    Zhang, Jinge
    Xu, Minghao
    Ma, Kaiming
    Wang, Linbao
    Lu, Yongtao
    Cheng, Liangliang
    Zhao, Dewei
    FRONTIERS IN BIOENGINEERING AND BIOTECHNOLOGY, 2023, 11
  • [30] Femoral neck system and cannulated compression screws in the treatment of non-anatomical reduction Pauwels type-III femoral neck fractures: A finite element analysis
    Zhong, Zhou
    Lan, Xuejun
    Xiang, Zhou
    Duan, Xin
    CLINICAL BIOMECHANICS, 2023, 108