Comparison of 4-Class and Continuous Fat/Water Methods for Whole-Body, MR-Based PET Attenuation Correction

被引:71
|
作者
Wollenweber, S. D. [1 ]
Ambwani, S. [1 ]
Lonn, A. H. R. [2 ]
Shanbhag, D. D. [3 ]
Thiruvenkadam, S. [3 ]
Kaushik, S. [3 ]
Mullick, R. [3 ]
Qian, H. [4 ]
Delso, G. [5 ,6 ]
Wiesinger, F. [7 ]
机构
[1] GE Healthcare, Waukesha, WI 53005 USA
[2] GE Healthcare, Chalfont St Giles HP8 4SP, Bucks, England
[3] GE Global Res, Bangalore 560066, Karnataka, India
[4] GE Global Res, Niskayuna, NY 12309 USA
[5] GE Healthcare, CH-8006 Zurich, Switzerland
[6] Univ Zurich Hosp, CH-8006 Zurich, Switzerland
[7] GE Global Res, D-80331 Munich, Germany
关键词
Image analysis; magnetic resonance imaging; positron emission tomography; reconstruction algorithms; CLASSIFICATION; FAT;
D O I
10.1109/TNS.2013.2278759
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
The goal of this study was to compare two approaches for MR-based PET patient attenuation correction (AC) in whole-body FDG-PET imaging using a tri-modality PET/CT and MR setup. Sixteen clinical whole-body FDG patients were included in this study. Mean standard uptake values (SUVs) were measured for liver and lung volumes-of-interest for comparison. Maximum SUVs were measured in 18 FDG-avid features in 10 of the patients. The AC methods compared to gold-standard CT-based AC were segmentation of the CT (air, lung, fat, water), MR image segmentation with four tissue classes (air, lung, fat, water), and segmentation with air, lung and a continuous fat/water method. Results show that magnitude differences of the uptake values induced by CT-based image segmentation were similar but lower on average than those found using MR-derived AC methods. The average liver SUV difference as compared to using CTAC was 1.3%, 10.4%, and 5.7% for 4-class segmented CT, 4-class MRAC, and continuous fat/water MRAC methods, respectively. The average FDG-avid feature SUVmax difference was -0.5%, 1.7%, and -1.6% for 4-class segmented CT, 4-classMRAC, and continuous fat/water MRAC methods, respectively. The results demonstrated that both 4-class and continuous fat/water AC methods provided adequate quantitation in the body, and that the continuous fat/watermethod was within 5.7% on average for SUVmean in liver and 1.6% on average for SUVmax for FDG-avid features.
引用
收藏
页码:3391 / 3398
页数:8
相关论文
共 50 条
  • [31] Quantitative accuracy of attenuation correction in the Philips Ingenuity TF whole-body PET/MR system: a direct comparison with transmission-based attenuation correction
    Schramm, Georg
    Langner, Jens
    Hofheinz, Frank
    Petr, Jan
    Beuthien-Baumann, Bettina
    Platzek, Ivan
    Steinbach, Joerg
    Kotzerke, Joerg
    van den Hoff, Joerg
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2015, 28 (01) : 101 - 101
  • [32] Quantitative accuracy of attenuation correction in the Philips Ingenuity TF whole-body PET/MR system: a direct comparison with transmission-based attenuation correction
    Schramm, Georg
    Langner, Jens
    Hofheinz, Frank
    Petr, Jan
    Beuthien-Baumann, Bettina
    Platzek, Ivan
    Steinbach, Joerg
    Kotzerke, Joerg
    van den Hoff, Joerg
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2013, 26 (01) : 115 - 126
  • [33] Quantitative accuracy of attenuation correction in the Philips Ingenuity TF whole-body PET/MR system: a direct comparison with transmission-based attenuation correction
    Georg Schramm
    Jens Langner
    Frank Hofheinz
    Jan Petr
    Bettina Beuthien-Baumann
    Ivan Platzek
    Jörg Steinbach
    Jörg Kotzerke
    Jörg van den Hoff
    Magnetic Resonance Materials in Physics, Biology and Medicine, 2013, 26 : 115 - 126
  • [34] Potential influence of Gadolinium contrast on image segmentation in MR-based attenuation correction with Dixon sequences in whole-body 18F-FDG PET/MR
    Ruhlmann, Verena
    Heusch, Philipp
    Kuehl, Hilmar
    Beiderwellen, Karsten
    Antoch, Gerald
    Forsting, Michael
    Bockisch, Andreas
    Buchbender, Christian
    Quick, Harald H.
    MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE, 2016, 29 (02) : 301 - 308
  • [35] Deep learning based attenuation correction for whole-body PET studies
    Hwang, Donghwi
    Kim, Kyeong Yun
    Kang, Seung Kwan
    Seo, Seongho
    Paeng, Jin Chul
    Lee, Dong Soo
    Lee, Jae Sung
    JOURNAL OF NUCLEAR MEDICINE, 2018, 59
  • [36] Systematic evaluation of human soft tissue attenuation correction in whole-body PET/MR: Implications from PET/CT for optimization of MR-based AC in patients with normal lung tissue
    Lindemann, Maike E.
    Gratz, Marcel
    Grafe, Hong
    Jannusch, Kai
    Umutlu, Lale
    Quick, Harald H.
    MEDICAL PHYSICS, 2024, 51 (01) : 192 - 208
  • [37] MR-based attenuation correction for cardiac FDG PET on a hybrid PET/MRI scanner: comparison with standard CT attenuation correction
    Jan Vontobel
    Riccardo Liga
    Mathias Possner
    Olivier F. Clerc
    Fran Mikulicic
    Patrick Veit-Haibach
    Edwin E. G. W. ter Voert
    Tobias A. Fuchs
    Julia Stehli
    Aju P. Pazhenkottil
    Dominik C. Benz
    Christoph Gräni
    Oliver Gaemperli
    Bernhard Herzog
    Ronny R. Buechel
    Philipp A. Kaufmann
    European Journal of Nuclear Medicine and Molecular Imaging, 2015, 42 : 1574 - 1580
  • [38] MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation
    Akbarzadeh, A.
    Ay, M. R.
    Ahmadian, A.
    Alam, N. Riahi
    Zaidi, H.
    ANNALS OF NUCLEAR MEDICINE, 2013, 27 (02) : 152 - 162
  • [39] MR-based attenuation correction for cardiac FDG PET on a hybrid PET/MRI scanner: comparison with standard CT attenuation correction
    Vontobel, Jan
    Liga, Riccardo
    Possner, Mathias
    Clerc, Olivier F.
    Mikulicic, Fran
    Veit-Haibach, Patrick
    ter Voert, Edwin E. G. W.
    Fuchs, Tobias A.
    Stehli, Julia
    Pazhenkottil, Aju P.
    Benz, Dominik C.
    Graeni, Christoph
    Gaemperli, Oliver
    Herzog, Bernhard
    Buechel, Ronny R.
    Kaufmann, Philipp A.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2015, 42 (10) : 1574 - 1580
  • [40] MRI-guided attenuation correction in whole-body PET/MR: assessment of the effect of bone attenuation
    A. Akbarzadeh
    M. R. Ay
    A. Ahmadian
    N. Riahi Alam
    H. Zaidi
    Annals of Nuclear Medicine, 2013, 27 : 152 - 162