Accuracy of Nodule Volume and Airway Wall Thickness Measurement Using Low-Dose Chest CT on a Photon-Counting Detector CT Scanner

被引:15
|
作者
Dunning, Chelsea A. S. [1 ]
Marsh, Jeffrey F., Jr. [1 ]
Winfree, Timothy [1 ]
Rajendran, Kishore [1 ]
Leng, Shuai [1 ]
Levin, David L. [1 ]
Johnson, Tucker F. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
关键词
computed tomography; x-ray; lung cancer; pulmonary nodules; dose reduction; chronic obstructive pulmonary disease; OBSTRUCTIVE PULMONARY-DISEASE; NOISE-REDUCTION METHOD; IMAGE; IMPACT;
D O I
10.1097/RLI.0000000000000933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesA comparison of high-resolution photon-counting detector computed tomography (PCD-CT) versus energy-integrating detector (EID) CT via a phantom study using low-dose chest CT to evaluate nodule volume and airway wall thickness quantification.Materials and MethodsTwelve solid and ground-glass lung nodule phantoms with 3 diameters (5 mm, 8 mm, and 10 mm) and 2 shapes (spherical and star-shaped) and 12 airway tube phantoms (wall thicknesses, 0.27-1.54 mm) were placed in an anthropomorphic chest phantom. The phantom was scanned with EID-CT and PCD-CT at 5 dose levels (CTDIvol = 0.1-0.8 mGy at Sn-100 kV, 7.35 mGy at 120 kV). All images were iteratively reconstructed using matched kernels for EID-CT and medium-sharp kernel (MK) PCD-CT and an ultra-sharp kernel (USK) PCD-CT kernel, and image noise at each dose level was quantified. Nodule volumes were measured using semiautomated segmentation software, and the accuracy was expressed as the percentage error between segmented and reference volumes. Airway wall thicknesses were measured, and the root-mean-square error across all tubes was evaluated.ResultsMK PCD-CT images had the lowest noise. At 0.1 mGy, the mean volume accuracy for the solid and ground-glass nodules was improved in USK PCD-CT (3.1% and 3.3% error) compared with MK PCD-CT (9.9% and 10.2% error) and EID-CT images (11.4% and 9.2% error), respectively. At 0.2 mGy and 0.8 mGy, the wall thickness root-mean-square error values were 0.42 mm and 0.41 mm for EID-CT, 0.54 mm and 0.49 mm for MK PCD-CT, and 0.23 mm and 0.16 mm for USK PCD-CT.ConclusionsUSK PCD-CT provided more accurate lung nodule volume and airway wall thickness quantification at lower radiation dose compared with MK PCD-CT and EID-CT.
引用
收藏
页码:283 / 292
页数:10
相关论文
共 50 条
  • [11] Optimization of the Reconstruction Settings for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lungs
    Graafen, Dirk
    Halfmann, Moritz C.
    Emrich, Tilman
    Yang, Yang
    Kreuter, Michael
    Dueber, Christoph
    Kloeckner, Roman
    Mueller, Lukas
    Jorg, Tobias
    DIAGNOSTICS, 2023, 13 (23)
  • [12] Quantum Iterative Reconstruction for Low-Dose Ultra-High-Resolution Photon-Counting Detector CT of the Lung
    Sartoretti, Thomas
    Racine, Damien
    Mergen, Victor
    Jungblut, Lisa
    Monnin, Pascal
    Flohr, Thomas G.
    Martini, Katharina
    Frauenfelder, Thomas
    Alkadhi, Hatem
    Euler, Andre
    DIAGNOSTICS, 2022, 12 (02)
  • [13] Low-dose chest CT and the impact on nodule visibility
    Tugwell-Allsup, J.
    Owen, B. W.
    England, A.
    RADIOGRAPHY, 2021, 27 (01) : 24 - 30
  • [14] Feasibility of very low iodine dose aortoiliac CT angiography using dual-source photon-counting detector CT
    Oechsner, Tim
    Soschynski, Martin
    Schlett, Christopher L.
    Krauss, Tobias
    Schupppert, Christopher
    Mueller-Peltzer, Katharina
    Vecsey-Nagy, Milan
    Kravchenko, Dmitrij
    Varga-Szemes, Akos
    Emrich, Tilman
    Scheu, Raphael
    Taron, Jana
    Bamberg, Fabian
    Hagar, Muhammad Taha
    EUROPEAN JOURNAL OF RADIOLOGY, 2025, 183
  • [15] High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose
    Rajiah, Prabhakar Shantha
    Dunning, Chelsea A. S.
    Rajendran, Kishore
    Tandon, Yasmeen K.
    Ahmed, Zaki
    Larson, Nicholas B.
    Collins, Jeremy D.
    Thorne, Jamison
    Williamson, Eric
    Fletcher, Joel G.
    McCollough, Cynthia
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2023, 58 (09) : 681 - 690
  • [16] Pediatric contrast-enhanced chest CT on a photon-counting detector CT: radiation dose and image quality compared to energy-integrated detector CT
    El-Ali, Alexander M.
    Strubel, Naomi
    Pinkney, Lynne
    Xue, Christine
    Dane, Bari
    Lala, Shailee V.
    PEDIATRIC RADIOLOGY, 2024, 54 (12) : 1984 - 1995
  • [17] Optimizing Chest CT With Photon-counting Detector Technology: Comparison to Established Energy-integrating Detector CT Protocols
    Alarab, N.
    Guo, J.
    Atha, J.
    Sieren, J. C.
    Puliyakote, A. S. Kizhakke
    Hoffman, E. A.
    Abadi, E.
    Fain, S. B.
    AMERICAN JOURNAL OF RESPIRATORY AND CRITICAL CARE MEDICINE, 2024, 209
  • [18] Pulmonary nodule visualization and evaluation of AI-based detection at various ultra-low-dose levels using photon-counting detector CT
    Jungblut, Lisa
    Euler, Andre
    Landsmann, Anna
    Englmaier, Vanessa
    Mergen, Victor
    Sefirovic, Medina
    Frauenfelder, Thomas
    ACTA RADIOLOGICA, 2024, 65 (10) : 1238 - 1245
  • [19] Editorial Comment: Better Images of the Pelvis With Less Dose Using Photon-Counting Detector CT
    Roth, Trenton D.
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2024, 222 (01)
  • [20] Technical feasibility proof for high-resolution low-dose photon-counting CT of the breast
    Kalender, Willi A.
    Kolditz, Daniel
    Steiding, Christian
    Ruth, Veikko
    Lueck, Ferdinand
    Rossler, Ann-Christin
    Wenkel, Evelyn
    EUROPEAN RADIOLOGY, 2017, 27 (03) : 1081 - 1086