Accuracy of Nodule Volume and Airway Wall Thickness Measurement Using Low-Dose Chest CT on a Photon-Counting Detector CT Scanner

被引:15
|
作者
Dunning, Chelsea A. S. [1 ]
Marsh, Jeffrey F., Jr. [1 ]
Winfree, Timothy [1 ]
Rajendran, Kishore [1 ]
Leng, Shuai [1 ]
Levin, David L. [1 ]
Johnson, Tucker F. [1 ]
Fletcher, Joel G. [1 ]
McCollough, Cynthia H. [1 ]
Yu, Lifeng [1 ]
机构
[1] Mayo Clin, Dept Radiol, Rochester, MN USA
关键词
computed tomography; x-ray; lung cancer; pulmonary nodules; dose reduction; chronic obstructive pulmonary disease; OBSTRUCTIVE PULMONARY-DISEASE; NOISE-REDUCTION METHOD; IMAGE; IMPACT;
D O I
10.1097/RLI.0000000000000933
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectivesA comparison of high-resolution photon-counting detector computed tomography (PCD-CT) versus energy-integrating detector (EID) CT via a phantom study using low-dose chest CT to evaluate nodule volume and airway wall thickness quantification.Materials and MethodsTwelve solid and ground-glass lung nodule phantoms with 3 diameters (5 mm, 8 mm, and 10 mm) and 2 shapes (spherical and star-shaped) and 12 airway tube phantoms (wall thicknesses, 0.27-1.54 mm) were placed in an anthropomorphic chest phantom. The phantom was scanned with EID-CT and PCD-CT at 5 dose levels (CTDIvol = 0.1-0.8 mGy at Sn-100 kV, 7.35 mGy at 120 kV). All images were iteratively reconstructed using matched kernels for EID-CT and medium-sharp kernel (MK) PCD-CT and an ultra-sharp kernel (USK) PCD-CT kernel, and image noise at each dose level was quantified. Nodule volumes were measured using semiautomated segmentation software, and the accuracy was expressed as the percentage error between segmented and reference volumes. Airway wall thicknesses were measured, and the root-mean-square error across all tubes was evaluated.ResultsMK PCD-CT images had the lowest noise. At 0.1 mGy, the mean volume accuracy for the solid and ground-glass nodules was improved in USK PCD-CT (3.1% and 3.3% error) compared with MK PCD-CT (9.9% and 10.2% error) and EID-CT images (11.4% and 9.2% error), respectively. At 0.2 mGy and 0.8 mGy, the wall thickness root-mean-square error values were 0.42 mm and 0.41 mm for EID-CT, 0.54 mm and 0.49 mm for MK PCD-CT, and 0.23 mm and 0.16 mm for USK PCD-CT.ConclusionsUSK PCD-CT provided more accurate lung nodule volume and airway wall thickness quantification at lower radiation dose compared with MK PCD-CT and EID-CT.
引用
收藏
页码:283 / 292
页数:10
相关论文
共 50 条
  • [31] Bullet characterization using Photon-Counting detector CT: A phantom with intact bullets
    Hardy, Joey C. A.
    Crombag, Genevieve A. J. C.
    Peters, Nicky H. G. M.
    Hermsen, Rob
    Willigers, Jef L.
    Nobel, J. Martijn
    Wildberger, Joachim E.
    Flohr, Thomas G.
    Postma, Alida A.
    EUROPEAN JOURNAL OF RADIOLOGY, 2025, 183
  • [32] The Eye of the CT Scanner: The story of learning to see the invisible or from the fluorescent screen to the photon-counting detector
    Schlemmer, Heinz-Peter
    ROFO-FORTSCHRITTE AUF DEM GEBIET DER RONTGENSTRAHLEN UND DER BILDGEBENDEN VERFAHREN, 2021, 193 (09): : 1034 - 1048
  • [33] Low-Dose High-Resolution Photon-Counting CT of the Lung: Radiation Dose and Image Quality in the Clinical Routine
    Woeltjen, Matthias Michael
    Niehoff, Julius Henning
    Michael, Arwed Elias
    Horstmeier, Sebastian
    Moenninghoff, Christoph
    Borggrefe, Jan
    Kroeger, Jan Robert
    DIAGNOSTICS, 2022, 12 (06)
  • [34] Photon-counting detector CT allows significant reduction in radiation dose while maintaining image quality and noise on non-contrast chest CT
    Donuru, Achala
    Araki, Tetsuro
    Dako, Farouk
    Dave, Jaydev K.
    Perez, Raul Porto
    Xu, Dongming
    Nachiappan, Arun C.
    Barbosa Jr, Eduardo Mortani
    Noel, Peter
    Litt, Harold
    Knollman, Friedrich
    EUROPEAN JOURNAL OF RADIOLOGY OPEN, 2023, 11
  • [35] Quantification accuracy in photon-counting detector CT for coronary artery calcium score: a pilot study
    Wang, Mengzhen
    Zhang, Xu
    Li, Jiqang
    Dong, Haipeng
    Xu, Zhihan
    Yan, Fuhua
    Yang, Wenjie
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2024, 40 (10): : 2181 - 2191
  • [36] Detectability and Volumetric Accuracy of Pulmonary Nodules in Low-Dose Photon-Counting Detector Computed Tomography: An Anthropomorphic Phantom Study
    Hop, Joost F.
    Walstra, Anna N. H.
    Pelgrim, Gert-Jan
    Xie, Xueqian
    Panneman, Noor A.
    Schurink, Niels W.
    Faby, Sebastian
    van Straten, Marcel
    de Bock, Geertruida H.
    Vliegenthart, Rozemarijn
    Greuter, Marcel J. W.
    DIAGNOSTICS, 2023, 13 (22)
  • [37] Liver Iodine Quantification With Photon-Counting Detector CT: Accuracy in an Abdominal Phantom and Feasibility in Patients
    Sartoretti, Thomas
    Mergen, Victor
    Jungblut, Lisa
    Alkadhi, Hatem
    Euler, Andre
    ACADEMIC RADIOLOGY, 2023, 30 (03) : 461 - 469
  • [38] Potential dose reduction and image quality improvement in chest CT with a photon-counting CT compared to a new dual-source CT
    Greffier, Joel
    Dabli, Djamel
    Faby, Sebastian
    Pastor, Maxime
    de Oliveira, Fabien
    Croisille, Cedric
    Erath, Julien
    Beregi, Jean-Paul
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 127
  • [39] Feasibility of Dose-reduced Chest CT with Photon-counting Detectors: Initial Results in Humans
    Symons, Rolf
    Pourmorteza, Amir
    Sandfort, Veit
    Ahlman, Mark A.
    Cropper, Tracy
    Mallek, Marissa
    Kappler, Steffen
    Ulzheimer, Stefan
    Mahesh, Mahadevappa
    Jones, Elizabeth C.
    Malayeri, Ashkan A.
    Folio, Les R.
    Bluemke, David A.
    RADIOLOGY, 2017, 285 (03) : 980 - 989
  • [40] Dose Reduction for Sinus and Temporal Bone Imaging Using Photon-Counting Detector CT With an Additional Tin Filter
    Rajendran, Kishore
    Voss, Benjamin A.
    Zhou, Wei
    Tao, Shengzhen
    DeLone, David R.
    Lane, John, I
    Weaver, Jayse M.
    Carlson, Matthew L.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2020, 55 (02) : 91 - 100