Detectability and Volumetric Accuracy of Pulmonary Nodules in Low-Dose Photon-Counting Detector Computed Tomography: An Anthropomorphic Phantom Study

被引:3
|
作者
Hop, Joost F. [1 ]
Walstra, Anna N. H. [1 ]
Pelgrim, Gert-Jan [1 ]
Xie, Xueqian [2 ]
Panneman, Noor A. [1 ]
Schurink, Niels W. [3 ]
Faby, Sebastian [4 ]
van Straten, Marcel [5 ]
de Bock, Geertruida H. [6 ]
Vliegenthart, Rozemarijn [1 ]
Greuter, Marcel J. W. [1 ]
机构
[1] Univ Med Ctr Groningen, Dept Radiol, NL-9713 GZ Groningen, Netherlands
[2] Shanghai Jiao Tong Univ, Shanghai Gen Hosp, Dept Radiol, Sch Med, Shanghai 200080, Peoples R China
[3] Siemens Healthineers Nederland BV, Nl-2595 BN Den Hague, Netherlands
[4] Siemens Healthcare GmbH, Computed Tomog, D-91301 Forchheim, Germany
[5] Erasmus Univ, Med Ctr, Dept Radiol, NL-3015 GD Rotterdam, Netherlands
[6] Univ Med Ctr Groningen, Dept Epidemiol, NL-9713 GZ Groningen, Netherlands
关键词
computed tomography; X-ray; photon-counting detector; low-dose CT; lung cancer; imaging phantom; pulmonary nodules; dose reduction; LUNG-CANCER; CT; AGREEMENT;
D O I
10.3390/diagnostics13223448
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this phantom study was to assess the detectability and volumetric accuracy of pulmonary nodules on photon-counting detector CT (PCD-CT) at different low-dose levels compared to conventional energy-integrating detector CT (EID-CT). In-house fabricated artificial nodules of different shapes (spherical, lobulated, spiculated), sizes (2.5-10 mm and 5-1222 mm(3)), and densities (-330 HU and 100 HU) were randomly inserted into an anthropomorphic thorax phantom. The phantom was scanned with a low-dose chest protocol with PCD-CT and EID-CT, in which the dose with PCD-CT was lowered from 100% to 10% with respect to the EID-CT reference dose. Two blinded observers independently assessed the CT examinations of the nodules. A third observer measured the nodule volumes using commercial software. The influence of the scanner type, dose, observer, physical nodule volume, shape, and density on the detectability and volumetric accuracy was assessed by a multivariable regression analysis. In 120 CT examinations, 642 nodules were present. Observer 1 and 2 detected 367 (57%) and 289 nodules (45%), respectively. With PCD-CT and EID-CT, the nodule detectability was similar. The physical nodule volumes were underestimated by 20% (range 8-52%) with PCD-CT and 24% (range 9-52%) with EID-CT. With PCD-CT, no significant decrease in the detectability and volumetric accuracy was found at dose reductions down to 10% of the reference dose (p > 0.05). The detectability and volumetric accuracy were significantly influenced by the observer, nodule volume, and a spiculated nodule shape (p < 0.05), but not by dose, CT scanner type, and nodule density (p > 0.05). Low-dose PCD-CT demonstrates potential to detect and assess the volumes of pulmonary nodules, even with a radiation dose reduction of up to 90%.
引用
收藏
页数:13
相关论文
共 50 条
  • [1] Ultra-low-dose spectral-detector computed tomography for the accurate quantification of pulmonary nodules: an anthropomorphic chest phantom study
    Zhou, Xiuxiu
    Zhang, Hanxiao
    Jin, Xiaoxing
    Zhang, Xiaohui
    Lu, Xiaomei
    Han, Qun
    Xiong, Xiaoge
    Liu, Tian
    Feng, Yan
    Tu, Wenting
    Zhou, Taohu
    Ge, Yanming
    Dong, Peng
    Liu, Shiyuan
    Fan, Li
    DIAGNOSTIC AND INTERVENTIONAL RADIOLOGY, 2023, 29 (05): : 691 - 703
  • [2] Lesion Detectability and Radiation Dose in Spiral Breast CT With Photon-Counting Detector Technology A Phantom Study
    Shim, Sojin
    Saltybaeva, Natalia
    Berger, Nicole
    Marcon, Magda
    Alkadhi, Hatem
    Boss, Andreas
    INVESTIGATIVE RADIOLOGY, 2020, 55 (08) : 515 - 523
  • [3] Image Quality and Pulmonary Nodule Detectability at Low-dose Computed Tomography (low kVp and mAs): A phantom study
    Iranmakani, Sepideh
    Jahanshahi, Amir Reza
    Mehnati, Parinaz
    Mortezazadeh, Tohid
    Khezerloo, Davood
    JOURNAL OF MEDICAL SIGNALS & SENSORS, 2022, 12 (01): : 64 - 68
  • [4] Performance of calcium quantifications on low-dose photon-counting detector CT with high-pitch: A phantom study
    Zhou, Shanshui
    Liu, Peng
    Dong, Haipeng
    Li, Jiqiang
    Xu, Zhihan
    Schmidt, Bernhard
    Lin, Shushen
    Yang, Wenjie
    Yan, Fuhua
    Qin, Le
    HELIYON, 2024, 10 (12)
  • [5] Microcalcification and Mass Detection with a Photon-Counting Si Detector for Low Dose Breast Computed Tomography
    Ding, H.
    Cho, H.
    Masaki, F.
    Barber, W.
    Iwanczyk, J.
    Molloi, S.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [6] Pulmonary nodule volumetric accuracy of a deep learning-based reconstruction algorithm in low-dose computed tomography: A phantom study
    Watanabe, Shota
    Sakaguchi, Kenta
    Kitaguchi, Shigetoshi
    Ishii, Kazunari
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2022, 104 : 1 - 9
  • [7] Novel photon-counting low-dose computed tomography using a multi-pixel photon counter
    Morita, H.
    Oshima, T.
    Kataoka, J.
    Arimoto, M.
    Nitta, H.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 857 : 58 - 65
  • [8] CT technology: photon-counting detector computed tomography
    Layer, Yannik Christian
    Kravchenko, Dmitrij
    Dell, Tatjana
    Kuetting, Daniel
    RADIOLOGIE, 2023, 63 (07): : 497 - 506
  • [9] Photon-counting detector computed tomography in cardiac imaging
    Sharma, Simran P.
    Lemmens, Marie-Julie D. K.
    Smulders, Martijn W.
    Budde, Ricardo P. J.
    Hirsch, Alexander
    Mihl, Casper
    NETHERLANDS HEART JOURNAL, 2024, 32 (11) : 405 - 416
  • [10] Subsolid Pulmonary Nodules in Low-dose Computed Tomography
    Hekmat, K.
    Bruns, C. J.
    CHIRURG, 2019, 90 (Suppl 2): : S113 - S113