Detectability and Volumetric Accuracy of Pulmonary Nodules in Low-Dose Photon-Counting Detector Computed Tomography: An Anthropomorphic Phantom Study

被引:3
|
作者
Hop, Joost F. [1 ]
Walstra, Anna N. H. [1 ]
Pelgrim, Gert-Jan [1 ]
Xie, Xueqian [2 ]
Panneman, Noor A. [1 ]
Schurink, Niels W. [3 ]
Faby, Sebastian [4 ]
van Straten, Marcel [5 ]
de Bock, Geertruida H. [6 ]
Vliegenthart, Rozemarijn [1 ]
Greuter, Marcel J. W. [1 ]
机构
[1] Univ Med Ctr Groningen, Dept Radiol, NL-9713 GZ Groningen, Netherlands
[2] Shanghai Jiao Tong Univ, Shanghai Gen Hosp, Dept Radiol, Sch Med, Shanghai 200080, Peoples R China
[3] Siemens Healthineers Nederland BV, Nl-2595 BN Den Hague, Netherlands
[4] Siemens Healthcare GmbH, Computed Tomog, D-91301 Forchheim, Germany
[5] Erasmus Univ, Med Ctr, Dept Radiol, NL-3015 GD Rotterdam, Netherlands
[6] Univ Med Ctr Groningen, Dept Epidemiol, NL-9713 GZ Groningen, Netherlands
关键词
computed tomography; X-ray; photon-counting detector; low-dose CT; lung cancer; imaging phantom; pulmonary nodules; dose reduction; LUNG-CANCER; CT; AGREEMENT;
D O I
10.3390/diagnostics13223448
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The aim of this phantom study was to assess the detectability and volumetric accuracy of pulmonary nodules on photon-counting detector CT (PCD-CT) at different low-dose levels compared to conventional energy-integrating detector CT (EID-CT). In-house fabricated artificial nodules of different shapes (spherical, lobulated, spiculated), sizes (2.5-10 mm and 5-1222 mm(3)), and densities (-330 HU and 100 HU) were randomly inserted into an anthropomorphic thorax phantom. The phantom was scanned with a low-dose chest protocol with PCD-CT and EID-CT, in which the dose with PCD-CT was lowered from 100% to 10% with respect to the EID-CT reference dose. Two blinded observers independently assessed the CT examinations of the nodules. A third observer measured the nodule volumes using commercial software. The influence of the scanner type, dose, observer, physical nodule volume, shape, and density on the detectability and volumetric accuracy was assessed by a multivariable regression analysis. In 120 CT examinations, 642 nodules were present. Observer 1 and 2 detected 367 (57%) and 289 nodules (45%), respectively. With PCD-CT and EID-CT, the nodule detectability was similar. The physical nodule volumes were underestimated by 20% (range 8-52%) with PCD-CT and 24% (range 9-52%) with EID-CT. With PCD-CT, no significant decrease in the detectability and volumetric accuracy was found at dose reductions down to 10% of the reference dose (p > 0.05). The detectability and volumetric accuracy were significantly influenced by the observer, nodule volume, and a spiculated nodule shape (p < 0.05), but not by dose, CT scanner type, and nodule density (p > 0.05). Low-dose PCD-CT demonstrates potential to detect and assess the volumes of pulmonary nodules, even with a radiation dose reduction of up to 90%.
引用
收藏
页数:13
相关论文
共 50 条
  • [31] Low-dose lung cancer screening with photon-counting CT: a feasibility study
    Symons, Rolf
    Cork, Tyler E.
    Sahbaee, Pooyan
    Fuld, Matthew K.
    Kappler, Steffen
    Folio, Les R.
    Bluemke, David A.
    Pourmorteza, Amir
    PHYSICS IN MEDICINE AND BIOLOGY, 2017, 62 (01): : 202 - 213
  • [32] Photon-counting detector computed tomography Paradigm shift in cardiac CT imaging
    Kravchenko, Dmitrij
    Layer, Yannik Christian
    Vecsey-Nagy, Milan
    Tremamunno, Giuseppe
    Varga-Szemes, Akos
    Schlett, Christopher L.
    Bamberg, Fabian
    Emrich, Tilman
    Hagar, Muhammad Taha
    RADIOLOGIE, 2024, 64 (12): : 928 - 934
  • [33] Consistency of radiologists in identifying pulmonary nodules based on low-dose computed tomography
    Ming, Shuai
    Yang, Wei
    Cui, Si-Jia
    Huang, Shuai
    Gong, Xiang-Yang
    JOURNAL OF THORACIC DISEASE, 2019, 11 (07) : 2973 - 2980
  • [34] Assessment of Bone Mineral Density From a Computed Tomography Topogram of Photon-Counting Detector Computed Tomography-Effect of Phantom Size and Tube Voltage
    Euler, Andre
    Nowak, Tristan
    Bucher, Brian
    Eberhard, Matthias
    Schmidt, Bernhard
    Flohr, Thomas G.
    Frey, Diana
    Distler, Oliver
    Alkadhi, Hatem
    INVESTIGATIVE RADIOLOGY, 2021, 56 (10) : 614 - 620
  • [35] Subsolide Lungenrundherde im Low-dose-ComputertomogrammSubsolid pulmonary nodules in low-dose computed tomography
    K. Hekmat
    C. J. Bruns
    Der Chirurg, 2018, 89 : 1008 - 1008
  • [36] Subsolide Lungenrundherde im Low-dose-ComputertomogrammSubsolid pulmonary nodules in low-dose computed tomography
    K. Hekmat
    C. J. Bruns
    Der Chirurg, 2019, 90 : 113 - 113
  • [37] Photon-Counting Detector Computed Tomography Versus Energy-Integrating Detector Computed Tomography for Coronary Artery Calcium Quantitation
    McCollough, Cynthia H.
    Winfree, Tim N.
    Melka, Elnata F.
    Rajendran, Kishore
    Carter, Rickey E.
    Leng, Shuai
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2024, 48 (02) : 212 - 216
  • [38] Reduction of Metal Artifacts and Improvement in Dose Efficiency Using Photon-Counting Detector Computed Tomography and Tin Filtration
    Zhou, Wei
    Bartlett, David J.
    Diehn, Felix E.
    Glazebrook, Katrina N.
    Kotsenas, Amy L.
    Carter, Rickey E.
    Fletcher, Joel G.
    McCollough, Cynthia H.
    Leng, Shuai
    INVESTIGATIVE RADIOLOGY, 2019, 54 (04) : 204 - 211
  • [39] Radiation Dose Reduction for Coronary Artery Calcium Scoring Using a Virtual Noniodine Algorithm on Photon-Counting Detector Computed-Tomography Phantom Data
    Fink, Nicola
    Zsarnoczay, Emese
    Schoepf, U. Joseph
    O'Doherty, Jim
    Griffith III, Joseph P. P.
    Pinos, Daniel
    Tesche, Christian
    Ricke, Jens
    Willemink, Martin J.
    Varga-Szemes, Akos
    Emrich, Tilman
    DIAGNOSTICS, 2023, 13 (09)
  • [40] Innovative advances in pediatric radiology: computed tomography reconstruction techniques, photon-counting detector computed tomography, and beyond
    Mese, Ismail
    Mese, Ceren Altintas
    Demirsoy, Ugur
    Anik, Yonca