Preclinical evaluation of semi-automated laser ablation for pulmonary vein isolation: A comparative study

被引:1
|
作者
Kuroki, Kenji [1 ]
Reddy, Vivek Y. [1 ]
Iwasawa, Jin [1 ]
Kawamura, Iwanari [1 ]
Neuzil, Petr [2 ]
Estabrook, Brian [3 ]
Melsky, Gerald [3 ]
Dukkipati, Srinivas R. [1 ]
Koruth, Jacob [1 ,4 ]
机构
[1] Icahn Sch Med Mt Sinai, Helmsley Electrophysiol Ctr, New York, NY USA
[2] Hosp Na Homolce, Dept Cardiol, Prague, Czech Republic
[3] CardioFocus Inc, Marlborough, MA USA
[4] Icahn Sch Med Mt Sinai, Helmsley Electrophysiol Ctr, One Gustave L Levy Pl,POB 1030, New York, NY 10029 USA
关键词
atrial fibrillation; automated; catheter ablation; endoscopic; laser balloon; POINT-BY-POINT; CATHETER ABLATION; PAROXYSMAL AF; MULTICENTER; BALLOON; TRIAL;
D O I
10.1111/jce.15777
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
IntroductionVisually-guided laser balloon ablation (VGLA) currently requires careful manual rotation of the laser to create overlapping lesions. A novel semi-automated VGLA may reduce ablation times and lesion gaps. We aimed to compare semi-automated (SA) VGLA to that of manual (MN) VGLA. MethodsAcute: Nine swine underwent right superior pulmonary vein isolation (PVI) using either SA (n = 3, 13-18 W), MN (n = 3, 8.5-12 W), or radiofrequency (RF, n = 3, 25-40 W) and were killed acutely. Chronic: 16 swine, underwent PVI using either SA (n = 8, 15 W) or MN (n = 8, 10 W), and were survived for 1 month before being killed. All hearts were then submitted for pathological evaluation. ResultsAcute: PVI was successful in all 9/9 swine with lesion counts significantly lower in the SA arm (5.3 +/- 5.9, 33.7 +/- 10.0, and 28.0 +/- 4.4 in SA, MN, and RF arms; p = .007 for SA and MN). At necropsy, circumferentiality and transmurality were 98% and 94% in SA, 98% and 80% in MN, and 100% and 100% in RF arms. A single steam pop was noted on sectioning in the SA arm swine and occurred in the high dose (18 W) strategy. Chronic: PVI was acutely successful in 16/16 swine with no difference in PVI durability rates (62.5% vs. 75.0%), lesion transmurality (95.8 +/- 17.4% vs.91.9 +/- 25.9%), and circumferentiality (95.8 +/- 6.6% vs. 94.8 +/- 6.3%) between SA and MN arms. Catheter use time and lesion counts were lower in the SA arm compared to the MN arm (11.5 +/- 12.7 vs. 21.8 +/- 3.8 min, p = .046 and 4.8 +/- 3.83 vs. 35.4 +/- 4.4, p < .001). ConclusionMotor-assisted semi-automated laser balloon ablation can improve upon procedural efficiency by reducing ablation time.
引用
收藏
页码:315 / 324
页数:10
相关论文
共 50 条
  • [31] A simple, robust and semi-automated parasite egg isolation protocol
    Mes, Ted H. M.
    Eysker, Maarten
    Ploeger, Harm W.
    NATURE PROTOCOLS, 2007, 2 (03) : 486 - 489
  • [32] Pulmonary vein stenosis after pulmonary vein ablation catheter-guided pulmonary vein isolation
    De Greef, Yves
    Schwagten, Bruno
    De Keulenaer, Gilles
    Stockman, Dirk
    HEART RHYTHM, 2010, 7 (09) : 1306 - 1308
  • [33] A simple, robust and semi-automated parasite egg isolation protocol
    Ted HM Mes
    Maarten Eysker
    Harm W Ploeger
    Nature Protocols, 2007, 2 : 486 - 489
  • [34] A multicentered evaluation of ablation at higher power guided by ablation index: Establishing ablation targets for pulmonary vein isolation
    Dhillon, Gurpreet
    Ahsan, Syed
    Honarbakhsh, Shohreh
    Lim, Wei
    Baca, Marco
    Graham, Adam
    Srinivasan, Neil
    Sawhney, Vinit
    Sporton, Simon
    Schilling, Richard J.
    Chow, Anthony
    Ginks, Matthew
    Sohal, Manav
    Gallagher, Mark M.
    Hunter, Ross J.
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2019, 30 (03) : 357 - 365
  • [35] A semi-automated method for evaluation of touch-escape behavior in isolation-reared mice
    Kuchiiwa, Toshiko
    Miura, Kae
    Kuchiiwa, Satoshi
    JOURNAL OF PHARMACOLOGICAL SCIENCES, 2015, 128 (03) : S163 - S163
  • [36] Pulmonary vein isolation alone versus pulmonary vein isolation with additional extensive ablation for atrial fibrillation
    Junarta, Joey
    Upadhyay, Naman
    Molin, Andrea
    Bodempudi, Sairamya
    Dikdan, Sean
    Warner, Eric
    Joffe, Daniel
    Pang, Zachary
    Frisch, Daniel
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2022, 33 (04) : 799 - 799
  • [37] Preclinical evaluation of a novel single-shot pulsed field ablation system for pulmonary vein and atrial ablation
    Aryana, Arash
    Ji, Sang Yong
    Hata, Cary
    de la Rama, Alan
    Nguyen, Ken
    Panescu, Dorin
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2023, 34 (11) : 2203 - 2212
  • [38] Evaluation of Pulmonary Vein Stenosis After Pulmonary Vein Isolation Using a Novel Circular Mapping and Ablation Catheter (PVAC)
    von Bary, Christian
    Weber, Stefan
    Dornia, Christian
    Eissnert, Christoph
    Fellner, Claudia
    Latzin, Philipp
    Fredersdorf, Sabine
    Stadler, Stefan
    Hamer, Okka W.
    CIRCULATION-ARRHYTHMIA AND ELECTROPHYSIOLOGY, 2011, 4 (05): : 630 - 636
  • [39] Automated Annotation-Guided Radiofrequency Ablation Improves Learning Curve for Pulmonary Vein Isolation
    Tanaka, Nobuaki
    Inoue, Koichi
    Tanaka, Koji
    Onishi, Toshinari
    Hirao, Yuko
    Oka, Takafumi
    Okada, Masato
    Yoshimoto, Issei
    Kitagaki, Ryo
    Iwakura, Katsuomi
    Fujii, Kenshi
    CIRCULATION, 2019, 140
  • [40] Pulmonary vein isolation using laser balloon as compared to standard radiofrequency catheter ablation
    Koopman, Pieter
    Vanhentenrijk, Simon
    Schurmans, Joris
    Dining, Dagmara
    Vijgen, Johan
    ACTA CARDIOLOGICA, 2017, 72 (01) : 89 - 90