Motion artifact variability in biomagnetic wearable devices

被引:0
|
作者
Arekhloo, Negin Ghahremani [1 ,2 ]
Wang, Huxi [1 ,2 ]
Parvizi, Hossein [2 ]
Tanwear, Asfand [1 ]
Zuo, Siming [1 ]
McKinlay, Michael [2 ]
Nunez, Carlos Garcia [2 ]
Nazarpour, Kianoush [1 ,3 ]
Heidari, Hadi [1 ,2 ]
机构
[1] Neuranics Ltd, Glasgow, Scotland
[2] Univ Glasgow, James Watt Sch Engn, Microelect Lab, Glasgow City, Scotland
[3] Univ Edinburgh, Sch Informat, Edinburgh, Scotland
来源
FRONTIERS IN MEDICAL TECHNOLOGY | 2024年 / 6卷
基金
英国工程与自然科学研究理事会;
关键词
biomagnetic measurements; motion artifacts; wearable sensors; gradient background field; homogeneous background field; SIGNAL; ECG;
D O I
10.3389/fmedt.2024.1457535
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Motion artifacts can be a significant noise source in biomagnetic measurements when magnetic sensors are not separated from the signal source. In ambient environments, motion artifacts can be up to ten times stronger than the desired signals, varying with environmental conditions. This study evaluates the variability of these artifacts and the effectiveness of a gradiometer in reducing them in such settings. To achieve these objectives, we first measured the single channel output in varying magnetic field conditions to observe the effect of homogeneous and gradient background fields. Our analysis revealed that the variability in motion artifact within an ambient environment is primarily influenced by the gradient magnetic field rather than the homogeneous one. Subsequently, we configured a gradiometer in parallel and vertical alignment with the direction of vibration (X-axis). Our findings indicated that in a gradient background magnetic field ranging from 1 nT/mm to 10 nT/mm, the single-channel sensor output exhibited a change of 164.97 pT per mm unit increase, while the gradiometer output showed a change of only 0.75 pT/mm within the same range. Upon repositioning the gradiometer vertically (Y direction), perpendicular to the direction of vibration, the single-channel output slope increased to 196.85 pT, whereas the gradiometer output only increased by 1.06 pT/mm for the same range. Our findings highlight the influence of ambient environments on motion artifacts and demonstrate the potential of gradiometers to mitigate these effects. In the future, we plan to record biomagnetic signals both inside and outside the shielded room to compare the efficacy of different gradiometer designs under varying environmental conditions.
引用
收藏
页数:11
相关论文
共 50 条
  • [11] Motion Artifact Reduction Algorithm for Wearable Electrocardiogram Monitoring Systems
    Lee, Shuenn-Yuh
    Su, Po-Han
    Hung, Yi-Wen
    Lee, I-Pei
    Li, Szu-Ju
    Chen, Ju-Yi
    IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2023, 69 (03) : 533 - 547
  • [12] A Wearable Electrooculogram System with Parallel Motion Artifact Sensing and Reduction
    Debbarma, Shibam
    Bhadra, Sharmistha
    2022 IEEE INTERNATIONAL SYMPOSIUM ON CIRCUITS AND SYSTEMS (ISCAS 22), 2022, : 205 - 209
  • [13] Motion Artifact Removal Techniques for Wearable EEG and PPG Sensor Systems
    Seok, Dongyeol
    Lee, Sanghyun
    Kim, Minjae
    Cho, Jaeouk
    Kim, Chul
    FRONTIERS IN ELECTRONICS, 2021, 2
  • [14] Two-Pronged Motion Artifact Reduction for Wearable Photoplethysmographic Biosensors
    Jiang, Yuan
    IEEE SENSORS LETTERS, 2018, 2 (04)
  • [15] A Smart ECG Sensor with In-Situ Adaptive Motion-Artifact Compensation for Dry-Contact Wearable Healthcare Devices
    Zhu, Shuang
    Song, Jingyi
    Chellappa, Balaji
    Enteshari, Ali
    Shan, Tuo
    He, Mengxun
    Chiu, Yun
    PROCEEDINGS OF THE SEVENTEENTH INTERNATIONAL SYMPOSIUM ON QUALITY ELECTRONIC DESIGN ISQED 2016, 2016, : 450 - 455
  • [16] Evaluating the impact of motion artifact on noninvasive blood pressure devices
    Alpert, Bruce S.
    Quinn, David E.
    Friedman, Bruce C.
    Matsumura, Paul M.
    Dart, Richard A.
    Donehoo, Robert F., III
    JOURNAL OF CLINICAL HYPERTENSION, 2020, 22 (04): : 585 - 589
  • [17] Energy Harvesting from Human Motion for Wearable Devices
    Cheng, Qijia
    Peng, Zhuoteng
    Lin, Jie
    Li, Shanshan
    Wang, Fei
    2015 IEEE 10TH INTERNATIONAL CONFERENCE ON NANO/MICRO ENGINEERED AND MOLECULAR SYSTEMS (NEMS), 2015, : 409 - 412
  • [18] Investigation of Motion Artifacts for Biopotential Measurement in Wearable Devices
    Li, Xian
    Hui, Huang
    Sun, Ye
    2016 IEEE 13TH INTERNATIONAL CONFERENCE ON WEARABLE AND IMPLANTABLE BODY SENSOR NETWORKS (BSN), 2016, : 218 - 223
  • [19] Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography
    Han, Hyonyoung
    Kim, Min-Joon
    Kim, Jung
    2007 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-16, 2007, : 1538 - +
  • [20] Wearable EMG Measurement Device Using Polyurethane Foam for Motion Artifact Suppression
    Takagi, Takuma
    Tomita, Naoto
    Sato, Suguru
    Yamamoto, Michitaka
    Takamatsu, Seiichi
    Itoh, Toshihiro
    SENSORS, 2024, 24 (10)