Motion artifact variability in biomagnetic wearable devices

被引:0
|
作者
Arekhloo, Negin Ghahremani [1 ,2 ]
Wang, Huxi [1 ,2 ]
Parvizi, Hossein [2 ]
Tanwear, Asfand [1 ]
Zuo, Siming [1 ]
McKinlay, Michael [2 ]
Nunez, Carlos Garcia [2 ]
Nazarpour, Kianoush [1 ,3 ]
Heidari, Hadi [1 ,2 ]
机构
[1] Neuranics Ltd, Glasgow, Scotland
[2] Univ Glasgow, James Watt Sch Engn, Microelect Lab, Glasgow City, Scotland
[3] Univ Edinburgh, Sch Informat, Edinburgh, Scotland
来源
基金
英国工程与自然科学研究理事会;
关键词
biomagnetic measurements; motion artifacts; wearable sensors; gradient background field; homogeneous background field; SIGNAL; ECG;
D O I
10.3389/fmedt.2024.1457535
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Motion artifacts can be a significant noise source in biomagnetic measurements when magnetic sensors are not separated from the signal source. In ambient environments, motion artifacts can be up to ten times stronger than the desired signals, varying with environmental conditions. This study evaluates the variability of these artifacts and the effectiveness of a gradiometer in reducing them in such settings. To achieve these objectives, we first measured the single channel output in varying magnetic field conditions to observe the effect of homogeneous and gradient background fields. Our analysis revealed that the variability in motion artifact within an ambient environment is primarily influenced by the gradient magnetic field rather than the homogeneous one. Subsequently, we configured a gradiometer in parallel and vertical alignment with the direction of vibration (X-axis). Our findings indicated that in a gradient background magnetic field ranging from 1 nT/mm to 10 nT/mm, the single-channel sensor output exhibited a change of 164.97 pT per mm unit increase, while the gradiometer output showed a change of only 0.75 pT/mm within the same range. Upon repositioning the gradiometer vertically (Y direction), perpendicular to the direction of vibration, the single-channel output slope increased to 196.85 pT, whereas the gradiometer output only increased by 1.06 pT/mm for the same range. Our findings highlight the influence of ambient environments on motion artifacts and demonstrate the potential of gradiometers to mitigate these effects. In the future, we plan to record biomagnetic signals both inside and outside the shielded room to compare the efficacy of different gradiometer designs under varying environmental conditions.
引用
收藏
页数:11
相关论文
共 50 条
  • [21] Motion artifact removal based on periodical property for ECG monitoring with wearable systems
    Zou, Chen
    Qin, Yajie
    Sun, Chenglu
    Li, Wei
    Chen, Wei
    PERVASIVE AND MOBILE COMPUTING, 2017, 40 : 267 - 278
  • [22] Development of Wearable ECG Measurement System Using EMD for Motion Artifact Removal
    Jung, Hae-Kyung
    Jeong, Do-Un
    2012 7TH INTERNATIONAL CONFERENCE ON COMPUTING AND CONVERGENCE TECHNOLOGY (ICCCT2012), 2012, : 299 - 304
  • [23] A Wearable Pulse Oximeter With Wireless Communication and Motion Artifact Tailoring for Continuous Use
    Chacon, Pedro J.
    Pu, Limeng
    da Costa, Tallis H.
    Shin, Young-Ho
    Ghomian, Taher
    Shamkhalichenar, Hamed
    Wu, Hsiao-Chun
    Irving, Brian A.
    Choi, Jin-Woo
    IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 2019, 66 (06) : 1505 - 1513
  • [24] Research on Motion Artifact Elimination Algorithm for Heart Rate Monitoring in Wearable Application
    Du, JiaTao
    Li, XiaoHui
    Nie, ZeDong
    Ding, Yuemin
    Cai, Bin
    He, Jie
    2023 35TH CHINESE CONTROL AND DECISION CONFERENCE, CCDC, 2023, : 5361 - 5366
  • [25] Evaluation of measurement accuracy of wearable devices for heart rate variability
    Li, Xiangchen
    Song, Yuting
    Wang, Huang
    Su, Xinyu
    Wang, Mengyao
    Li, Jing
    Ren, Zhiqiang
    Zhong, Daidi
    Huang, Zhiyong
    ISCIENCE, 2023, 26 (11)
  • [26] Dealing with human variability in motion based, wearable activity recognition
    Kreil, Matthias
    Sick, Bernhard
    Lukowicz, Paul
    2014 IEEE INTERNATIONAL CONFERENCE ON PERVASIVE COMPUTING AND COMMUNICATIONS WORKSHOPS (PERCOM WORKSHOPS), 2014, : 36 - 40
  • [27] A Physiological Information Extraction Method Based on Wearable PPG Sensors with Motion Artifact Removal
    Lin, Wei-Jheng
    Ma, Hsi-Pin
    2016 IEEE INTERNATIONAL CONFERENCE ON COMMUNICATIONS (ICC), 2016,
  • [28] Combining Nonlinear Adaptive Filtering and Signal Decomposition for Motion Artifact Removal in Wearable Photoplethysmography
    Ye, Yalan
    Cheng, Yunfei
    He, Wenwen
    Hou, Mengshu
    Zhang, Zhilin
    IEEE SENSORS JOURNAL, 2016, 16 (19) : 7133 - 7141
  • [29] Active motion artifact reduction for wearable sensors using Laguerre expansion and signal separation
    Wood, Levi B.
    Asada, H. Harry
    2005 27TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2005, : 3571 - 3574
  • [30] A Combined Method to Reduce Motion Artifact and Power Line Interference for Wearable Healthcare Systems
    Hong, Sunjoo
    Song, Kiseok
    Yan, Long
    Yoo, Hoi-Jun
    PROCEEDINGS OF THE 2010 IEEE ASIA PACIFIC CONFERENCE ON CIRCUIT AND SYSTEM (APCCAS), 2010, : 508 - 511