Development of real-time motion artifact reduction algorithm for a wearable photoplethysmography

被引:68
|
作者
Han, Hyonyoung [1 ]
Kim, Min-Joon [2 ]
Kim, Jung [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Mech Engn, Taejon 305701, South Korea
[2] H3 Syst, Daejeon, South Korea
关键词
D O I
10.1109/IEMBS.2007.4352596
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
This paper presents a motion artifact reduction algorithm for a real-time, wireless and wearable photoplethysmography (PPG) device for measuring heart beats. A wearable finger band PPG device consists of a 3-axis accelerometer, infrared LED, photo diode, a microprocessor and wireless module. Sources of the motion artifacts were investigated from the hand motions, through computing the correlations between the three directional finger motions and distorted PPG signals. A two-dimensional active noise cancellation algorithm was applied to compensate the distorted signals by motions, using the directional accelerometer data. NLMS (Normalized Least Mean Square) adaptive filter (4th order) was employed in the algorithm. As a result, the signals' distortion rates were reduced from 52.34% to 3.53%, at frequencies between 1 and 2.5 Hz, which representing daily motions such walking and jogging. The wearable health monitoring device equipped with the motion artifact reduction algorithm can be integrated as a terminal in a so-called ubiquitous healthcare system, which provides a continuous health monitoring without interrupting a daily life.
引用
收藏
页码:1538 / +
页数:2
相关论文
共 50 条
  • [1] Motion Artifact Reduction Algorithm in Wearable Healthcare System
    Lee, Seung Chul
    Kim, Sung Min
    JOURNAL OF MEDICAL DEVICES-TRANSACTIONS OF THE ASME, 2016, 10 (02):
  • [2] Motion Artifact Cancellation in Wearable Photoplethysmography Using Gyroscope
    Lee, H.
    Chung, H.
    Lee, J.
    IEEE SENSORS JOURNAL, 2019, 19 (03) : 1166 - 1175
  • [3] Motion Artifact Reduction Algorithm for Wearable Electrocardiogram Monitoring Systems
    Lee, Shuenn-Yuh
    Su, Po-Han
    Hung, Yi-Wen
    Lee, I-Pei
    Li, Szu-Ju
    Chen, Ju-Yi
    IEEE TRANSACTIONS ON CONSUMER ELECTRONICS, 2023, 69 (03) : 533 - 547
  • [4] Development of a real-time adaptive delta compression algorithm for photoplethysmography system
    Chong, Kim Soon
    Gan, Kok Beng
    Zahedi, Edmond
    IEEJ TRANSACTIONS ON ELECTRICAL AND ELECTRONIC ENGINEERING, 2018, 13 (10) : 1454 - 1460
  • [5] CycleGAN Based Motion Artifact Cancellation for Photoplethysmography Wearable Device
    Nguyen Mai Hoang Long
    Kim, Jong Jin
    Lee, Boon Giin
    Chung, Wan Young
    INTELLIGENT HUMAN COMPUTER INTERACTION, IHCI 2021, 2022, 13184 : 138 - 144
  • [6] Motion Artifact Reduction in Wearable Photoplethysmography Based on Multi-Channel Sensors with Multiple Wavelengths
    Lee, Jongshill
    Kim, Minseong
    Park, Hoon-Ki
    Kim, In Young
    SENSORS, 2020, 20 (05)
  • [7] A MOTION ARTIFACT IN REAL-TIME ULTRASOUND SCANNERS
    MORRISON, DC
    MCDICKEN, WN
    SMITH, DSA
    ULTRASOUND IN MEDICINE AND BIOLOGY, 1983, 9 (02): : 201 - 203
  • [8] SELECTIVE IR MR-ANGIOGRAPHY WITH REAL-TIME MOTION ARTIFACT REDUCTION
    WANG, SJ
    SACHS, TS
    LI, KC
    MACOVSKI, AE
    NISHIMURA, DG
    RADIOLOGY, 1995, 197 : 187 - 187
  • [9] Motion Artifact Reduction In Photoplethysmography For Reliable Signal Selection
    Mao, Runyu
    Tweardy, Mackenzie
    Wegerich, Stephan W.
    Goergen, Craig J.
    Wodicka, George R.
    Zhu, Fengqing
    2021 43RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE & BIOLOGY SOCIETY (EMBC), 2021, : 5625 - 5630
  • [10] Development of a Wearable Motion Detector for Telemonitoring and Real-Time Identification of Physical Activity
    Yang, Che-Chang
    Hsu, Yeh-Liang
    TELEMEDICINE JOURNAL AND E-HEALTH, 2009, 15 (01): : 62 - 72