Effect of a prototype 2-dimensional antiscatter grid on image quality obtained with a dental cone-beam computed tomography scanner

被引:0
|
作者
Li, Boyuan [1 ,2 ]
Somerkivi, Villeseveri [3 ,4 ,5 ]
Bayat, Farhang [6 ]
Huynh, Carolyn [7 ]
Altunbas, Cem [6 ]
机构
[1] M2 Technol, Denver, CO USA
[2] Stanford Univ, Dept Radiol, Stanford, CA USA
[3] Planmeca, Helsinki, Finland
[4] Tech Univ Munich, Sch Nat Sci, Dept Phys, Garching, Germany
[5] Tech Univ Munich, Munich Inst Biomed Engn, Garching, Germany
[6] Univ Colorado, Dept Radiat Oncol, Sch Med, 1665 Aurora Court Suite 1032, Aurora, CO 80045 USA
[7] UTHealth Houston, Dept Diagnost & Biomed Sci, Sch Dent, Houston, TX USA
关键词
Cone-Beam Computed Tomography; X-rays; Diagnostic Imaging; Image quality enhancement; SCATTER CORRECTION METHODS; GENERAL FRAMEWORK; CT;
D O I
10.5624/isd.20240182
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: X-ray scattering adversely affects cone-beam computed tomography (CBCT) image quality, generating image artifacts, causing inaccurate tissue density representation, and reducing contrast. This study evaluated the performance of a 2-dimensional antiscatter grid (2D grid) prototype in a dental CBCT system. Materials and Methods: A focused 2D grid prototype was fabricated from tungsten and integrated with the detector of a dental CBCT system. Residual scatter transmitted through the 2D grid was corrected using a measurement-based scatter correction method. Phantom imaging experiments were performed in anatomical regions relevant to dental and head imaging with and without this grid. Following image reconstruction via filtered back projection, attenuation coefficients were converted to Hounsfield units (HU). Subsequently, scatter suppression performance, HU consistency, image artifacts, and contrast resolution were evaluated. Results: The 2D grid reduced scatter intensity by a factor of 10-20 in CBCT projections. Consequently, the grid substantially increased contrast, reduced image artifacts, and improved HU consistency. The contrast increased by 27% and 48% in bone- and soft tissue-equivalent regions, respectively. HU value deviations among teeth decreased from 510 to 146 HU. These results indicate improved visualization and tissue density representation fidelity in CBCT images acquired with the 2D grid. Conclusion: Use of a 2D grid could substantially improve the accuracy of tissue density representation and the contrast of dental and head anatomy in 3-dimensional images obtained with dental CBCT. Such improvements may translate to better quantitative evaluation of bone quality, enhanced tissue visualization, and more accurate model generation for surgical planning and guidance. (Imaging Sci Dent 20240182)
引用
收藏
页数:11
相关论文
共 50 条
  • [1] Evaluation of a two-dimensional Moire-free antiscatter grid for cone-beam computed tomography
    Kim, Jongkuk
    Kang, Yoonseok
    Hwang, Taesoon
    Park, Myungkyu
    Chung, Woohyun
    MEDICAL PHYSICS, 2023, 50 (06) : 3435 - 3444
  • [2] Two-dimensional antiscatter grid: A novel scatter rejection device for Cone-beam computed tomography
    Alexeev, Timur
    Kavanagh, Brian
    Miften, Moyed
    Altunbas, Cem
    MEDICAL PHYSICS, 2018, 45 (02) : 529 - 534
  • [3] Quality Control in Dental Cone-Beam Computed Tomography
    Huang, Jin-Feng
    Chen, Xiao-Zhao
    Wang, Hong
    APPLIED SCIENCES-BASEL, 2021, 11 (17):
  • [4] Removing grid artifacts from a two-dimensional crisscrossed grid in cone-beam computed tomography to enhance image quality
    Jeon, Duhee
    Kim, Woosung
    Cho, Hyosung
    Park, Myeongkyu
    JOURNAL OF INSTRUMENTATION, 2024, 19 (03)
  • [5] Investigation of effect of reconstruction filters on cone-beam computed tomography image quality
    Srinivasan, Kavitha
    Mohammadi, Mohammad
    Shepherd, Justin
    AUSTRALASIAN PHYSICAL & ENGINEERING SCIENCES IN MEDICINE, 2014, 37 (03) : 607 - 614
  • [6] Investigation of effect of reconstruction filters on cone-beam computed tomography image quality
    Kavitha Srinivasan
    Mohammad Mohammadi
    Justin Shepherd
    Australasian Physical & Engineering Sciences in Medicine, 2014, 37 : 607 - 614
  • [7] Evaluation of radiation dose and image quality for dental cone-beam computed tomography in pediatric patients
    Ito, Misaki
    Chida, Koichi
    Onodera, Shu
    Kojima, Ikuho
    Iikubo, Masahiro
    Kato, Toshiki
    Fujisawa, Masaki
    Zuguchi, Masayuki
    JOURNAL OF RADIOLOGICAL PROTECTION, 2023, 43 (03)
  • [8] A study on image quality provided by a kilovoltage cone-beam computed tomography
    Garayoa, Julia
    Castro, Pablo
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2013, 14 (01): : 239 - 257
  • [9] Image quality produced by different cone-beam computed tomography settings
    Kwong, Jeffrey C.
    Palomo, J. Martin
    Landers, Michael A.
    Figueroa, Alex
    Hans, Mark G.
    AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 2008, 133 (02) : 317 - 327
  • [10] TransCBCT: Improving the Image Quality of Cone-Beam Computed Tomography with Transformer
    Chen, X.
    Liu, Y.
    Yang, B.
    Zhu, J.
    Liu, Y.
    Dai, J.
    Men, K.
    MEDICAL PHYSICS, 2022, 49 (06) : E791 - E791