Fatigue lifetime of reduced-diameter implants placed in different bone models

被引:0
|
作者
Tretto, Pedro Henrique Wentz [1 ]
Griggs, Jason Alan [2 ]
Jodha, Kartikeya Singh [2 ]
Duan, Yuanyuan [2 ]
Marocho, Susana Maria Salazar [2 ]
Della Bona, Alvaro [1 ]
机构
[1] Univ Passo Fundo, Sch Dent, Postgrad Program Dent, Campus I,BR 285, BR-99001970 Passo Fundo, RS, Brazil
[2] Univ Mississippi Med Ctr, Biomed Mat Sci, 2500 North State St, Jackson, MS 39216 USA
关键词
Dental implants; Fatigue fractures; Fractographic analysis; COMPUTED-TOMOGRAPHY; VOLUME FRACTION; PARAMETERS; STIFFNESS; FRACTURE;
D O I
10.1016/j.dental.2024.09.003
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Objectives: This study assessed the fatigue lifetime of reduced-diameter implants placed in either bovine rib or polymer-based bone model. Methods: Bovine ribs were classified according to the criteria proposed by Lekholm and Zarb and were analyzed for bone fraction. Fourteen dental implants (3.25 mm in diameter x 15 mm in length) were placed in bovine ribs used as a bone model. They were subjected to resonance frequency analysis. Stainless steel loading hemisphere caps were bonded on the abutments position at 30-degree angle and with a moment arm of 11 mm. Accelerated life testing using the step-stress method was conducted at 2 Hz with a stress ratio of 0.1 until fracture on a servohydraulic load frames machine (MTS). Results were compared with those of a previous study wherein implants were placed in a polymer-based bone model. Fatigue lifetime statistics (characteristic lifetime and Weibull modulus) of physical specimens were estimated in a reliability analysis software (ALTA PRO). Fractured specimens were examined under an electron scanning microscope to determine the failure mode. Result: The implants exhibited high stability quotient values (75.07 +/- 3.81). Implants placed in bovine ribs showed better data dispersion and longer fatigue lifetime than those placed in polymer-based bone models, with no significant difference between groups. All fractures occurred in the implant body near the bone level and were indicative of fatigue fractures. Significance: Bovine ribs appear to be a more suitable material for accelerated life testing than the polymer-based material because of better data dispersion.
引用
收藏
页码:1920 / 1926
页数:7
相关论文
共 50 条
  • [1] Fatigue lifetime prediction of a reduced-diameter dental implant system: Numerical and experimental study
    Duan, Yuanyuan
    Gonzalez, Jorge A.
    Kulkarni, Pratim A.
    Nagy, William W.
    Griggs, Jason A.
    DENTAL MATERIALS, 2018, 34 (09) : 1299 - 1309
  • [2] Fracture of Reduced-Diameter Zirconia Dental Implants Following Repeated Insertion
    Karl, Matthias
    Scherg, Stefan
    Grobecker-Karl, Tanja
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2017, 32 (05) : 971 - 975
  • [3] Primary Stability of Implants Placed at Different Angulations in Artificial Bone
    Kashi, Ajay
    Gupta, Bhumija
    Malmstrom, Hans
    Romanos, Georgios E.
    IMPLANT DENTISTRY, 2015, 24 (01) : 92 - 95
  • [4] Healing at implants placed in bone of different morphology: an experimental study in dogs
    Caroprese, Marino
    Soldini, Claudio
    Ricci, Sara
    Rea, Massimiliano
    Lang, Niklaus P.
    Botticelli, Daniele
    CLINICAL ORAL IMPLANTS RESEARCH, 2017, 28 (08) : 961 - 965
  • [5] Longevity of different abutments placed on narrow diameter implants: Assessment of structural damage and loosening
    Lenz, Ulysses
    Santos, Renan Brandenburg
    Griggs, Jason Alan
    Estrela, Carlos
    dos Reis Bueno, Mike
    Porto, Olavo Cesar Lyra
    Della Bona, Alvaro
    DENTAL MATERIALS, 2024, 40 (09) : 1332 - 1340
  • [6] In Vitro Effect of Drilling Speed on the Primary Stability of Narrow Diameter Implants with Varying Thread Designs Placed in Different Qualities of Simulated Bone
    Romanos, Georgios E.
    Bastardi, Daniel J.
    Moore, Rachel
    Kakar, Apoorv
    Herin, Yaro
    Delgado-Ruiz, Rafael A.
    MATERIALS, 2019, 12 (08)
  • [7] Investigation of Phase Transformation and Fracture Pattern as a Result of Long-Term Chewing Simulation and Static Loading of Reduced-Diameter Zirconia Implants
    Atalay Seckiner, Pelin
    Gonuldas, Fehmi
    Akat, Bora
    Buyuksungur, Arda
    Orhan, Kaan
    MATERIALS, 2024, 17 (19)
  • [8] Retrospective Evaluation of Crestal Bone Changes Around Implants With Reduced Abutment Diameter Placed Non-Submerged and at Subcrestal Positions: The Effect of Bone Grafting at Implant Placement
    Koutouzis, Theofilos
    Fetner, Michael
    Fetner, Alan
    Lundgren, Tord
    JOURNAL OF PERIODONTOLOGY, 2011, 82 (02) : 234 - 242
  • [9] Immediate Versus Non-immediate Loading Protocols for Reduced-Diameter Implants Supporting Overdentures: A Systematic Review and Meta-analysis
    Liu, Ying
    He, Fengxiao
    Zhao, Yaoyu
    Sun, Quan
    Xia, Haibin
    Xia, Dahong
    Bai, Yi
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2024, 39 (05)
  • [10] Comparative study of wide-diameter implants placed after dental extraction and implants positioned in mature bone for molar replacement
    Penarrocha-Diago, Miguel
    Carrillo-Garcia, Celia
    Boronat-Lopez, Araceli
    Garcia-Mira, Berta
    INTERNATIONAL JOURNAL OF ORAL & MAXILLOFACIAL IMPLANTS, 2008, 23 (03) : 497 - 501