Anatomy-based fitting improves speech perception in noise for cochlear implant recipients with single-sided deafness

被引:0
|
作者
Kurz, Anja [1 ]
Herrmann, David [1 ]
Mueller-Graff, Franz-Tassilo [1 ]
Voelker, Johannes [1 ]
Hackenberg, Stephan [1 ]
Rak, Kristen [1 ]
机构
[1] Univ Hosp Wurzburg, Comprehens Hearing Ctr, Dept Oto Rhino Laryngol Head & Neck Surg, Josef Schneider Str 11, D-97080 Wurzburg, Germany
关键词
Cochlear implant in single sided deafness; Anatomy-based fitting; Frequency-to-place mismatch; Speech perception; SPIRAL GANGLION; PITCH;
D O I
10.1007/s00405-024-08984-4
中图分类号
R76 [耳鼻咽喉科学];
学科分类号
100213 ;
摘要
ObjectiveTo evaluate objective and subjective hearing outcomes in experienced cochlear implant users with single sided deafness (SSD CI) who used fitting maps created via anatomy-based fitting (ABF) and clinically-based fitting (CBF).ParticipantsTwelve SSD CI users with postlingual hearing loss.InterventionOTOPLAN (Version 3. (MED-EL) was used to determine intracochlear electrode contact positions using post-operative high-resolution flat panel volume computed tomography. From these positions, the corresponding center frequencies and bandwidths were derived for each channel. These were implemented in the clinical fitting software MAESTRO to yield an ABF map individualized to each user.Main Outcome MeasuresABF and CBF maps were compared. Objective speech perception in quiet and in noise, binaural effects, and self-perceived sound quality were evaluated.ResultsSignificantly higher speech perception in noise scores were observed with the ABF map compared to the CBF map (mean SRT50: -6.49 vs. -4.8 dB SNR for the S0NCI configuration and - 3.85 vs. -2.75 dB SNR for the S0N0 configuration). Summation and squelch effects were significantly increased with the ABF map (0.86 vs. 0.21 dB SNR for summation and 0.85 vs. -0.09 dB SNR for squelch). No improvement in speech perception in quiet or spatial release from masking were observed with the ABF map. A similar level of self-perceived sound quality was reported for each map. Upon the end of the study, all users opted to keep the ABF map. This preference was independent of the angular insertion depth of the electrode array.ConclusionsExperienced SSD CI users preferred using the ABF map, which gave them significant improvements in binaural hearing and some aspects of speech perception.
引用
收藏
页码:467 / 479
页数:13
相关论文
共 50 条
  • [21] Can anatomy-based fitting improve musical perception in adult cochlear implant users?
    Lassaletta, Luis
    Calvino, Miryam
    Sanchez-Cuadrado, Isabel
    Munoz, Elena
    Gavilan, Javier
    BRAZILIAN JOURNAL OF OTORHINOLARYNGOLOGY, 2025, 91 (02)
  • [22] Masked Speech Recognition as a Function of Masker Location for Cochlear Implant Users With Single-Sided Deafness
    Anderson, Kelly M.
    Buss, Emily
    Rooth, Meredith A.
    Richter, Margaret E.
    Overton, Andrea B.
    Brown, Kevin D.
    Dillon, Margaret T.
    AMERICAN JOURNAL OF AUDIOLOGY, 2022, 31 (03) : 757 - 763
  • [23] Speech Perception Outcomes with the Anatomy-Based Fitting Map among Experienced, Adult Cochlear Implant Users: A Longitudinal Study
    Wangchuk, Pelden
    Umat, Cila
    Chong, Foong Yen
    Zaki, Faizah Mohd
    Abdullah, Asma
    AUDIOLOGY AND NEUROTOLOGY, 2024,
  • [24] Speech Recognition in Noise in Single-Sided Deaf Cochlear Implant Recipients Using Digital Remote Wireless Microphone Technology
    Wesarg, Thomas
    Arndt, Susan
    Wiebe, Konstantin
    Schmid, Frauke
    Huber, Annika
    Mulder, Hans E.
    Laszig, Roland
    Aschendorff, Antje
    Speck, Iva
    JOURNAL OF THE AMERICAN ACADEMY OF AUDIOLOGY, 2019, 30 (07) : 607 - 618
  • [25] Using aided cortical assessment as an objective tool to evaluate cochlear implant fitting in users with single-sided deafness
    Tavora-Vieira, Dayse
    Wedekind, Andre
    Marino, Roberta
    Purdy, Suzanne C.
    Rajan, Gunesh P.
    PLOS ONE, 2018, 13 (02):
  • [26] From the expert opinion practice: Cochlear implant (CI) versus CROS fitting for single-sided deafness (SSD)
    Brusis, Tilman
    Schmidt, Sandra
    LARYNGO-RHINO-OTOLOGIE, 2024, 103 (09) : 665 - 667
  • [27] Influence of cochlear coverage on speech perception in single sided deafness, bimodal, and bilateral implanted cochlear implant patients
    Spiegel, Jennifer L.
    Mueller, Joachim
    Boehnlein, Rebecca
    Hempel, John-Martin
    Spiro, Judith E.
    Weiss, Bernhard G.
    Bertlich, Mattis
    Canis, Martin
    Rader, Tobias
    EUROPEAN ARCHIVES OF OTO-RHINO-LARYNGOLOGY, 2024, : 1853 - 1863
  • [28] Speech Perception Performance Growth and Benchmark Score Achievement After Cochlear Implantation for Single-Sided Deafness
    Nassiri, Ashley M.
    Wallerius, Katherine P.
    Lohse, Christine M.
    Marinelli, John P.
    Saoji, Aniket A.
    Driscoll, Colin L. W.
    Neff, Brian A.
    Carlson, Matthew L.
    OTOLOGY & NEUROTOLOGY, 2022, 43 (01) : E64 - E71
  • [29] Spatial Speech-in-Noise Performance in Bimodal and Single-Sided Deaf Cochlear Implant Users
    Williges, Ben
    Wesarg, Thomas
    Jung, Lorenz
    Geven, Leontien I.
    Radeloff, Andreas
    Juergens, Tim
    TRENDS IN HEARING, 2019, 23
  • [30] Restoration of spatial hearing in adult cochlear implant users with single-sided deafness
    Litovsky, Ruth Y.
    Moua, Keng
    Godar, Shelly
    Kan, Alan
    Misurelli, Sara M.
    Lee, Daniel J.
    HEARING RESEARCH, 2019, 372 : 69 - 79