AN EVALUATION OF THE SYSMEX NE-8000 HEMATOLOGY ANALYZER

被引:14
|
作者
HALLAWELL, R
OMALLEY, C
HUSSEIN, S
DAUER, RJ
TANTI, M
WOOTTON, AM
MCGRATH, KM
机构
[1] ROYAL MELBOURNE HOSP,DEPT DIAGNOST HEMATOL,C-RMH PO,PARKVILLE,VIC 3050,AUSTRALIA
[2] ROYAL MELBOURNE HOSP,DEPT BIOCHEM,PARKVILLE,VIC 3050,AUSTRALIA
关键词
AUTOMATED LEUKOCYTE 5-PART DIFFERENTIAL COUNT; SYSMEX NE-8000; COULTER COUNTER-S PLUS-IV; NCCLS REFERENCE DIFFERENTIAL METHOD;
D O I
10.1093/ajcp/96.5.594
中图分类号
R36 [病理学];
学科分类号
100104 ;
摘要
The Sysmex NE-8000(R) is a new, fully automated hematology analyzer capable of providing a five-part white blood cell differential count and identifying abnormal specimens. This instrument was evaluated on 5,000 consecutive blood specimens and compared to the Coulter S Plus-IV(R) analyzer and manual differential cell counts to determine the efficacy of its five-cell differential and screening capabilities. There was a high correlation between the commercial counters for the standard parameters, white blood cell count, red blood cell count, hemoglobin level, hematocrit, mean corpuscular volume, mean corpuscular hemoglobin concentration, and platelet count (r > 0.95), except for the mean corpuscular hemoglobin concentration (r = 0.51), for which the NE-8000 was considered the more accurate measurement. Precision and linearity studies were excellent. The white blood cell count, red blood cell count, hemoglobin level, and platelet count were reproducible on specimens stored at 4-degrees-C or room temperature for 72 hours and the differential counts were reproducible for 12 hours. The correlations between automated and manual counts for neutrophils, eosinophils, basophils, and lymphocytes were excellent: r = 0.912, 0.945, 0.332, and 0.964, respectively. The monocyte correlation improved with software modification from 0.306 to 0.801. The NE-8000 gave accurate and reproducible differential counts for neutrophils, lymphocytes, and monocytes on specimens with white blood cell counts as low as 0.6 X 10(9)/1. The ability of the instrument to 'flag' abnormal specimens was excellent. The false-positive rate on normal samples was 1.8%, and the false-negative rate on known abnormal samples was 0.3%, due only to nonrecognition of a mild left shift. The identification of specific abnormalities was less precise. The NE-8000 is a powerful hematology analyzer that can perform a five-part white blood cell differential count accurately for a wide range of WBCs and reliably indicate abnormal specimens. It is an excellent screening tool for distinguishing between normal and abnormal specimens and identifying those that require microscopy. Its reliability significantly reduces the need for manual film examination.
引用
收藏
页码:594 / 601
页数:8
相关论文
共 50 条
  • [1] EVALUATION OF THE AUTOMATED HEMATOLOGY ANALYZER SYSMEX NE-8000
    DEVREESE, K
    DELOGI, E
    FRANCART, C
    HEYNDRICKX, B
    PHILIPPE, J
    LEROUXROELS, G
    EUROPEAN JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1991, 29 (05): : 339 - 345
  • [2] AUTOMATED DIFFERENTIAL LEUKOCYTE COUNTING ON THE SYSMEX NE-8000 ANALYZER
    SIVAKUMARAN, M
    ALLEN, B
    WOOD, JK
    CLINICAL AND LABORATORY HAEMATOLOGY, 1994, 16 (02): : 206 - 207
  • [3] The characterisation of leukaemias with the Sysmex NE-8000
    vanderMeer, W
    Swinkels, DW
    Willems, HL
    ACTA HAEMATOLOGICA, 1997, 98 (04) : 195 - 198
  • [4] EVALUATION OF A NEW HEMATOLOGY ANALYZER FOR WHOLE-BLOOD COUNT AND FULL DIFFERENTIAL (NE-8000)
    LAHARRAGUE, PF
    FILLOLA, G
    CORBERAND, JX
    NOUVELLE REVUE FRANCAISE D HEMATOLOGIE, 1992, 34 (04): : 303 - 307
  • [5] Time dependent increase of differential monocyte count on the Sysmex NE-8000
    Bartels, PCM
    Schoorl, M
    CLINICAL AND LABORATORY HAEMATOLOGY, 1998, 20 (03): : 165 - 168
  • [6] EVALUATION OF PERFORMANCE FOR AUTOMATED DIFFERENTIAL LEUKOCYTE COUNTING ON SYSMEX NE-8000 BY NCCLS RECOMMENDED PROTOCOL, H20-T
    HU, CY
    WANG, CH
    CHUANG, HM
    SHEN, MC
    CLINICAL AND LABORATORY HAEMATOLOGY, 1993, 15 (04): : 287 - 299
  • [7] PERFORMANCE EVALUATION OF SYSMEX XN-550 AUTOMATIC HEMATOLOGY ANALYZER - COMPARATIVE STUDY WITH HEMATOLOGY ANALYZER SYSMEX XN-10
    Yan, Jie
    Ortiz, Ashley
    DeMarco, Julie
    Zhang, Qian-Yun
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 : 129 - 129
  • [8] EVALUATION OF THE AUTOMATED HEMATOLOGY ANALYZER SYSMEX M-2000
    POHLAND, D
    JOURNAL OF CLINICAL CHEMISTRY AND CLINICAL BIOCHEMISTRY, 1989, 27 (01): : 41 - 47
  • [9] EVALUATION OF THE AUTOMATED TOA SYSMEX CC800 HEMATOLOGY ANALYZER
    MUNOZ, LE
    REIMER, SM
    GAMBARDELLA, RL
    CLINICAL CHEMISTRY, 1984, 30 (06) : 952 - 952
  • [10] Evaluation of the Sysmex XE-2100® hematology analyzer in hospital use
    Nakul-Aquaronne, D
    Sudaka-Sammarcelli, I
    Ferrero-Vacher, C
    Starck, B
    Bayle, J
    JOURNAL OF CLINICAL LABORATORY ANALYSIS, 2003, 17 (04) : 113 - 123