Rapid and simultaneous determination of 25-OH-vitamin D-2 and D-3 in human serum by LC/MS/MS: Validation and uncertainty assessment

被引:0
|
作者
Goren, Ahmet C. [1 ]
Bilsel, Gokhan [1 ]
Bilsel, Mine [1 ]
机构
[1] TUBITAK, UME, P Box 54, TR-41470 Gebze, Turkey
来源
JOURNAL OF CHEMICAL METROLOGY | 2007年 / 1卷 / 01期
关键词
25-OH Vitamin D-3; 25-OH Vitamin D-2; LC/MS/MS; method validation; uncertainty;
D O I
暂无
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
25-OH Vitamin D-3 and 25-OH Vitamin D-2 were determined by rapid, simple and validated LC/MS/MS method without using SPE and isotopic standard. The relative standard deviations (RSD) were found to be 3.1 and 2.8 % for 25-OH Vitamin D-3 and 25-OH Vitamin D-2, respectively. The correlation coefficient was grater than 0.99 for both analytes in the calibration range. The recoveries at 100 mu g L-1 concentration averaged to 92 % for 25-OH vitamin D-3 and 94 % for 25-OH-vitamin D-2. The uncertainty of the measurements for 25-OH vitamin D-3 and for 25-OH-vitamin D-2 was 7.77 % and 7.24 %, respectively.
引用
收藏
页码:1 / 9
页数:9
相关论文
共 50 条
  • [1] Analytical measurement of serum 25-OH-vitamin D3, 25-OH-vitamin D2 and their C3-epimers by LC-MS/MS in infant and pediatric specimens
    Yazdanpanah, Mehrdad
    Bailey, Dana
    Walsh, Warren
    Wan, Betty
    Adeli, Khosrow
    CLINICAL BIOCHEMISTRY, 2013, 46 (13-14) : 1264 - 1271
  • [2] LC-MS/MS Method for Simultaneous Measurement of Vitamins D2, D3, 25(OH) D2, and 25(OH) D3 in Human Serum
    Burson, Michael A.
    Puhl, Jennifer
    Bodor, Geza S.
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 2009, 132 (03) : 469 - 469
  • [3] Rapid and reliable 25-OH vitamin D2 and 25-OH vitamin D3 measurements by multitasker LC-MS/MS
    Goren, Ahmet C.
    Sahin, Gozde
    Gumilcineli, Ismail
    Binici, Burcu
    JOURNAL OF CHEMICAL METROLOGY, 2018, 12 (01): : 17 - 25
  • [4] Performance evaluating of Abbott 25-OH-vitamin D assay: comparison with HPLC and LC-MS/MS systems
    Avci, E.
    Nar, R.
    Demir, S.
    Senol, H.
    CLINICA CHIMICA ACTA, 2019, 493 : S514 - S515
  • [5] A fast and simple method for simultaneous measurements of 25(OH)D, 24, 25(OH)2D and the Vitamin D Metabolite Ratio (VMR) in serum samples by LC-MS/MS
    Fabregat-Cabello, Neus
    Farre-Segura, Jordi
    Huyghebaert, Loreen
    Peeters, Stephanie
    Le Goff, Caroline
    Souberbielle, Jean-Claude
    Cavalier, Etienne
    CLINICA CHIMICA ACTA, 2017, 473 : 116 - 123
  • [6] ANALYSIS OF 25-HYDROXY VITAMIN D2 AND D3 IN HUMAN SERUM BY 2 DIMENSIONAL LC-MS/MS
    Ray, J. A.
    Kushnir, M. M.
    Rockwood, A. L.
    Meikle, A. W.
    JOURNAL OF INVESTIGATIVE MEDICINE, 2010, 58 (01) : 212 - 212
  • [7] Ultra-Fast LC-MS-MS Analysis of 25-OH Vitamin D2 and D3 from Serum
    Koerner, Philip J.
    LC GC NORTH AMERICA, 2011, : 20 - 20
  • [8] Simultaneous analysis of 25OHD3 and 24,25(OH)2D3 both in human serum and cerebrospinal fluid by LC-MS/MS
    He, Xin
    Jiang, Pei
    Xue, Ying
    Zhu, Wen-Ye
    Deng, Yang
    Yan, Miao
    Li, Huan-De
    Dang, Rui-Li
    Tang, Mi-Mi
    ANALYTICAL METHODS, 2016, 8 (11) : 2400 - 2407
  • [9] Determination of vitamin D metabolites in serum using LC-MS and LC-MS/MS
    Bedner, Mary
    Phinney, Karen W.
    Sander, Lane C.
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 2009, 237 : 301 - 301
  • [10] 25-OH Vitamin D concentrations measured by LC-MS/MS are equivalent in serum and EDTA plasma
    van der Vorm, Lisa N.
    Le Goff, Caroline
    Peeters, Stephanie
    Makris, Konstantinos
    Cavalier, Etienne
    Heijboer, Annemieke C.
    STEROIDS, 2022, 187