Optimum Radius Size between Cylindrical Ion Trap and Quadrupole Ion Trap

被引:3
|
作者
Chaharborj, Sarkhosh Seddighi [1 ,2 ,3 ]
Kiai, Seyyed Mahmod Sadat [3 ]
Arifin, Norihan Md [1 ]
Gheisari, Yousof [2 ]
机构
[1] Univ Purta Malaysia UPM, Fac Sci, Dept Math, Serdang 43400, Malaysia
[2] Islamic Azad Univ, Dept Math, Bushehr Branch, Bushehr, Iran
[3] Nucl Sci & Technol Res Inst NSTRI, Plasma Phys & Nucl Fus Res Sch, POB 14395, Tehran, Iran
关键词
Quadrupole ion trap; cylindrical ion trap; optimum radius size; fifth order Runge Kutta method; stability regions; ion trajectory; fractional mass resolution;
D O I
10.5478/MSL.2015.6.3.59
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Quadrupole ion trap mass analyzer with a simplified geometry, namely, the cylindrical ion trap (CIT), has been shown to be well-suited using in miniature mass spectrometry and even in mass spectrometer arrays. Computation of stability regions is of particular importance in designing and assembling an ion trap. However, solving CIT equations are rather more difficult and complex than QIT equations, so, analytical and matrix methods have been widely used to calculate the stability regions. In this article we present the results of numerical simulations of the physical properties and the fractional mass resolutions of the confined ions in the first stability region was analyzed by the fifth order Runge-Kutta method (RKM5) at the optimum radius size for both ion traps. Because of similarity the both results, having determining the optimum radius, we can make much easier to design CIT. Also, the simulated results has been performed a high precision in the resolution of trapped ions at the optimum radius size.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 50 条
  • [31] Ion excitation in a linear quadrupole ion trap with an added octopole field
    Michaud, AL
    Frank, AJ
    Ding, C
    Zhao, XZ
    Douglas, DJ
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2005, 16 (06) : 835 - 849
  • [32] Advances in Ion Dissociation within Quadrupole Ion Trap Mass Spectrometry
    Chu, Shi-Ying
    Xie, Jie
    Jiang, You
    Wang, Fang-Jun
    Zhai, Rui
    Feng, Lu-Lu
    Gong, Xiao-Yun
    Dai, Xin-Hua
    Tian, Di
    Fang, Xiang
    Qiu, Chun-Ling
    Huang, Ze-Jian
    Journal of Chinese Mass Spectrometry Society, 2019, 40 (04): : 391 - 400
  • [33] RECOIL ION CONFINEMENT IN A RADIO-FREQUENCY QUADRUPOLE ION TRAP
    CHURCH, DA
    HOLZSCHEITER, HM
    KENEFICK, RA
    SHORT, CSORT
    BERRY, SD
    ELSTON, SB
    BREINIG, M
    DESERIO, R
    SELLIN, IA
    THOMAS, B
    APPLIED PHYSICS LETTERS, 1985, 47 (03) : 330 - 332
  • [34] Nonlinear resonance effects during ion storage in a quadrupole ion trap
    1600, Publ by Elsevier Science Publ Co Inc, New York, NY, USA (04):
  • [35] ION REMEASUREMENT IN THE RADIO-FREQUENCY QUADRUPOLE ION-TRAP
    GOERINGER, DE
    CRUTCHER, RI
    MCLUCKEY, SA
    ANALYTICAL CHEMISTRY, 1995, 67 (22) : 4164 - 4169
  • [36] Design and simulation of a wire quadrupole ion trap for ion trapping and cooling
    Li, Ze-Kai
    Zhou, Yijie
    Dong, Changwu
    Hou, Gao-Lei
    CHINESE JOURNAL OF CHEMICAL PHYSICS, 2024, 37 (02) : 137 - 146
  • [37] Miniature cylindrical ion trap mass spectrometer
    Patterson, GE
    Guymon, AJ
    Riter, LS
    Everly, M
    Griep-Raming, J
    Laughlin, BC
    Zheng, OY
    Cooks, RG
    ANALYTICAL CHEMISTRY, 2002, 74 (24) : 6145 - 6153
  • [38] A parallel miniature cylindrical ion trap array
    Badman, ER
    Cooks, RG
    ANALYTICAL CHEMISTRY, 2000, 72 (14) : 3291 - 3297
  • [39] Miniaturization of cylindrical ion trap mass analyzers
    Zhang, Xiangnan
    Zhang, Hong
    Yu, Kai
    Liu, Yuning
    He, Jing
    Jiang, Jie
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2020, 455
  • [40] Gas-phase ion/ion interactions between peptides or proteins and iron ions in a quadrupole ion trap
    Payne, AH
    Glish, GL
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 2001, 204 (1-3) : 47 - 54