Optimum Radius Size between Cylindrical Ion Trap and Quadrupole Ion Trap

被引:3
|
作者
Chaharborj, Sarkhosh Seddighi [1 ,2 ,3 ]
Kiai, Seyyed Mahmod Sadat [3 ]
Arifin, Norihan Md [1 ]
Gheisari, Yousof [2 ]
机构
[1] Univ Purta Malaysia UPM, Fac Sci, Dept Math, Serdang 43400, Malaysia
[2] Islamic Azad Univ, Dept Math, Bushehr Branch, Bushehr, Iran
[3] Nucl Sci & Technol Res Inst NSTRI, Plasma Phys & Nucl Fus Res Sch, POB 14395, Tehran, Iran
关键词
Quadrupole ion trap; cylindrical ion trap; optimum radius size; fifth order Runge Kutta method; stability regions; ion trajectory; fractional mass resolution;
D O I
10.5478/MSL.2015.6.3.59
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
Quadrupole ion trap mass analyzer with a simplified geometry, namely, the cylindrical ion trap (CIT), has been shown to be well-suited using in miniature mass spectrometry and even in mass spectrometer arrays. Computation of stability regions is of particular importance in designing and assembling an ion trap. However, solving CIT equations are rather more difficult and complex than QIT equations, so, analytical and matrix methods have been widely used to calculate the stability regions. In this article we present the results of numerical simulations of the physical properties and the fractional mass resolutions of the confined ions in the first stability region was analyzed by the fifth order Runge-Kutta method (RKM5) at the optimum radius size for both ion traps. Because of similarity the both results, having determining the optimum radius, we can make much easier to design CIT. Also, the simulated results has been performed a high precision in the resolution of trapped ions at the optimum radius size.
引用
收藏
页码:59 / 64
页数:6
相关论文
共 50 条
  • [21] Determination of cooling rates in a quadrupole ion trap
    Black, David M.
    Payne, Anne H.
    Glish, Gary L.
    JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 2006, 17 (07) : 932 - 938
  • [22] A dual quadrupole ion trap mass spectrometer
    Zerega, Y
    Perrier, P
    Carette, M
    Brincourt, G
    Nguema, T
    Andre, J
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY, 1999, 191 : 59 - 68
  • [23] The application of quadrupole ion trap in environmental research
    March, RE
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1999, 217 : U132 - U132
  • [24] QUADRUPOLE ION TRAP MASS-SPECTROMETRY
    COOKS, RG
    KAISER, RE
    ACCOUNTS OF CHEMICAL RESEARCH, 1990, 23 (07) : 213 - 219
  • [25] An introduction to quadrupole ion trap mass spectrometry
    March, RE
    JOURNAL OF MASS SPECTROMETRY, 1997, 32 (04): : 351 - 369
  • [26] Cylindrical ion trap array with mass selection by variation in trap dimensions
    Badman, ER
    Cooks, RG
    ANALYTICAL CHEMISTRY, 2000, 72 (20) : 5079 - 5086
  • [27] A COMPARISON BETWEEN ION TRAP AND TRIPLE QUADRUPOLE COLLISION-RESULTS
    EVANS, C
    TRALDI, P
    MELE, A
    SOTTANI, C
    ORGANIC MASS SPECTROMETRY, 1991, 26 (04): : 347 - 349
  • [28] A cylindrical quadrupole ion trap in combination with an electrospray ion source for gas-phase luminescence and absorption spectroscopy
    Stockett, Mark H.
    Houmoller, Jorgen
    Stochkel, Kristian
    Svendsen, Annette
    Nielsen, Steen Brondsted
    REVIEW OF SCIENTIFIC INSTRUMENTS, 2016, 87 (05):
  • [30] Ion mobility measurement by dc tomography in an rf quadrupole ion trap
    Plass, WR
    Gill, LA
    Bui, HA
    Cooks, RG
    JOURNAL OF PHYSICAL CHEMISTRY A, 2000, 104 (21): : 5059 - 5065