A MOSAIC CRYSTAL BACKSCATTERING MONOCHROMATOR FOR HIGH-RESOLUTION POWDER DIFFRACTION

被引:2
|
作者
WROBLEWSKI, T
TERNES, W
机构
[1] Hamburger Synchrotronstrahlungslabor (HASYLAB), Deutsches Elektronensynchrotron (DESY), 2000 Hamburg 52
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 1992年 / 63卷 / 01期
关键词
D O I
10.1063/1.1143105
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The resolution function of a powder diffractometer at a synchrotron radiation source increases with increasing Bragg angle of the monochromator. Consequently, a monochromator close to backscattering has been designed and built. Two mosaic crystals of beryllium in nondispersive setting serve as optical elements. They are mounted on goniometer heads inside a tube-like housing which may be evacuated. Fine adjustment of one of the crystals is done with a micro-goniometer driven by a magnetostrictive actuator. The fact that the housing is a tube allows the use of the assembly as part of a multi-purpose monochromator system for powder diffraction.
引用
收藏
页码:1091 / 1093
页数:3
相关论文
共 50 条
  • [41] The crystal structure of γ-phase p-dichlorobenzene at low temperature and high pressure by high-resolution neutron powder diffraction
    Ibberson, RM
    Wilson, CC
    JOURNAL OF PHYSICS-CONDENSED MATTER, 2002, 14 (31) : 7287 - 7295
  • [43] COMPOSITE GERMANIUM CRYSTALS WITH ANISOTROPIC MOSAIC AS A FOCUSSING MONOCHROMATOR FOR NEUTRON POWDER DIFFRACTION.
    Vogt, T.
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 1996, 52 : C492 - C492
  • [44] HIGH-RESOLUTION NEUTRON POWDER DIFFRACTION STUDY OF ICE-IC
    KUHS, WF
    BLISS, DV
    FINNEY, JL
    JOURNAL DE PHYSIQUE, 1987, 48 (C-1): : 631 - 636
  • [45] High-Resolution Powder Neutron Diffraction on Mn3C
    Dierkes, Hannes
    Dronskowski, Richard
    ZEITSCHRIFT FUR ANORGANISCHE UND ALLGEMEINE CHEMIE, 2014, 640 (15): : 3148 - 3152
  • [46] Correcting the axial divergence from high-resolution powder diffraction data
    Dejoie, C.
    Fitch, A.
    ACTA CRYSTALLOGRAPHICA A-FOUNDATION AND ADVANCES, 2024, 80
  • [47] A twelve-analyzer detector system for high-resolution powder diffraction
    Lee, Peter L.
    Shu, Deming
    Ramanathan, Mohan
    Preissner, Curt
    Wang, Jun
    Beno, Mark A.
    Von Dreele, Robert B.
    Ribaud, Lynn
    Kurtz, Charles
    Antao, Sytle M.
    Jiao, Xuesong
    Toby, Brian H.
    JOURNAL OF SYNCHROTRON RADIATION, 2008, 15 (427-432) : 427 - 432
  • [48] The microstructure of biogenic calcite: A view by high-resolution synchrotron powder diffraction
    Pokroy, Boaz
    Fitch, Andy N.
    Zolotoyabko, Emil
    ADVANCED MATERIALS, 2006, 18 (18) : 2363 - +
  • [49] Dedicated high-resolution powder diffraction beamline at the advanced photon source
    Lee, PL
    Beno, MA
    Shu, DM
    Ramanathan, M
    Mitchell, JF
    Jorgensen, JD
    Von Dreele, RB
    SYNCHROTRON RADIATION INSTRUMENTATION, 2004, 705 : 388 - 391
  • [50] APPLICATION OF SYNCHROTRON RADIATION TO HIGH-RESOLUTION POWDER DIFFRACTION AND RIETVELD REFINEMENT
    COX, DE
    HASTINGS, JB
    THOMLINSON, W
    PREWITT, CT
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1983, 208 (1-3): : 573 - 578