The feasibility of sub-millisievert coronary CT angiography with low tube voltage, prospective ECG gating, and a knowledge-based iterative model reconstruction algorithm

被引:0
|
作者
Chul Hwan Park
Joohee Lee
Chisuk Oh
Kyung Hwa Han
Tae Hoon Kim
机构
[1] Yonsei University College of Medicine,Department of Radiology and Research Institute of Radiological Science, Gangnam Severance Hospital
[2] Yonsei University College of Medicine,Department of Radiology, Research Institute of Radiological Science, Severance Hospital
关键词
Multidetector computed tomography; Iterative model reconstruction; Low tube voltage; Prospective ECG gating; Coronary artery disease;
D O I
暂无
中图分类号
学科分类号
摘要
We evaluated the feasibility of sub-millisievert (mSv) coronary CT angiography (CCTA) using low tube voltage, prospective ECG gating, and a knowledge-based iterative model reconstruction algorithm. Twenty-four non-obese healthy subjects (M:F 13:11; mean age 50.2 ± 7.8 years) were enrolled. Three sets of CT images were reconstructed using three different reconstruction methods: filtered back projection (FBP), iterative reconstruction (IR), and knowledge-based iterative model reconstruction (IMR). The scanning parameters were as follows: step-and-shoot axial scanning, 80 kVp, and 200 mAs. On the three sets of CT images, the attenuation and image noise values were measured at the aortic root. The signal-to-noise ratio (SNR) and the contrast-to-noise ratio (CNR) were calculated at the proximal right coronary artery and the left main coronary artery. The qualitative image quality of the CCTA with IMR was assessed using a 4-point grading scale (grade 1, poor; grade 4, excellent). The mean radiation dose of the CCTA was 0.89 ± 0.09 mSv. The attenuation values with IMR were not different from those of other reconstruction methods. The image noise with IMR was significantly lower than with IR and FBP. Compared to FBP, the noise reduction rate of IMR was 69 %. The SNR and CNR of CCTA with IMR were significantly higher than with FBP or IR. On the qualitative analysis with IMR, all included segments were diagnostic (grades 2, 3, and 4), and the mean image quality score was 3.6 ± 0.6. In conclusion, CCTA with low tube voltage, prospective ECG gating, and an IMR algorithm might be a feasible method that allows for sub-millisievert radiation doses and good image quality when used with non-obese subjects.
引用
收藏
页码:197 / 203
页数:6
相关论文
共 50 条
  • [31] The Impact of Combining a Low-Tube Voltage Acquisition with Iterative Reconstruction on Total Iodine Dose in Coronary CT Angiography
    Van Cauteren, Toon
    Van Gompel, Gert
    Tanaka, Kaoru
    Verdries, Douwe E.
    Belsack, Dries
    Nieboer, Koenraad H.
    Willekens, Inneke
    Evans, Paul
    Macholl, Sven
    Verfaillie, Guy
    Droogmans, Steven
    de Mey, Johan
    Buls, Nico
    BIOMED RESEARCH INTERNATIONAL, 2017, 2017
  • [32] Evaluation of an iterative model-based reconstruction algorithm for low-tube-voltage (80 kVp) computed tomography angiography
    Noel, Peter B.
    Koehler, Thomas
    Fingerle, Alexander A.
    Brown, Kevin M.
    Zabic, Stanislav
    Muenzel, Daniela
    Haller, Bernhard
    Baum, Thomas
    Henninger, Martin
    Meier, Reinhard
    Rummeny, Ernst J.
    Dobritz, Martin
    JOURNAL OF MEDICAL IMAGING, 2014, 1 (03)
  • [33] Combination of a Low-Tube-Voltage Technique With Hybrid Iterative Reconstruction (iDose) Algorithm at Coronary Computed Tomographic Angiography
    Funama, Yoshinori
    Taguchi, Katsuyuki
    Utsunomiya, Daisuke
    Oda, Seitaro
    Yanaga, Yumi
    Yamashita, Yasuyuki
    Awai, Kazuo
    JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2011, 35 (04) : 480 - 485
  • [34] Feasibility study of low tube voltage (80 kVp) coronary CT angiography combined with contrast medium reduction using iterative model reconstruction (IMR) on standard BMI patients
    Zhang, Fan
    Yang, Li
    Song, Xiang
    Li, Ying-Na
    Jiang, Yan
    Zhang, Xing-Hua
    Ju, Hai-Yue
    Wu, Jian
    Chang, Rui-Ping
    BRITISH JOURNAL OF RADIOLOGY, 2016, 89 (1058):
  • [35] Coronary Computed Tomographic Angiography at 80 kVp and Knowledge-Based Iterative Model Reconstruction Is Non-Inferior to that at 100 kVp with Iterative Reconstruction
    Lee, Joohee
    Park, Chul Hwan
    Oh, Chi Suk
    Han, Kyunghwa
    Kim, Tae Hoon
    PLOS ONE, 2016, 11 (09):
  • [36] Improved image quality with simultaneously reduced radiation exposure: Knowledge-based iterative model reconstruction algorithms for coronary CT angiography in a clinical setting
    Andre, Florian
    Fortner, Philipp
    Vembar, Mani
    Mueller, Dirk
    Stiller, Wolfram
    Buss, Sebastian J.
    Kauczor, Hans-Ulrich
    Katus, Hugo A.
    Korosoglou, Grigorios
    JOURNAL OF CARDIOVASCULAR COMPUTED TOMOGRAPHY, 2017, 11 (03) : 213 - 220
  • [37] A Knowledge-based Iterative Model Reconstruction Algorithm: Can Super-Low-Dose Cardiac CT Be Applicable in Clinical Settings?
    Oda, Seitaro
    Utsunomiya, Daisuke
    Funama, Yoshinori
    Katahira, Kazuhiro
    Honda, Keiichi
    Tokuyasu, Shinichi
    Vembar, Mani
    Yuki, Hideaki
    Noda, Katsuo
    Oshima, Shuichi
    Yamashita, Yasuyuki
    ACADEMIC RADIOLOGY, 2014, 21 (01) : 104 - 110
  • [38] A deep-learning reconstruction algorithm that improves the image quality of low-tube-voltage coronary CT angiography
    Wang, Mengzhen
    Fan, Jing
    Shi, Xiaofeng
    Qin, Le
    Yan, Fuhua
    Yang, Wenjie
    EUROPEAN JOURNAL OF RADIOLOGY, 2022, 146
  • [39] Iterative model reconstruction (IMR) algorithm for reduced radiation dose renal artery CT angiography with different tube voltage protocols
    Qin, Le
    Ma, ZePeng
    Yan, FuHua
    Yang, WenJie
    RADIOLOGIA MEDICA, 2018, 123 (02): : 83 - 90
  • [40] Iterative model reconstruction (IMR) algorithm for reduced radiation dose renal artery CT angiography with different tube voltage protocols
    Le Qin
    ZePeng Ma
    FuHua Yan
    WenJie Yang
    La radiologia medica, 2018, 123 : 83 - 90