Determination of transmission curves of X-ray spectra for contrast-enhanced digital mammography using Monte Carlo simulations

被引:0
|
作者
Marques J.B. [1 ]
Leite J.P.S. [2 ]
da Cunha D.M. [1 ]
机构
[1] Institute of Physics, Federal University of Uberlândia, Uberlândia, 38400-902, MG
[2] Department of Radiotherapy, Sírio-Libanês Hospital, Brasília, 70200-730, DF
关键词
Contrast-enhanced digital mammography; Monte Carlo; Radiation protection; Shielding;
D O I
10.1007/s42600-021-00143-6
中图分类号
学科分类号
摘要
Purpose: To determine transmission curves of X-ray broad beams through common shielding materials, for the contrast-enhanced digital mammography (CEDM) imaging technique, using Monte Carlo simulations. Simulations were performed for anode/filter combinations of W/Rh and W/Ag, for tube potentials between 25 and 35 kVp, employed for low energy (LE) imaging, and for W/Cu, W/Ti and Mo/Cu between 35 and 49 kVp, for high energy (HE) imaging. Fitting parameters for each transmission curve were also obtained. Methods: Transmission curves were simulated using the Monte Carlo code PENELOPE. The geometric model adopted was composed by the breast compression paddle and the following common shielding materials: lead, steel, plate glass, concrete, gypsum, and wood. The obtained curves were fitted to the Archer equation, in order to determine their fitting parameters α, β, and γ. Results: Transmission curves for LE and HE spectra showed different behaviors. On average, values of half-value layer at high attenuation for LE spectra, at 35 kVp, are 2.36 times greater than those at 25 kVp, while for HE spectra, at 49 kVp, excepting for wood, are 2.39 times greater than those at 35 kVp. Results for fitting parameters showed that α values decrease with tube potential, whereas the behavior of β and γ values depends on the spectrum and shielding material. Conclusions: The results obtained and the fitting parameters provided can be useful for evaluating the shielding design of this technique, since there are significant differences between results for the LE and HE spectra, evidentiating the importance of considering appropriate transmission data. © 2021, Sociedade Brasileira de Engenharia Biomedica.
引用
收藏
页码:319 / 328
页数:9
相关论文
共 50 条
  • [21] Material Decomposition in Contrast-enhanced Digital Mammography Based on Photon Counting: Monte Carlo Simulation Studies
    Choi, Yu-Na
    Cho, Hyo-Min
    Lee, Seung-Wan
    Ryu, Hyun-Ju
    Lee, Young-Jin
    Kim, Hee-Joung
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2011, 59 (01) : 161 - 168
  • [22] Monte Carlo simulations in X-ray imaging
    Giersch, Juergen
    Durst, Juergen
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2008, 591 (01): : 300 - 305
  • [23] Monte Carlo performance on the x-ray converter thickness in digital mammography using software breast models
    Liaparinos, P.
    Bliznakova, K.
    MEDICAL PHYSICS, 2012, 39 (11) : 6638 - 6651
  • [24] A microCT X-ray head model for spectra generation with Monte Carlo simulations
    Taschereau, R.
    Chow, P. L.
    Cho, J. S.
    Chatziioannou, A. F.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 569 (02): : 373 - 377
  • [25] High Energy Resolution Hyperspectral X-Ray Imaging for Low-Dose Contrast-Enhanced Digital Mammography
    Pani, Silvia
    Saifuddin, Sarene C.
    Ferreira, Filipa I. M.
    Henthorn, Nicholas
    Seller, Paul
    Sellin, Paul J.
    Stratmann, Philipp
    Veale, Matthew C.
    Wilson, Matthew D.
    Cernik, Robert J.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2017, 36 (09) : 1784 - 1795
  • [26] Dual-energy contrast-enhanced digital mammography: Glandular dose estimation using a Monte Carlo code and voxel phantom
    Tzamicha, E.
    Yakoumakis, E.
    Tsalafoutas, I. A.
    Dimitriadis, A.
    Georgiou, E.
    Tsapaki, V.
    Chalazonitis, A.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2015, 31 (07): : 785 - 791
  • [27] Analysis of multilayer and single layer X-ray detectors for contrast-enhanced mammography using imaging task
    Allec, Nicholas
    Abbaszadeh, Shiva
    Karim, Karim S.
    MEDICAL IMAGING 2011: PHYSICS OF MEDICAL IMAGING, 2011, 7961
  • [28] Monte Carlo simulation of x-ray spectra in diagnostic radiology and mammography using MCNP4C
    Ay, MR
    Shahriari, M
    Sarkar, S
    Adib, M
    Zaidi, H
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (21): : 4897 - 4917
  • [29] X-ray Microanalysis of Porous Materials Using Monte Carlo Simulations
    Poirier, Dominique
    Gauvin, Raynald
    SCANNING, 2011, 33 (03) : 126 - 134
  • [30] X-Ray Microanalysis of Real Materials Using Monte Carlo Simulations
    Raynald Gauvin
    Eric Lifshin
    Microchimica Acta, 2004, 145 : 41 - 47