Determination of transmission curves of X-ray spectra for contrast-enhanced digital mammography using Monte Carlo simulations

被引:0
|
作者
Marques J.B. [1 ]
Leite J.P.S. [2 ]
da Cunha D.M. [1 ]
机构
[1] Institute of Physics, Federal University of Uberlândia, Uberlândia, 38400-902, MG
[2] Department of Radiotherapy, Sírio-Libanês Hospital, Brasília, 70200-730, DF
关键词
Contrast-enhanced digital mammography; Monte Carlo; Radiation protection; Shielding;
D O I
10.1007/s42600-021-00143-6
中图分类号
学科分类号
摘要
Purpose: To determine transmission curves of X-ray broad beams through common shielding materials, for the contrast-enhanced digital mammography (CEDM) imaging technique, using Monte Carlo simulations. Simulations were performed for anode/filter combinations of W/Rh and W/Ag, for tube potentials between 25 and 35 kVp, employed for low energy (LE) imaging, and for W/Cu, W/Ti and Mo/Cu between 35 and 49 kVp, for high energy (HE) imaging. Fitting parameters for each transmission curve were also obtained. Methods: Transmission curves were simulated using the Monte Carlo code PENELOPE. The geometric model adopted was composed by the breast compression paddle and the following common shielding materials: lead, steel, plate glass, concrete, gypsum, and wood. The obtained curves were fitted to the Archer equation, in order to determine their fitting parameters α, β, and γ. Results: Transmission curves for LE and HE spectra showed different behaviors. On average, values of half-value layer at high attenuation for LE spectra, at 35 kVp, are 2.36 times greater than those at 25 kVp, while for HE spectra, at 49 kVp, excepting for wood, are 2.39 times greater than those at 35 kVp. Results for fitting parameters showed that α values decrease with tube potential, whereas the behavior of β and γ values depends on the spectrum and shielding material. Conclusions: The results obtained and the fitting parameters provided can be useful for evaluating the shielding design of this technique, since there are significant differences between results for the LE and HE spectra, evidentiating the importance of considering appropriate transmission data. © 2021, Sociedade Brasileira de Engenharia Biomedica.
引用
收藏
页码:319 / 328
页数:9
相关论文
共 50 条
  • [41] Detection of radium in water by x-ray fluorescence using Monte Carlo simulations
    Burille, F.
    Correa, J. J. M.
    Zambianchi, P.
    Zambianchi, J. K.
    Antoniassi, M.
    RADIATION PHYSICS AND CHEMISTRY, 2020, 167
  • [42] Characterization of X-ray scatter and glandular dose in digital tomosynthesis for breast imaging using Monte Carlo simulations
    Sechopoulos, I.
    Suryanarayanan, S.
    Vedantham, S.
    Karellas, A.
    MEDICAL PHYSICS, 2006, 33 (06) : 1990 - 1990
  • [43] X-ray radiography of HPGe detectors for Monte Carlo simulations
    Pohuliai, Serhii
    Vlasenko, Andrey
    Malgin, Vladislav
    Kanapelka, Marina
    APPLIED RADIATION AND ISOTOPES, 2023, 198
  • [44] Contrast-enhanced x-ray microscopy of articular cartilage
    Zhu, Ying
    Ponjevic, Dragana
    Matyas, John R.
    Boyd, Steven K.
    CONNECTIVE TISSUE RESEARCH, 2021, 62 (05) : 542 - 553
  • [45] Coating Thickness Determination Using X-ray Fluorescence Spectroscopy: Monte Carlo Simulations as an Alternative to the Use of Standards
    Giurlani, Walter
    Berretti, Enrico
    Innocenti, Massimo
    Lavacchi, Alessandro
    COATINGS, 2019, 9 (02):
  • [46] Monte Carlo simulation of X-ray spectra produced by Linac
    Permatasari, Intan Dyah Ayu
    Suharyana
    Riyatun
    9TH INTERNATIONAL CONFERENCE ON PHYSICS AND ITS APPLICATIONS (ICOPIA), 2019, 1153
  • [47] Monte Carlo simulation of x-ray tube spectra with PENELOPE
    Jia, Pengxiang
    Xie, Yaoqin
    Bao, Shanglian
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 503 - 506
  • [48] Monte Carlo simulation of X-ray source characteristics using MCPETS and MCNPX codes for mammography
    Flegontova, E. Yu.
    Bakaleinikov, L. A.
    Nam, Ki-Yong
    Park, Jung Gon
    Kim, Kyong-Woo
    Chon, Kwon Su
    Yoon, Kwon-Ha
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2007, 51 (01) : 65 - 77
  • [49] Contrast-enhanced digital mammography (CEDM): Imaging modeling, computer simulations, and phantom study
    Chen, B
    Jing, ZX
    Smith, A
    MEDICAL IMAGING 2005: PHYSICS OF MEDICAL IMAGING, PTS 1 AND 2, 2005, 5745 : 877 - 888
  • [50] Characterization of Kilovoltage X-Ray Spectra Using Measurements with a High-Purity Germanium Detector and Monte Carlo Simulations
    Moga, J.
    DeWerd, L.
    MEDICAL PHYSICS, 2010, 37 (06) : 3427 - +