Development of a Daily-Treatment Beam-Monitoring System Based Gafchromic EBT3 Film for Pencil-Beam Scanning Proton Therapy

被引:0
|
作者
Seonghoon Jeong
Myonggeun Yoon
Kwangzoo Chung
机构
[1] National Cancer Center,Proton Therapy Center
[2] Korea University,Department of Bio
[3] Sungkyunkwan University,convergence Engineering
来源
关键词
Proton therapy; Pencil-beam scanning; Gafchromic EBT3 film; Daily beam monitoring;
D O I
暂无
中图分类号
学科分类号
摘要
This study describes the development of a simple method to assess inter-fractional deviations of delivered proton beams in treatment rooms. To monitor the treatment beam, we measured the field-by-field beam fluences by attaching the EBT3 film to the snout, followed by a simple constancy check based on comparisons between the reference beam fluences (acquired during the pre-treatment quality assurance process) and the test beam fluences (acquired during treatment). The feasibility of the proposed treatment beam-monitoring system was confirmed by evaluating 12 treatment fields for each of six patients (brain, liver, prostate, lung, cranial and spinal area, and head and neck). The constancy of the treatment beams was verified by using a gamma index analysis to compare three measurements per field with the reference beam fluence. On the basis of the 3%/3 mm criterion, the average gamma-index passing rates for all measurements were over 99.6%. These results suggest that the constancy of fractional proton beams delivered to patients in treatment rooms can be verified with EBT3 film-based proton-beam monitoring system that can be easily attached to the treatment nozzle and is cost effective.
引用
收藏
页码:769 / 773
页数:4
相关论文
共 49 条
  • [21] The Applications and Pitfalls of Cone-Beam Computed Tomography-Based Synthetic Computed Tomography for Adaptive Evaluation in Pencil-Beam Scanning Proton Therapy
    Tsai, Pingfang
    Tseng, Yu-Lun
    Shen, Brian
    Ackerman, Christopher
    Zhai, Huifang A.
    Yu, Francis
    Simone II, Charles B.
    Choi, J. Isabelle
    Lee, Nancy Y.
    Kabarriti, Rafi
    Lazarev, Stanislav
    Johnson, Casey L.
    Liu, Jiayi
    Chen, Chin-Cheng
    Lin, Haibo
    CANCERS, 2023, 15 (20)
  • [22] Development of Optical Fiber Based Measurement System for the Verification of Entrance Dose Map in Pencil Beam Scanning Proton Beam
    Son, Jaeman
    Lee, Se Byeong
    Lim, Youngkyung
    Park, Sung Yong
    Cho, Kwanho
    Yoon, Myonggeun
    Shin, Dongho
    SENSORS, 2018, 18 (01):
  • [23] Technical note: Proton beam dosimetry at ultra-high dose rates (FLASH): Evaluation of GAFchromic™ (EBT3, EBT-XD) and OrthoChromic (OC-1) film performances
    Institut de Cancérologie de l'Ouest, Saint-Herblain, France
    不详
    不详
    Med. Phys., 4 (2732-2745):
  • [24] REDUCING DOSE TO DENTOFACIAL STRUCTURES: A PENCIL-BEAM SCANNING PROTON THERAPY AND VOLUMETRIC-ARC THERAPY PHOTON TREATMENT PLANNING COMPARISON STUDY
    Davies, Lucy Siew Chen
    Charlwood, Frances
    Hol, Marinka
    Aznar, Marianne C.
    Davey, Angela
    Eccles, Cynthia
    Foster-Thomas, Emma
    Gaito, Simona
    Gaze, Mark
    Indelicato, Danny
    Mandeville, Henry
    Slater, Olga
    Sitch, Peter
    Tang, Vivian
    Whitfield, Gillian
    Pan, Shermaine
    PEDIATRIC BLOOD & CANCER, 2023, 70 : S68 - S69
  • [25] Clinical application of a gantry-attachable plastic scintillating plate dosimetry system in pencil beam scanning proton therapy beam monitoring
    Jeong, Seonghoon
    Yoon, Myonggeun
    Chung, Kwangzoo
    Ahn, Sung Hwan
    Lee, Boram
    Seo, Jaehyeon
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 77 : 181 - 186
  • [26] Development of a new 3-dimensional scintillating detector for patient treatment quality control in pencil beam scanning proton therapy
    Frelin, Anne-Marie
    Daviau, G.
    Bui, My Hoang Hoa
    Fontbonne, Cathy
    Fontbonne, Jean-Marc
    Lebhertz, Dorothee
    Mainguy, Erwan
    Moignier, Cyril
    Vela, Anthony
    Thariat, Juliette
    13TH INTERNATIONAL CONFERENCE ON 3D DOSIMETRY, IC3DDOSE 2024, 2024, 2799
  • [27] Development and long-term stability of a comprehensive daily QA program for a modern pencil beam scanning (PBS) proton therapy delivery system
    Rana, Suresh
    Bennouna, Jaafar
    Samuel, E. James Jebaseelan
    Gutierrez, Onso N.
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (04): : 29 - 44
  • [28] Development and Validation of a TOPAS Monte Carlo Model of a Dynamic Collimation System for Pencil Beam Scanning Proton Therapy
    Nelson, N.
    Culberson, W.
    Hyer, D.
    Geoghegan, T.
    Patwardhan, K.
    Smith, B.
    Flynn, R.
    Yu, J.
    Rana, S.
    Gutierrez, A.
    Hill, P.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [29] Investigating Aperture-Based Approximations to Model a Dynamic Collimation System for Pencil Beam Scanning Proton Therapy
    Nelson, N.
    Smith, B.
    Culberson, W.
    Flynn, R.
    Hyer, D.
    Rana, S.
    Hill, P.
    MEDICAL PHYSICS, 2020, 47 (06) : E802 - E802
  • [30] Application of Monte Carlo Simulation for Treatment Planning System Commissioning and Dose Calculation for a Pencil Beam Scanning Proton Therapy System
    Hu, H.
    Liu, H.
    Jia, Y.
    Zheng, Y.
    Ding, S.
    Wang, B.
    Li, Y.
    Huang, X.
    Li, Z.
    MEDICAL PHYSICS, 2022, 49 (06) : E860 - E860