Parallax-free intra-operative X-ray image stitching

被引:33
|
作者
Wang, Lejing [1 ]
Traub, Joerg [1 ]
Weidert, Simon [2 ]
Heining, Sandro Michael [2 ]
Euler, Ekkehard [2 ]
Navab, Nassir [1 ]
机构
[1] Tech Univ Munich, Fak Informat, D-85748 Garching, Germany
[2] Ludwig Maximilians Univ Munchen, Klinikum Innenstadt, Dept Trauma Surg, Munich, Germany
关键词
X-ray image stitching; Camera Augmented Mobile C-arm; Computer aided surgery; FRACTURES; MALALIGNMENT; RADIOGRAPHY; ANGLE;
D O I
10.1016/j.media.2010.05.007
中图分类号
TP18 [人工智能理论];
学科分类号
081104 ; 0812 ; 0835 ; 1405 ;
摘要
We present a novel method to generate parallax-free panoramic X-ray images during surgery by enabling the mobile C-arm to rotate around its X-ray source center, relative to the patient's table. Rotating the mobile C-arm around its X-ray source center is impractical and sometimes impossible due to the mechanical design of mobile C-arm systems. In order to ensure that the C-arm motion is a relative pure rotation around its X-ray source center, we propose to move the table to compensate for the translational part of the motion based on C-arm pose estimation. For this we employ a visual marker pattern and a Camera Augmented Mobile C-arm system that is a standard mobile C-arm augmented by a video camera and mirror construction. We are able to produce a parallax-free panoramic X-ray image independent of the geometric configuration of imaged anatomical structures. Our method does not require a fronto-parallel setup or any overlap between the acquired X-ray images. This generated parallax-free panoramic Xray image preserves the linear perspective projection property. It also presents a negligible difference (below 2 pixels) in the overlapping area between two consecutive individual X-ray images and has a high visual quality. This promises suitability for intra-operative clinical applications in orthopedic and trauma surgery. The experiments on phantoms and ex-vivo bone structure demonstrate both the functionality and accuracy of the method. (C) 2010 Elsevier BM. All rights reserved.
引用
收藏
页码:674 / 686
页数:13
相关论文
共 50 条
  • [1] Parallax-Free Long Bone X-ray Image Stitching
    Wang, Lejing
    Traub, Joerg
    Weidert, Simon
    Heining, Sandro Michael
    Euler, Ekkehard
    Navab, Nassir
    MEDICAL IMAGE COMPUTING AND COMPUTER-ASSISTED INTERVENTION - MICCAI 2009, PT I, PROCEEDINGS, 2009, 5761 : 173 - +
  • [2] Parallax-free X-ray area detector with high spatial resolution
    Zanevsky, YV
    Chernenko, SP
    Cheremukhina, GA
    Donets, DE
    Fateev, OV
    Smykov, LP
    Vasiliev, SE
    Kheiker, DM
    Andrianova, ME
    Sulianov, SN
    NUCLEAR PHYSICS B, 1995, : 406 - 408
  • [3] The planispherical chamber: A parallax-free gaseous X-ray detector for imaging applications
    Brunbauer, F. M.
    Oliveri, E.
    Resnati, F.
    Ropelewski, L.
    Sauli, F.
    Thuiner, P.
    van Stenis, M.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2017, 875 : 16 - 20
  • [4] A MULTISTEP PARALLAX-FREE X-RAY-IMAGING COUNTER
    BRESKIN, A
    CHARPAK, G
    SANTIARD, JC
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 195 (03): : 469 - 473
  • [5] Radiation protection for an intra-operative X-ray device
    Eaton, D. J.
    Gonzalez, R.
    Duck, S.
    Keshtgar, M.
    BRITISH JOURNAL OF RADIOLOGY, 2011, 84 (1007): : 1034 - 1039
  • [6] Dosimetry measurements with an intra-operative x-ray device
    Eaton, D. J.
    Duck, S.
    PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (12): : N359 - N369
  • [7] Radiation protection for an intra-operative X-ray source
    Schneider, F.
    Clausen, S.
    Klabes, A.
    Steil, V
    Bludau, F.
    Suetterlin, M.
    Obertacke, U.
    Wenz, F.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2012, 188 : 171 - 171
  • [8] PARALLAX-FREE, HIGH-ACCURACY GASEOUS DETECTORS FOR X-RAY AND VUV LOCALIZATION
    CHARPAK, G
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH, 1982, 201 (01): : 181 - 192
  • [9] A PARALLAX-FREE GRATICULE FOR CATHODE RAY OSCILLOSCOPES
    KANTOROWICZ, O
    JOURNAL OF SCIENTIFIC INSTRUMENTS, 1947, 24 (03): : 80 - 80
  • [10] Fast, parallax-free, one-coordinate X-ray detector OD3
    Aulchenko, V.M.
    Bukin, M.A.
    Velikzhanin, Yu.S.
    Gaponenko, Ya.V.
    Dubrovin, M.S.
    Titov, V.M.
    Ancharov, A.I.
    Gaponov, Yu.A.
    Evdokov, O.V.
    Tolochko, B.P.
    Sharafutdinov, M.R.