Characterization of a 2D ionization chamber array for IMRT QA

被引:0
|
作者
Shtraus, N.
Curran, B.
Moran, J.
Litzenberg, D.
机构
[1] Tel Aviv Med Ctr & Sch Med, Tel Aviv, Israel
[2] Univ Michigan, Med Ctr, Ann Arbor, MI USA
关键词
D O I
10.1118/1.2761556
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Background and Purpose: Measurement‐based methods of quality assurance (QA) for Intensity Modulated Radiotherapy (IMRT) fields often include individual field verification at a neutral gantry angle using film or other 2D detector in addition to a composite ionization chamber check. Unfortunately, film is time‐consuming to use. Diode systems are currently available but these arrays are limited in size and spacing of the diodes. This work characterizes a new 2D ionization chamber array (MatriXX, Scanditronix‐Wellhofer) that can be used for treatment verification at neutral and planned gantry angles. Material and methods: The 2D array consists of 1020 ion chambers, arranged in a 32×32 cm2 grid. Each chamber is separated by 7.62 mm center to center and has a volume of 0.08 cm3. Properties such as linearity, reproducibility, pressure dependence, homogeneity, and accuracy of the 2D array have been evaluated for 6 and 16MV beams at neutral and planned gantry angles. Results: The 2D array response was found to be linear with dose from 0–800 cGy. The 2D array readings were reproducible to within a SD of ±0.14%. Examination of the array's pressure dependence, homogeneity, sensitivity as a function of ion chamber location, and the interpolated dose showed accuracies of better than 0.3%. Absolute dose measurements were done with the array and a calibrated ion chamber. Agreement was better than 0.6%. TPR measurements were made and found to agree with commissioning data in our planning system (UMplan). Measurements done using a 2 × 2 cm2 field placed randomly around the array show a standard deviation of 0.21%. Delivery of IMRT plans using multiple gantry angles showed excellent agreement with calculated results. Conclusions: The 2D array offers an excellent solution to simplifying QA and saving time, particularly for IMRT pre‐treatment QA. The device can be used for both relative and absolute dose measurements. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2600 / 2600
页数:1
相关论文
共 50 条
  • [1] Evaluation of Hybrid Plan IMRT QA with a 2D Ionization Chamber Array
    Depuydt, T.
    Gomola, I.
    Van den Heuvel, F.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [2] Characterization of a 2D ionization chamber array for IMRT plan verification
    Alashrah, S.
    Kandaiya, S.
    Yong, S. Y.
    Cheng, S. K.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 619 (1-3): : 181 - 185
  • [3] Guidelines for IMRT-QA using 2D-ionization chamber array
    Duran, Molina F. J.
    Schneider, F.
    Wenz, F.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2010, 186 : 107 - 107
  • [4] Dosimetric evaluation of a 2D ionisation chamber array for application for IMRT QA
    Herzen, J.
    Todorovic, M.
    Cremers, F.
    Platz, V.
    Albers, D.
    Fischer, M.
    Thom, E.
    Schmidt, R.
    RADIOTHERAPY AND ONCOLOGY, 2006, 81 : S463 - S463
  • [5] IMRT QA with a 2D diode array
    Ding, L
    Ahluwalia, W
    Liu, C
    Li, J
    Palta, J
    MEDICAL PHYSICS, 2003, 30 (06) : 1396 - 1396
  • [6] A new 2D ionization chamber array for dosimetry verification of IMRT
    Gomola, I
    Giordanengo, S
    Iliescu, S
    Cirio, R
    Donetti, M
    Marchetto, F
    Peroni, C
    Stasi M, M
    RADIOTHERAPY AND ONCOLOGY, 2005, 76 : S120 - S121
  • [7] Evaluation of a 2D diode array for IMRT QA
    Letourneau, D
    MEDICAL PHYSICS, 2003, 30 (06) : 1396 - 1396
  • [8] Application of a 2D Ionization Chamber Array for Monthly and Annual Linear Accelerator QA
    Buckey, C.
    Myers, P.
    Mavroidis, P.
    Esquivel, C.
    Mihailidis, D.
    Gutierrez, A.
    Papanikolaou, N.
    Stathakis, S.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [9] Dosimetric plan verification using a 2D ionization chamber array in helical IMRT
    Yango, A.
    Schoroth, F.
    Muedder, T.
    Poppe, B.
    Garbe, S.
    STRAHLENTHERAPIE UND ONKOLOGIE, 2010, 186 : 108 - 109
  • [10] Comparison of a 2D and 3D Array of Diodes for IMRT QA
    Lawrence, M.
    Potter, L.
    Cullip, T.
    Xu, Z.
    Deschesne, K.
    Chang, S.
    MEDICAL PHYSICS, 2009, 36 (06)