Odontoid plate fixation without C1-C2 arthrodesis: Biomechanical testing of a novel surgical technique and comparison to the conventional screw fixation procedure

被引:5
|
作者
Platzer, Patrick [1 ]
Eipeldauer, Stefan [1 ]
Vecsei, Vilmos [1 ]
机构
[1] Med Univ Vienna, Dept Trauma Surg, A-1090 Vienna, Austria
关键词
Odontoid fractures; Plate fixation; Biomechanical testing; FRACTURES; STABILIZATION;
D O I
10.1016/j.clinbiomech.2010.03.012
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Background: Odontoid plate fixation without C1-C2 arthrodesis appears to be a practicable option for the management of odontoid fractures that are not amenable for conventional screw fixation. The purpose of this study was to measure the mechanical stability of odontoid plate fixation using a specially designed plate construct, and to compare the results to those after conventional screw fixation. Methods: The second cervical vertebra was removed from twenty fresh human spinal columns. Stiffness and failure load of the intact odontoid were measured, and type II odontoid fractures were created. Afterward, the specimens were randomly assigned to one of the following four groups: Group I: plate-fixation; Group II: 2-AO-screw-fixation; Group III: 1-AO-screw-fixation; Group IV: Herbert-screw-fixation. In a second series, stiffness and failure load of the stabilized odontoid fractures were assessed for comparison and statistical analysis. Findings: Group I showed a significantly higher mean failure load than the other groups. The mean failure load of Group I after fixation of the odontoid fracture was 86% of the mean failure load of the intact odontoid. Comparing Groups II, III and IV. there was no significant difference regarding the failure load. In these three groups the mean failure load after odontoid fixation was approximately 50% of the mean failure load of the intact odontoid. Interpretation: Odontoid plate fixation as an alternative procedure in certain fracture patterns provided a significantly higher biomechanical stability than the technique of odontoid screw fixation. Using a specially designed plate construct, 86% of the original stability of the intact odontoid was restored. (C) 2010 Elsevier Ltd. All rights reserved.
引用
收藏
页码:623 / 627
页数:5
相关论文
共 50 条
  • [21] Biomechanical Analysis Comparing Three C1-C2 Transarticular Screw Salvaging Fixation Techniques
    Elgafy, Hossein
    Potluri, Tejaswy
    Goel, Vijay K.
    Foster, Scott
    Faizan, Ahmad
    Kulkarni, Nikhil
    SPINE, 2010, 35 (04) : 378 - 385
  • [22] Combined Anterior Odontoid Screw Fixation and C1-C2 Jefferson Fracture Anterior Fixation: A Step-by-Step Technical Note
    Frati, Alessandro
    Cimatti, Marco
    Ileyassoff, Hernan
    Capobianco, Mattia
    Santoro, Antonio
    Armocida, Daniele
    WORLD NEUROSURGERY, 2024, 184 : 112 - 118
  • [23] Modified Gallie technique versus transarticular screw fixation in C1-C2 fusion
    Farey, ID
    Nadkarni, S
    Smith, N
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (359) : 126 - 135
  • [24] C1-c2 transarticular screw fixation: Technical aspects
    Haid, RW
    NEUROSURGERY, 2001, 49 (01) : 71 - 74
  • [25] C1-C2 transarticular screw fixation for atlantoaxial instability
    Fountas, KN
    Kapsalaki, EZ
    Karampelas, I
    Dimopoulos, VG
    Feltes, CH
    Kassam, MA
    Boev, AN
    Johnston, KW
    Smisson, HF
    Troup, EC
    Robinson, JS
    SOUTHERN MEDICAL JOURNAL, 2004, 97 (11) : 1042 - 1048
  • [26] Biomechanical comparison of C1-C2 posterior arthrodesis techniques
    Papagelopoulos, Panayiotis J.
    Currier, Bradford L.
    Hokari, Yukitaka
    Neale, Patricia G.
    Zhao, Chunfeng
    Berglund, Lawrence J.
    Larson, Dirk R.
    An, Kai-Nan
    SPINE, 2007, 32 (13) : E363 - E370
  • [27] A comparison of biomechanical stability and pullout strength of two C1-C2 fixation constructs
    Savage, Jason W.
    Limthongkul, Worawat
    Park, Hyung-Soon
    Zhang, Li-Qun
    Karaikovic, Eldin E.
    SPINE JOURNAL, 2011, 11 (07): : 654 - 658
  • [28] C1-C2 transarticular screw fixation for atlantoaxial instability: A 6-year experience, and C1-C2 transarticular screw fixation-technical aspects
    Low, HL
    Redfern, RM
    NEUROSURGERY, 2002, 50 (05) : 1165 - 1166
  • [29] Biomechanical Evaluations of Various C1-C2 Posterior Fixation Techniques
    Sim, Hong Bo
    Lee, Jong Won
    Park, Jong Tae
    Mindea, Stefan A.
    Lim, Jesse
    Park, Jon
    SPINE, 2011, 36 (06) : E401 - E407
  • [30] Virtual placement of posterior C1-C2 transarticular screw fixation
    Spangenberg, P
    Coenen, V
    Gilsbach, JM
    Rohde, V
    NEUROSURGICAL REVIEW, 2006, 29 (02) : 114 - 117