Quality assurance for a treatment planning system in scanned ion beam therapy

被引:43
|
作者
Jäkel, O
Hartmann, GH
Karger, CP
Heeg, P
Rassow, J
机构
[1] Deutsch Krebsforschungszentrum, Abt E0400, D-69120 Heidelberg, Germany
[2] Univ Essen Gesamthsch, Radiol Ctr, Essen, Germany
关键词
quality assurance; treatment planning; scanned ion beam;
D O I
10.1118/1.599025
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Conformal radiation therapy using dynamic beam delivery systems like scanned ion beams requires concise quality assurance procedures for the complete treatment planning process. For the heavy ion therapy facility at GSI, Darmstadt, a quality assurance program for the treatment planning system (TPS) has been developed. It covers the development and updating of software, data protection and safety, and the application of soft- and hardware. The tests also apply to the geometrical precision of imaging devices and the geometrical and dosimetrical verification of dose distributions in different phantoms. The quality assurance program addresses acceptance and constancy tests of the treatment planning program. Results of the acceptance tests served as a basis for its governmental approval. Two main results of the acceptance tests are representative for the overall performance of the system. (1) The geometrical uncertainty that could be achieved for the target point definition, setup accuracy, field contouring, and field alignment is typically 1.5 mm. The uncertainty for the setup verification using digitally reconstructed radiographs (DRR's) is limited to 2 mm. (2) The mean deviations between measured and planned dose values is 3% for standardized cases in a water phantom and up to 6% for more complicated treatment configurations. (C) 2000 American Association of Physicists in Medicine. [S0094-2405(00)00707-0].
引用
收藏
页码:1588 / 1600
页数:13
相关论文
共 50 条
  • [21] Commissioning of a commercial treatment planning system for scanned carbon-ion radiotherapy
    Sun, Wei
    Wang, Weiwei
    Huang, Zhijie
    Zhao, Jingfang
    JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2025, 26 (03):
  • [22] Performance of a Beam Tracking System for Treatment of Moving Targets with Scanned Ion Beams
    Bert, C.
    Saito, N.
    Schmidt, A.
    Chaudhri, N.
    Schardt, D.
    Rietzel, E.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [23] Management of organ motion in scanned ion beam therapy
    Christoph Bert
    Klaus Herfarth
    Radiation Oncology, 12
  • [24] Management of organ motion in scanned ion beam therapy
    Bert, Christoph
    Herfarth, Klaus
    RADIATION ONCOLOGY, 2017, 12
  • [25] Adjustment procedure for beam alignment in scanned ion-beam therapy
    Saraya, Y.
    Takeshita, E.
    Furukawa, T.
    Hara, Y.
    Mizushima, K.
    Saotome, N.
    Tansho, R.
    Shirai, T.
    Noda, K.
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2016, 69 (06) : 948 - 952
  • [26] Adjustment procedure for beam alignment in scanned ion-beam therapy
    Y. Saraya
    E. Takeshita
    T. Furukawa
    Y. Hara
    K. Mizushima
    N. Saotome
    R. Tansho
    T. Shirai
    K. Noda
    Journal of the Korean Physical Society, 2016, 69 : 948 - 952
  • [27] Quality assurance and quality control in operable breast carcinoma external beam radiotherapy treatment planning
    Stankovic, J
    Saric, M
    Borojevic, N
    RadosevicJelic, L
    BALKAN CONGRESS OF ONCOLOGY, 1996, : 177 - 183
  • [28] Commissioning and quality assurance-of the Pinnacle3 radiotherapy treatment planning system for external beam photons
    Bedford, JL
    Childs, PJ
    Hansen, VN
    Mosleh-Shirazi, MA
    Verhaegen, F
    Warrington, AP
    BRITISH JOURNAL OF RADIOLOGY, 2003, 76 (903): : 163 - 176
  • [29] Online Quality Assurance of External Beam Radiation Therapy With an Integrated Quality Monitoring System
    Hoffman, D.
    Chung, E.
    Hess, C. B.
    Stern, R. L.
    Benedict, S. H.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2015, 93 (03): : E582 - E582
  • [30] MITIGATION OF RESIDUAL MOTION EFFECTS IN SCANNED ION BEAM THERAPY
    Richter, D.
    Steidl, P.
    Trautmann, J.
    Schwarzkopf, A.
    Kraemer, M.
    Gemmel, A.
    Naumann, J.
    Panse, R.
    Saito, N.
    Jaekel, O.
    Durante, M.
    Bert, C.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S72 - S73