Ion-selective optical sensor for continuous on-line monitoring of dialysate sodium during dialysis

被引:1
|
作者
Sharma, Manoj K. [1 ]
Frijns, Arjan J. H. [1 ]
Mandamparambil, Rajesh [2 ]
Kooman, Jeroen P. [3 ]
Smeulders, David M. J. [1 ]
机构
[1] Eindhoven Univ Technol, Dept Mech Engn, Eindhoven, Netherlands
[2] TNO, Eindhoven, Netherlands
[3] Maastricht Univ Med Ctr MUMC, Maastricht, Netherlands
关键词
Optical sensor; Micro-optofluidics; On-line electrolyte monitoring; Dialysis; Photoinduced electron transfer; FLUORESCENT; SYSTEMS;
D O I
10.1117/12.2251923
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Patients with end stage renal disease are dependent on dialysis. In most outpatient centers, the dialysate is prepared with a fixed electrolyte concentration without taking into account the inter-individual differences of essential electrolytes (sodium, potassium and calcium). This one-size fits all approach can lead to acute and chronic cardiovascular complications in dialysis patients. On-line monitoring of these essential electrolytes is an important physiological step towards patient specific dialysate leading to individualized treatment. Currently, changes in electrolyte concentrations are indirectly measured by conductivity measurements, which are not ion-specific. In this paper, we present a novel optical sensor for on-line monitoring of sodium concentrations in dialysate. This sensor is ion-specific and can detect up to a single ion. The working principle is based on the selective fluorescence quenching of photo-induced electron transfer (PET) molecules. The PET molecules when complexed with sodium ions start fluorescing upon laser excitation. The emission intensity is directly correlated to the sodium concentration. To prove the working principle, a micro-optofluidic device has been fabricated in polydimethylsiloxane (PDMS) with integrated optical fibers for fluorescence light collection. The PET molecules are covalently grafted in the PDMS microchannel for continuous monitoring of the sodium dialysate concentrations. The experimental setup consists of a laser module (lambda=450nm) operating at 4.5mW, a syringe pump to precisely control the sample flow and a spectrometer for fluorescence collection. The performance of the sensor has been evaluated for sodium ions ranging from 0-50mM. A clear signal and good response time was observed.
引用
收藏
页数:8
相关论文
共 50 条
  • [1] On-line blood and dialysate sampling system for multiparametric monitoring during dialysis
    Lurzer, W
    Scharfetter, H
    Hutten, H
    PROCEEDINGS OF THE 18TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOL 18, PTS 1-5, 1997, 18 : 77 - 78
  • [2] Biomass monitoring by use of a continuous on-line optical sensor
    Salgado, AM
    Folly, ROM
    Valdman, B
    SENSORS AND ACTUATORS B-CHEMICAL, 2001, 75 (1-2) : 24 - 28
  • [3] Potentiometric ion-selective silicon sensors for the on-line monitoring of blood electrolytes
    Scheipers, A
    Wassmus, O
    Sundermeier, C
    Eshold, J
    Weiss, T
    Gitter, M
    Ross, B
    Knoll, M
    ANALYTICA CHIMICA ACTA, 2001, 439 (01) : 29 - 38
  • [4] Photoresponsive ion-selective optical sensor
    Shvarev, Alexey
    Journal of the American Chemical Society, 2006, 128 (22): : 7138 - 7139
  • [5] Photoresponsive ion-selective optical sensor
    Shvarev, A
    JOURNAL OF THE AMERICAN CHEMICAL SOCIETY, 2006, 128 (22) : 7138 - 7139
  • [6] CONTINUOUS MONITORING USING ION-SELECTIVE ELECTRODES
    ROSS, JW
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1976, 172 (SEP3): : 97 - 97
  • [7] INTRADIALYTIC ON-LINE MULTICOMPONENT REDUCTION RATIO MONITORING IN SPENT DIALYSATE BY A NOVEL MINIATURIZED OPTICAL SENSOR
    Pilt, Kristjan
    Arund, Jurgen
    Adoberg, Annika
    Leis, Liisi
    Fridolin, Ivo
    Luman, Merike
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2019, 34
  • [8] Estimation of delivered dialysis dose by on-line monitoring of the ultraviolet absorbance in the spent dialysate
    Uhlin, F
    Fridolin, I
    Lindberg, LG
    Magnusson, M
    AMERICAN JOURNAL OF KIDNEY DISEASES, 2003, 41 (05) : 1026 - 1036
  • [9] ON-LINE MONITORING OF URIC ACID CONCENTRATION IN SPENT DIALYSATE DURING HEMODIALYSIS ACCOMPANIED BY GRADUATED PHYSICAL EXERCISES WITH THE BISPECTRAL OPTICAL SENSOR
    Vasilevsky, Aleksandr
    Konoplev, Georgy
    Lopatenko, Oksana
    Komashnya, Artemy
    Visnevsky, Konstantin
    Gerasimchuk, Roman
    Neivelt, Indrek
    Frorip, Aleksandr
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 2012, 27 : 208 - 209
  • [10] CONTINUOUS MONITORING FOR AMMONIA AND NITRATE WITH ION-SELECTIVE ELECTRODES
    STEWART, MD
    ADVANCES IN INSTRUMENTATION AND CONTROL, VOL 44, PT 1-4, 1989, 44 : 287 - 303