Ion-selective optical sensor for continuous on-line monitoring of dialysate sodium during dialysis

被引:1
|
作者
Sharma, Manoj K. [1 ]
Frijns, Arjan J. H. [1 ]
Mandamparambil, Rajesh [2 ]
Kooman, Jeroen P. [3 ]
Smeulders, David M. J. [1 ]
机构
[1] Eindhoven Univ Technol, Dept Mech Engn, Eindhoven, Netherlands
[2] TNO, Eindhoven, Netherlands
[3] Maastricht Univ Med Ctr MUMC, Maastricht, Netherlands
关键词
Optical sensor; Micro-optofluidics; On-line electrolyte monitoring; Dialysis; Photoinduced electron transfer; FLUORESCENT; SYSTEMS;
D O I
10.1117/12.2251923
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Patients with end stage renal disease are dependent on dialysis. In most outpatient centers, the dialysate is prepared with a fixed electrolyte concentration without taking into account the inter-individual differences of essential electrolytes (sodium, potassium and calcium). This one-size fits all approach can lead to acute and chronic cardiovascular complications in dialysis patients. On-line monitoring of these essential electrolytes is an important physiological step towards patient specific dialysate leading to individualized treatment. Currently, changes in electrolyte concentrations are indirectly measured by conductivity measurements, which are not ion-specific. In this paper, we present a novel optical sensor for on-line monitoring of sodium concentrations in dialysate. This sensor is ion-specific and can detect up to a single ion. The working principle is based on the selective fluorescence quenching of photo-induced electron transfer (PET) molecules. The PET molecules when complexed with sodium ions start fluorescing upon laser excitation. The emission intensity is directly correlated to the sodium concentration. To prove the working principle, a micro-optofluidic device has been fabricated in polydimethylsiloxane (PDMS) with integrated optical fibers for fluorescence light collection. The PET molecules are covalently grafted in the PDMS microchannel for continuous monitoring of the sodium dialysate concentrations. The experimental setup consists of a laser module (lambda=450nm) operating at 4.5mW, a syringe pump to precisely control the sample flow and a spectrometer for fluorescence collection. The performance of the sensor has been evaluated for sodium ions ranging from 0-50mM. A clear signal and good response time was observed.
引用
收藏
页数:8
相关论文
共 50 条
  • [41] On-line electrochemical sensor for selective continuous measurement of acetylcholine in cultured brain tissue
    Niwa, O
    Horiuchi, T
    Kurita, R
    Torimitsu, K
    ANALYTICAL CHEMISTRY, 1998, 70 (06) : 1126 - 1132
  • [42] A highly sensitive ion-selective chemiresistive sensor for online monitoring of lead ions in water
    Darestani-Farahani, Maryam
    Montealegre, Isabella Mendoza
    Gilavan, Mehraneh Tavakkoli
    Kirby, Thomas
    Selvaganapathy, Ponnambalam Ravi
    Kruse, Peter
    ANALYST, 2024, 149 (10) : 2915 - 2924
  • [43] Use of on-line conductivity monitoring to study individually set dialysate sodium concentrations: A randomised controlled study.
    Lambie, SH
    McIntyre, CW
    JOURNAL OF THE AMERICAN SOCIETY OF NEPHROLOGY, 2003, 14 : 55A - 56A
  • [44] INTRACELLULAR SODIUM ACTIVITY AND BRETSCHNEIDER CARDIOPLEGIA - CONTINUOUS MEASUREMENT BY ION-SELECTIVE MICROELECTRODES AT INITIAL EQUILIBRATION
    STINNER, B
    KROHN, E
    GEBHARD, MM
    BRETSCHNEIDER, HJ
    BASIC RESEARCH IN CARDIOLOGY, 1989, 84 (02) : 197 - 207
  • [45] DESIGN AND CHARACTERIZATION OF A NOVEL AMMONIUM ION-SELECTIVE OPTICAL SENSOR BASED ON NEUTRAL IONOPHORES
    SEILER, K
    MORF, WE
    RUSTERHOLZ, B
    SIMON, W
    ANALYTICAL SCIENCES, 1989, 5 (05) : 557 - 561
  • [46] HV switchgear temperature on-line monitoring system with distributed optical fiber sensor
    Shi, Bin
    Gaodianya Jishu/High Voltage Engineering, 2007, 33 (08): : 169 - 173
  • [47] Optical System for On-Line Monitoring and Temperature Control in Selective Laser Melting Technology
    Doubenskaia, Maria
    Pavlov, Mikhail
    Chivel, Yuri
    MEASUREMENT TECHNOLOGY AND INTELLIGENT INSTRUMENTS IX, 2010, 437 : 458 - 461
  • [48] Non-Equilibrium Diffusion Controlled Ion-Selective Optical Sensor for Blood Potassium Determination
    Du, Xinfeng
    Xie, Xiaojiang
    ACS SENSORS, 2017, 2 (10): : 1410 - 1414
  • [49] A lithium ion-selective optical chemical sensor based on the inner-filter effect of fluorescence
    Zhong, Jing-Rong
    Liu, Miao-Gen
    Ao, Bing-Yun
    Yang, Xiao-Hai
    Wang, Ke-Min
    Xiao, Dan
    Kao Teng Hsueh Hsiao Hua Heush Hsueh Pao/ Chemical Journal of Chinese Universities, 2001, 22 (02): : 191 - 196
  • [50] A lithium ion-selective optical chemical sensor based on the inner-filter effect of fluorescence
    Zhong, JR
    Liu, MG
    Ao, BY
    Yang, XH
    Wang, KM
    Xiao, D
    CHEMICAL JOURNAL OF CHINESE UNIVERSITIES-CHINESE, 2001, 22 (02): : 191 - 196